GUILCOR AR006712 Instruction Manual

Co2 concentration, temperature, humidity and other calculated humidity variables regulator with rs232 serial output

Advertisement

Quick Links

CO
concentration, temperature, humidity
2
and other calculated humidity variables regulator
with RS232 serial output
AR006712
AR006711
AR006713
Instruction Manual

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the AR006712 and is the answer not in the manual?

Questions and answers

Summary of Contents for GUILCOR AR006712

  • Page 1 RS232 serial output AR006712 AR006711 AR006713 Instruction Manual...
  • Page 2: Table Of Contents

    9. PREVENTIVE MAINTENANCE AND CALIBRATION ........... 22 10. TECHNICAL SPECIFICATION.................... 23 10.1........................23 ECHNICAL PARAMETERS 10.1.1. Common parameters ............................. 23 AR006712 - CO 10.1.2. regulator ..........................23 AR006711 - CO 10.1.3. regulator ........................... 23 AR006713 - temperature, relative humidity and CO 10.1.4.
  • Page 3 Do not use the instrument in an explosive environment. Read instruction manual before the first device connection, please. Copyright: . It is prohibited to copy and edit this instruction manual and make any changes at all without explicit agreement of Guilcor. All rights reserved.
  • Page 4: General Description

    The regulators are designed to be operated in a chemically non-aggressive environment. Type Temperature Humidity Computed value  AR006712  AR006711     AR006713 The device is equipped with two relay outputs for alarming or controlling of external devices.
  • Page 5: Dimensions

    Similar effect can occur under water aerosol conditions. The communication cable must be placed as far as possible from potential interference sources. Electrical system (wiring) may do only worker with required qualification by rules in operation. 3. Dimensions AR006712...
  • Page 6 AR006711 AR006713...
  • Page 7: Typical Application Wiring, Connection Of Terminals

    4. Typical application wiring, connection of terminals...
  • Page 8: Description Of Device Function And Setting

    5. Description of device function and setting Information from regulator is available directly through LCD or you can read them with one of next communication protocols ( Modbus RTU, Advantech-ADAM ). 5.1. Device setting from the manufacturer communication protocol: Modbus RTU regulator address: communication speed: 9600Bd, no parity, 2 stop bits...
  • Page 9: Actual Regulator Preset

    applications. The dual wavelength NDIR CO2 sensing procedure compensates automatically for ageing effects. The CO2 module is highly resistant to pollution and offers maintenance free operation and outstanding long term stability. Measured values can be read in „SLOW mode“ (filtered, averaged) or in „FAST mode“...
  • Page 10: When Close Output Relay

    Far condition 1 FAR0 – „far condition 0“: after device’s power on the relay is open FAR1 – „far condition 1“: after device’s power on the relay is close If “far condition” is assigned to output relay, then there is possible to change the output relay status by Modbus communication protocol only.
  • Page 11: Regulator Setting Change

    5.5.1. Setting with PC and TSensor software User’s software TSensor provides setting of all regulator parameters. The program is available free to download at web page www.guilcor.com. Run the TSensor program and follow the program instructions. Regulator includes hardware write protection of internal setting , new setting is possible to write only while: 1.
  • Page 12: Extended Setting Mode

    5.6. Extended setting mode During regulator operation it is possible to change parameters available through “Actual regulator preset” only (see chapter 5.4 at page 9). Rest of regulator parameters are available through “Extended setting mode”. In this mode regulator does NOT communicate, does NOT do any measurements and does NOT service output relays.
  • Page 13 Relay response for error Regulator continuously checks its state during operation. In case error of assigned value measurement is found then output Relay1 (Relay2): stays unchanged (in previous status) Relay1 switch OFF Relay1 switch On stays unchanged (in previous status) Relay2 switch OFF Relay2 switch On password setting...
  • Page 14 Temperature unit choice It is possible to measure temperature at °C or °F. There is shown preset unit. It is possible to change it after „SET“ key pressing. preset to °C preset to °F measurement mode selection This item provides to choice if measured value is displayed and read in “SLOW” or “FAST” mode.
  • Page 15 Show temperature at LCD display This item provides to choice if actual measured temperature may be displayed at LCD display or not. It is possible to change it after „SET“ key pressing. show value on LCD don’t show value on LCD Show relative humidity at LCD display This item provides to choice if actual measured relative humidity may be displayed at LCD display or not.
  • Page 16: Restore To „Factory Default

