DATA SHEET Blackbody Radiator SW40 / SW40NT Serial number (radiator) ..............Installed immersion circulator Serial number immersion circulator ............................Setting of immersion circulator Set point min Set point max ............................Options ................................................................................Accessories ................................................................................
Contents ........................5 General Safety Instructions................6 Identification of symbols and terms ............... 9 Intended use ....................10 Description of SW40 / SW40NT ..............10 Installation ...................... 12 Set up the radiator .................... 12 Fill the radiator ....................13 Operation ......................14 Determine temperature ..................
This manual is addressed to qualified experts who are competent in setting up measuring and control systems. HEITRONICS measuring instruments are characterized by accurate temperature measurement. This is only possible if the instruments are installed, operated and maintained properly. Handling by insufficiently trained personnel may cause incorrect measured temperature values which are misinterpreted and which provoke defects within the production process.
Page 7
110 °C. « The protection must not cover the radiator aperture. « The protection must not interfere with the sight path of the instrument to be calibrated. SW40 / SW40NT 95583226 General Safety Instructions HEITRONICS Infrarot Messtechnik GmbH...
Page 8
Observe the safety data sheet of the bath fluid used. If you have any questions about how to use the radiator without any risk and in accordance with its function, please contact HEITRONICS Infrarot Messtechnik GmbH. Our employees will be available to support you!
2 Identification of symbols and terms Important information which supports a safe and correct operation of the blackbody radiator SW40 / SW40NT: SYMBOL KEYWORD HAZARDOUS SITUATION WARNING Hazards marked in this way may lead to serious injury or even death if they occur.
3 Intended use Due to its high emissivity (ε ≥ 0,995) which is independent of wavelength in the infrared range, SW40 / SW40NT is suitable for calibrating infrared radiation thermometers, LineScanner and thermal imaging cameras. The device to be calibrated is positioned in front of the radiator aperture. The radiated temperature is determined either with a reference radiation thermometer or with a contact probe (e.g.
Page 11
Contact probe option for measuring the actual temperature of the radiator, independent of the setting and temperature indication of the immersion circulator. The choice and quality of the contact probe determines the thermodynamic temperature accuracy of the radiator. SW40 / SW40NT 95583226 Description of SW40 / SW40NT HEITRONICS Infrarot Messtechnik GmbH 01/08/21e...
» This prevents the radiator from tipping over during operation. 3. Align the radiator horizontally. » This prevents bath fluid from flowing over the edge of the cylinder. » The radiator is securely set up. SW40 / SW40NT Installation 95583226 HEITRONICS Infrarot Messtechnik GmbH 01/08/21e...
When using glycols containing additives (such as coolants for cars), make sure that the water-glycol-mixture is compatible with the materials of the radiator. The interior of the SW40NT is made of stainless steel and copper. SW40 / SW40NT 95583226 Installation...
∗ with a contact probe and emissivity adjustment or ∗ with a reference radiation thermometer (steps 1. to 3.). 5. Document the actual temperature of the radiator. » The calibration has been performed successfully. SW40 / SW40NT Operation 95583226 HEITRONICS Infrarot Messtechnik GmbH 01/08/21e...
» The nitrogen tube ends at the end of the radiator tube. » Icing inside the radiator tube is largely suppressed. Icing inside the radiator tube may be suppressed alternatively with dry, clean and oil-free air. SW40 / SW40NT 95583226 Upkeep and maintenance HEITRONICS Infrarot Messtechnik GmbH...
2. Carefully pour out the bath fluid. 3. If you use silicone oil and it drips, remove the silicone oil from the ground immediately. » The radiator is emptied. SW40 / SW40NT Upkeep and maintenance 95583226 HEITRONICS Infrarot Messtechnik GmbH 01/08/21e...
« The radiator is filled. 1. Empty the radiator. 2. Rinse the radiator. 3. Fill the radiator with the bath fluid. » The bath fluid has been changed. SW40 / SW40NT 95583226 Upkeep and maintenance HEITRONICS Infrarot Messtechnik GmbH 01/08/21e...
HEITRONICS is not responsible. Warranty does not include batteries. The device must be returned to HEITRONICS in the original packing, carriage paid. Costs for taxes, fees and customs duties are to be paid by sender. Shipping damage is borne by the sender.
Vous pouvez trouver les coordonnées de nos représentants régionaux sur notre site Web www.heitronics.eu. Organizacion de la venta Informaciónes referente a nuestros regionales colaboradores de venta se encuentran en nuestra pagina web www.heitronics.eu. SW40 / SW40NT Service Adress 95583226 HEITRONICS Infrarot Messtechnik GmbH...