Percussionaire TXP 5 Quick Reference Manual

Percussionaire TXP 5 Quick Reference Manual

High-frequency percussive ventilator
Hide thumbs Also See for TXP 5:

Advertisement

Quick Links

TXP
1. Mount TXP
5 to pole.
2. Connect AC power source.
3. Connect gas source.
4. Connect breathing circuit.
5. Always perform a pre-use check before
ventilating any patients.
6. Recommended HME/HMEF bacterial/viral
filter with an efficiency
of >99.999% with a
2 cmH
O or less
2
flow resistance
at 60 L/min.
Optional PEEP
valve with or
without filter
Optional
one-way
valve
QUICK REFERENCE GUIDE
Simple System Setup
Optional secondary
filters may be added
to the two 15 mm ports,
either directly or with tubing
connectors.
This device is
NOT an oscillator (HFOV) NOT a jet (HFJV)
NOT a Conventional Ventilator
It is a High-Frequency
Percussive Ventilator (HFPV)
7. Set Frequency arrow straight
up for starting position, which
will produce a high-frequency
rate of approximately 500.
8. Connect to patient invasively or
non-invasively.
9. Turn Amplitude counterclockwise
to start ventilation.
Display will turn on as soon as ventilation pressure
is detected above 2.5 cmH
Alarms default
set 45 and 3
10. Press once for alarm
silence if needed.
Supplemental oxygen can be
added after filter
Intubated patient with Phasitron®
O for 5 seconds.
2