    “Description of communication protocols”, which is available to download at www.guilcor.com, see detailed description of appropriate device type. After switching ON the power of the device it can last up to 3 s before the device starts to communicate...
  • Page 17: Modbus Rtu

    6.1. Modbus RTU Control units communicate on master-slave principle in half-duplex operation. Only master can send request and only addressed device responds. During sending of request no other slave station should respond. During communication, data transfer proceeds in binary format. Each Byte is sent as eight bit data word in format: 1 start bit, data word 8 bit (LSB first), 2 stop bits , without parity.
  • Page 18: Far Controlling Of Output Relays Status

    Legend:  * depends on device setting (by User’s software)  Int*10 register is in format integer *10 (likewise *100, *1000)  R register is designed only for reading  W** register is designed for writing, but direct write is not available, for details see description of communication protocols ...
  • Page 19 1) To address 0x0044 (68) write value1 – it enables Modbus setting. By the way it blocks regulator’s keyboard – local keyboard is blocked during remote control setting change. 2) Write the desired settings / changes to the addresses from 0x0045 (69) to 0x004E (78). New setting can be done simultaneously or each register can be written alone 3) To address 0x004F (79) write value 1.
  • Page 20: Protocol Compatible With Advantech-Adam Standard

    Address Address Function Unit Description Status [hex] [dec] 0 ... alarm occurs, when measured value is When close LOWER than preset alarm value 0x004B Relay2 1 ... alarm occurs, when measured value is HIGHER than preset alarm value Alarm value 0x004C preset alarm value (threshold) for Relay2...
  • Page 21: Error States Of The Device

    For communication with device which measure more than one value, there is necessary to add at the end of command, number of communication channel, where the measured value is mapped. Command for value reading is #AAx(CRC) CR, where AA is device address and x is number of communication channel by next table: Measured value Number of communication channel...
  • Page 22: Readings On Cd Display

     Computed value – calculation of computed value is not possible (error during measurement of temperature or relative humidity). Error 3 - there is a reading „Err3“ on LCD display upper line. Error of internal A/D converter appeared (converter does not respond, probably damage of A/D converter). At this state device does not measure temperature and relative humidity.
  • Page 23: Technical Specification

    60 VA Relays contact is not designed for direct control of line voltage! (output relay contact parameters: max 220Vdc, 125Vac, 2A, 60 W, 62.5 VA) EMC: EN 61326-1, EN 55011 10.1.2. AR006712 - CO regulator Concentration of CO Accuracy: ± (100 ppm + 5 % of measuring value) at 25°C (77°F) and 1013 hPa...
  • Page 24: Ar006713 - Temperature, Relative Humidity And Co Regulator

    10.1.4. AR006713 - temperature, relative humidity and CO regulator Temperature: Accuracy: ± 0,4 °C (±0,7 °F) Range: -30 to +60 °C (-22 to 140 °F) Resolution: 0,1 °C (0,2 °F) Relative humidity: Accuracy: ± 2,5 %RH from 5 to 95 %RH at 23 °C (73,4 °F) Range: 0 to 100 %RH Resolution:...
  • Page 25: Value Computed From Measured Relative Humidity And Temperature

    10.1.6. Value computed from measured relative humidity and temperature Dew point temperature Range: -60 to +80 °C (-76 to 176 °F) Accuracy: ±1,5°C (±2,7°F) at ambient temperature T<25°C (77°F) and relative humidity RH >30%, for more details see graphs below Absolute humidity Range: 0 to 400 g/m...
  • Page 26: Relative Humidity And Temperature Restriction

    10.1.7. Relative humidity and temperature restriction 10.2. Operating conditions Operating temperature and relative humidity: -30 to +80 °C (-22 to 176 °F) electronics AR006712: electronics AR006711, AR006713: -30 to +60 °C (-22 to 140 °F) measuring end of stem AR006713: -30 to +80 °C (-22 to +176 °F)
  • Page 27: End Of Operation

    Dimensions:see dimensional drawings Weight: approximately: AR006712/1m probe 440 g AR006712/2m probe 470 g AR006712/4m probe 530 g AR006711 340 g AR006713 360 g Weight of the device is given without RS232 communication cable (cable weight is 70g). Material of the case: ABS 10.3.
  • Page 28: Technical Support And Service

    The adjustment procedure is described at file „Calibration manual.pdf“ which is installed commonly with the TSensor software. Technical support and service is provided by distributor. For contact see warranty certificate. You can use discussion forum at web address: http://www.guilcor.com/. 11. Appendix A Connection of external power relay...

This manual is also suitable for:

Ar006711Ar006713

Table of Contents