User-Service Manual
Dolphin Fluid Immersion Simulation
System
®
To avoid potential injury, read user's manual before using.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Dolphin Fluid Immersion Simulation System and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for onCare Dolphin Fluid Immersion Simulation System

  • Page 1 User-Service Manual Dolphin Fluid Immersion Simulation System ® To avoid potential injury, read user’s manual before using.
  • Page 2: Important Precautions

    Immersion Simulation System, unplug it from its reading and understanding the product user manual power source. and contacting OnCare if anything in this manual is • Do not place or store the product where it can fall unclear. OnCare will not be held responsible for any or be pulled into a tub or sink.
  • Page 3: Table Of Contents

    User Manual Table of Contents Dolphin Fluid Immersion Simulation System ® Important Precautions .........................2 Introduction ............................4 Shear and Friction Reduction Indications for use ..........................4 Spinal Cord Injury Pressure Redistribution Pain Management Features ...............................5 Therapy Control Unit Features Therapy Mattress and Specialty Surface Features Therapy Pads Additional Features ..........................6 Battery Back Up...
  • Page 4: Introduction

    Dolphin Fluid Immersion Simulation System ® Introduction OnCare’ Dolphin System is suitable for the treatment and prevention of all stages of pressure ulcers, for patients who have been assessed to be at risk for The Dolphin Fluid Immersion Simulation (FIS) ®...
  • Page 5: Pressure Redistribution

    User Manual Dolphin Fluid Immersion Simulation System ® Features Pressure Redistribution Amputations Grafts The Dolphin FIS System is comprised of two Burns Neurology components: Dermatology Pressure Ulcers • Therapy control unit Flaps Rehabilitation • Therapy mattress system or specialty surface (wheelchair cushion, stretcher pad) Pain Management Therapy Control Unit Features (Figure 1)
  • Page 6: Therapy Mattress And Specialty Surface Features

    User Manual Dolphin Fluid Immersion Simulation System ® Additional Features Therapy Mattress System and Specialty Surface Features Warning: A possible fire hazard exists. This product • State-of-the-art pressure redistribution technology is suitable for use with oxygen administering designed to alleviate vertical shear forces. equipment of the nasal, mask, or half bed-length, •...
  • Page 7: Keyboard Functions

    User Manual Dolphin Fluid Immersion Simulation System ® Keyboard Functions Warning: For important precautions, see page two. Caution: The patient’s head should be positioned in the center of the top section of the therapy mattress. When using the therapy mattress system, always ensure that the patient is positioned properly within the confines of the bed.
  • Page 8 Comfort Adjust arrows. Figure 6 Comfort Adjust OnCare recommends that caregivers allow the Dolphin FIS System to set and control the immersion profile. However, to accommodate individual patient preference, caregivers can press the Comfort Adjust arrows to manually adjust comfort settings.
  • Page 9: Grounding Instructions

    If repair or replacement of the cord or plug is 250 lbs., moving the Comfort Adjust indicator necessary, please contact OnCare for assistance. to one LED above the Auto Feedback LED may improve comfort. If patient is less than 100 lbs., moving the Comfort Adjust indicator to one LED below Auto Feedback may improve comfort.
  • Page 10: Specialty Surface - Wheelchair Cushion

    User Manual Dolphin Fluid Immersion Simulation System ® 3. Hand Check: Place a hand with three (3) Troubleshooting fingers (if head of bed at 30° or higher) or four (4) fingers (if head of bed lower than Support Surface is Not Inflating 30°) stacked vertically beneath the cells of the mattress and above the safety mattress •...
  • Page 11: Changing The Therapy Pad

    User Manual Dolphin Fluid Immersion Simulation System ® Safety Information Changing the Therapy Pad • Place the therapy pad over the therapy cells, fitting When using the Dolphin FIS System, always ensure the corner of the cushions into the corner of the that the patient is positioned properly within the therapy pad (similar to a fitted sheet).
  • Page 12: Cleaning

    User Manual Dolphin Fluid Immersion Simulation System ® Cleaning Disinfecting When there is visible soilage and between patients, Warning: Unplug the control unit from its power we recommend that you disinfect the unit and source. Failure to do so could result in personal therapy mattress or specialty surface with a injury or equipment damage.
  • Page 13: Maintenance

    Call for Assistance Weight If you have any questions or require service Control Unit: ........10 lbs (4.5 Kg) on a OnCare product, please call OnCare at Therapy Mattress: ......22 lbs (10 Kg) 800.847.7368. Wheelchair Cushion: ......5 lbs (2.3 Kg)
  • Page 14 User Manual Dolphin Fluid Immersion Simulation System ® Appendix A: Electromagnetic Compatibility (EMC) Information AutoVector Fluid Immersion Simulator ® Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in this manual. Portable and mobile RF communications equipment can affect Medical Electrical Equipment.
  • Page 15 User Manual Dolphin Fluid Immersion Simulation System ® Guidance and Manufacturer’s Declaration - Electromagnetic Immunity: The AutoVector Fluid Immersion Simulator is intended for use in the electromagnetic environment specified ® below. The customer or the user of the Auto Vector Fluid Immersion Simulator should assure that it is used in such an environment.
  • Page 16 User Manual Dolphin Fluid Immersion Simulation System ® Guidance and Manufacturer’s Declaration – Electromagnetic Immunity The AutoVector Fluid Immersion Simulator is intended for use in the electromagnetic environment specified ® below. The customer or the user of the Auto Vector Fluid Immersion Simulator should assure that it is used in such an environment.
  • Page 17 User Manual Dolphin Fluid Immersion Simulation System ® Recommended separation distances between portable and mobile RF communications equipment and the Auto Vector Fluid Immersion Simulator The AutoVector Fluid Immersion Simulator is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
  • Page 18: Warranty

    (compressors, valves, printed circuit boards, hoses, option, any product which is found by OnCare to be and couplers). Damages arising from improper use defective. When requested by OnCare, parts must be will not be covered by this warranty.