Page 2
Careful internal layout and design minimize leakage and corona effects. All dividers are tested up to full voltage. The KVVB is equipped with three output binding posts (ground, low and output) and one ceramic post terminal (input). The input terminal a locking high voltage plug with high voltage cable terminated in a ring terminal.
Page 3
Connect the high voltage input lead wire terminated in a #10 ring terminal to the high voltage source. Connect the Low and Output binding posts of the KVVB to the meter or measuring system. Note that output Low is at Ground potential. See figure 1.
Page 4
To limit the error due to meter loading to less than 0.01 %, the meter impedance should be greater than 10 G For best accuracy, the loading effect on the KVVB can be calculated, and a correction can be applied to the measured result. To determine this correction, apply...
Page 5
2) Connect the high side of the decade resistor to the low terminal of the high voltage standard (HVS). 3) Connect an isolated null detector to the high side of the decade resistor and to the KVVB output terminal. 4) Connect high voltage to the HVS standard and to the input of the KVVB.
Page 6
Model KVVB Instructions Page 5 of 5 6) Service: The KVVB is sealed and contains no user serviceable parts. Replacement high voltage input cables are available in case the original is damaged or lost. 7) Warrantee: This divider is warranted against defects in manufacture for a period of two years from date of shipment.