The calibration system is a calibrated instrument qualified in a UKAS certified laboratory. The accuracy of the MCS7500 is calibrated to meet or exceed: 1% of Full Scale Deflection (FSD) from 2% to 8% of torque range and 1% of reading from 8% to 100% of torque range.
Digital display module –The digital display module facilitates the display, collection, archiving, and processing of torque readings. The module can run off battery power or plugged into an outlet. Each MCS7500 is equipped with a global outlet adaptor that allows the tool to be used with many electrical systems.
Fig. 3 - USB/ Battery charge port 1.3.1 DC Interface The interface for the 5V DC USB AC adapter is supplied with the MCS7500. Use the adapter if you plan on using the main power source. Use only the AC adapter provided with the MCS7500. If another power source is used it will void the warranty and may cause severe damage to the digital display module.
• The battery is charged any time the MCS7500 is plugged-in. The charge time is between 2 and 4 hours depending on the battery charge level. It is recommended that the user unplug the unit when it is not in use.
Read all introductions carefully. Follow all recommended safety precautions to avoid personal injury as well as damage to the product and / or damage to other property. Enerpac cannot be responsible for any damage or injury from unsafe use, lack of maintenance, or incorrect operation. Do not remove warning labels, tags, or decals.
• When operating, ensure full contact is achieved between the reaction arm and the reaction stop. • The maximum torque rating of the MCS7500 is 7375 Ft.lbs (10,000 Nm). Never exceed the torque rating of the MCS7500. Use and Care •...
SETUP AND ADJUSTMENTS Mount the Torque Wrench • Remove the MCS7500 components from the packaging and place on a flat surface. Ensure that the digital display module is plugged in or sufficiently charged to operate. • Select and position an appropriate adaptor onto the run down fixture to allow the proper mounting of the torque wrench.
Slide the reaction stop to the desired position. Lock into place by rotating the lock handle located directly under the reaction stop. Once the MCS7500 unit has been appropriately adjusted, complete the mounting of the tool. Handle Fig. 6 - Securing reaction arm...
Enerpac MCS7500 Mobile Calibration System Torque To Spreadsheet Software (TTS) The Torque To Spreadsheet (TTS) software is an intuitive application that allows users to connect to an MCS7500 or digital transducer through a USB cable. This application facilitates the display, collection, archiving, and processing of torque readings.
The digital display has five main buttons: ON/OFF, MENU, PEAK/TRACK, LEFT AND RIGHT (Figure 7, Items 1-4). Fig. 7 - MCS7500 Digital Display The display module can be set up to perform other functions, such as tracking the torque. The amount of time that the reading is displayed cannot be adjusted.
NOTE: The TTS software should detect the MCS7500 under “Available COM Ports”. If the TTS is launched prior to attaching the cable, the user must select “Rescan COM ports” to prompt the TTS to identify the MCS7500 as an available COM Port.
1. Ensure the correct drivers and software are loaded onto the PC to be used for the calibration, as stated in section 4.2 2. Connect the MCS7500 to the PC using the supplied USB type B to type A cable. 3. Run “EnerpacVses.exe” by double clicking the icon. The following window will be displayed: Fig.
Page 15
Enerpac MCS7500 Mobile Calibration System 5. Click the “Connect” button to connect the PC to the MCS7500 unit. The display window will now update and show “Listening on COM2” in the top left of the window, see Figure 9: Fig. 9 - TTS Display after connecting to PC (ETW) 6.
Page 16
Enerpac MCS7500 Mobile Calibration System 7. Enter in all the required information into the text fields. In “UUT Type” drop-down box select “Electric”. See Figure 11: Fig. 11 - Updating Calibration Mode (ETW) 8. Click the “Start Test Routine” button to start the calibration.
Page 17
Enerpac MCS7500 Mobile Calibration System 12. Actuate the trigger on the wrench. The dynamic joint fixing will start to rotate and the wrench will shut off when it hits its target torque. Once the cycle has finished, Sample 1 of data point 1 will be populated automatically, with the reading taken during the run down.
Page 18
Fig. 14 - Completed Test Routine Spreadsheet (ETW) 15. On the main Enerpac VSES program, click the “Export to pdf” button to create the certificate. The program will create a spreadsheet with an input tab and a certificate tab, at the bottom of the spreadsheet: Fig.
Page 19
Enerpac MCS7500 Mobile Calibration System 16. Select the certificate tab to view the calibration certificate. Fig. 16 - Calibration Certificate tab (ETW) Calibration result, readings taken and a graph to show deviation from target (command torque) are all displayed within the certificate in both Ft.lbs and Nm.
1. Ensure correct drivers and software are loaded onto the PC in use, as stated in section 4.2 2. Connect the MCS7500 to the PC using the supplied USB type B to type A cable. 3. Run “EnerpacVses.exe” by double clicking the icon. The following window will be displayed: Fig.
Page 21
Enerpac MCS7500 Mobile Calibration System 5. Click “Connect” button to connect the pc to the MCS7500 unit. The window will now update and show “Listening on COM2” in the top left of the window, see Figure 19: Fig. 19- TTS Display after connecting to PC (PTW) 6.
Page 22
Enerpac MCS7500 Mobile Calibration System 7. Enter in all the required information into the text fields. In “UUT Type” drop-down box, select “Pneumatic”. See Figure 21: Fig. 21 - Updating Calibration Mode (PTW) 8. Click the “Start test routine” button to start the calibration.
Page 23
Enerpac MCS7500 Mobile Calibration System 11. Set the target torque on the wrench controller based on the air pressure calculation from the previous step, as specified in blue on the spreadsheet for data point 1. In this case it is 200 Ft.lbs. See Figure 22: Fig.
Page 24
Enerpac MCS7500 Mobile Calibration System An average of the readings and status of reading will also be populated into the spreadsheet. The last reading can be cancelled by clicking the “Cancel last sample” button. This may be necessary if the joint was not backed off before the run down or an incorrect target torque was set.
Page 25
Enerpac MCS7500 Mobile Calibration System 15. On the main Enerpac VSES program, click the “Export to pdf” button to create the certificate. The program will create a spreadsheet with an input tab and a certificate tab, at the bottom of the spreadsheet: Fig.
Page 26
Enerpac MCS7500 Mobile Calibration System 16. Select the certificate tab to view the calibration certificate. Fig. 25- Calibration Certificate tab (PTW) Calibration result, readings taken and a graph to show torque vs air pressure in psi/ Ft.lbs and bar/Nm. Fig. 26- Completed Calibration Certificate (PTW) A copy can be printed, by selecting File, Print, and/or a copy can be saved to any preferred directory by selecting File, Save as and selecting the directory in which the copy needs to be saved.
Enerpac MCS7500 Mobile Calibration System DIGITAL DISPLAY MODULE Digital Display Buttons Fig. 27 - Digital Display Buttons Power key – Turns the display On and Off and zeros the transducer. Menu – Accesses Torque units and Filter settings. Left and Right keys – Navigate through the menu and allows access to stored readings.
The peak mode is recommended for rotary tools and wrenches. Track The track mode displays the torque as it is being applied to the transducer. This mode is used primarily for verifying the calibration of the MCS7500 unit. Peak Track Fig.
Enerpac MCS7500 Mobile Calibration System Filter Use this option to select the filter value. Press the Menu button twice and using the arrow keys, select the required Filter. Selectable filter values in Hz (125, 250, 500, 1000, 2000, 4000 and 5000).
Enerpac MCS7500 Mobile Calibration System MAINTENANCE The MCS7500 Calibration System requires very little maintenance, but periodic lubrication of the run down fixture is critical to the longevity of the tool. The run down bolt and disc spring should be periodically removed and inspected, and if necessary, coated with a high pressure anti-seize lubricant.
It is recommended that the unit is recalibrated every 12 months or 500 operations, whichever is sooner. The accuracy of the MCS7500 is calibrated to meet or exceed: 1% of FSD from 2% to 8% of torque range and 1% of reading from 8% to 100% of torque range.
Repair and (Re)Calibration All (re-)calibrations and all repairs on the Enerpac MCS7500 calibration stands will be handled by Enerpac Authorized Service Centers. ESSA: The nearest Enerpac Authorized Service Center can be found at: https://www.enerpac.com/en-gb/distributors...
Operating Temperature Range 32°F [0°C] to 122°F [50°C]. Data Communications The accuracy of the MCS7500 is calibrated to meet or exceed: 1% of Accuracy FSD from 2% to 8% of torque range and 1% of reading from 8% to 100% of torque range.
Need help?
Do you have a question about the MCS7500 and is the answer not in the manual?
Questions and answers