Monitor Calibration Unit
Operation Instructions
Figure 1. Desco
98220

Description

The Desco
Monitor Calibration Unit is designed to
98220
simplify the process of calibrating Workstation Continuous
Monitors. The unit allows the user to quickly and easily
verify whether a tester is operating within specifications. The
Monitor Calibration Unit is a passive device and requires no
power source.
"A Compliance Verification Plan shall be established
to ensure the Organization's fulfillment of the technical
requirements of the ESD Control Program Plan (ANSI/
ESD S20.20-2007 section 7.3). The product qualification
for the Continuous Monitors per ANSI/ESD S20.20-2007
Table 3 is "User defined" test method and required limits,
and compliance verification is per ESD TR53 with the
required limit being "Manufacturer defined." Per ESD TR53
Compiance Verification "Compliance verification of the
constant monitor device should follow the manufacturer's
instructions."

Packaging

1
Monitor Calibration Unit
1
Alligator Clip
1
Stacking Snap Banana Jack Adapter
1
Banana Jack to Wire Adapter
1
Certificate of Calibration
Calibration Test Procedures
The Monitor Calibration Unit was specifically designed for
use in calibration of Desco brand equipment. The following
step by step procedures will cover calibration for specific test
units. The procedures will not cover adjustment of the test
equipment. For detailed information regarding adjustment of
specific Desco brand testers contact our Customer Service
Department at 909-627-8178.
Rev. December 2009
TB-2082 Page 1 of 6
TECHNICAL BULLETIN TB-2082
Monitor Calibration Unit

Calibrating the Full Time Continuous Monitor

The Full Time Continuous Monitor is available as following
item number:
Item
19225
Step 1: Connect the Calibration Unit's banana plug wire
labeled "GROUND" to a known ground. Do not plug the
ground lead from the Calibration Unit to the monitor's
common point ground banana jack. There is a 47 kilohm
resistance to ground.
Step 2: Insert the Calibration Unit's banana plug labeled
"OPERATOR" into the jack labeled "OPERATOR" on the Full
Time Continuous Monitor (see Figure 2). Power the monitor.
Figure 2. Connecting the OPERATOR test lead from
the Monitor Calibration Unit to the Full Time Continuous
Monitor
Step 3: Start the Calibration Unit's selector knob at the full
counter-clockwise position. When testing the OPERATOR
circuit, test only the first four limits on the Calibration Unit.
Rotate the selector knob to the "LOW FAIL" position; the
red "OPERATOR GROUND" LED on the monitor should
illuminate and alarm.
Made in America
Voltage
120 VAC
© 2009 DESCO INDUSTRIES, INC.
Employee Owned

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the TB-2082 and is the answer not in the manual?

Questions and answers

Summary of Contents for Desco TB-2082

  • Page 1 Calibration Test Procedures Monitor The Monitor Calibration Unit was specifically designed for use in calibration of Desco brand equipment. The following Step 3: Start the Calibration Unit’s selector knob at the full step by step procedures will cover calibration for specific test counter-clockwise position.
  • Page 2 LED should illuminate. Step 4: Rotate the selector knob clockwise to the “LOW LIMIT” position of the green pass sector. The monitor’s corresponding green “OPR1” or “OPR2” LED should illuminate. TB-2082 Page 2 of 6 © 2009 DESCO INDUSTRIES, INC. Employee Owned...
  • Page 3 Calibration Unit. Rotate the selector knob to the “MAT1” or “MAT2” green LED should illuminate. “LOW FAIL” position; the red “OPERATOR GROUND” LED on the monitor should illuminate and alarm. © 2009 DESCO INDUSTRIES, INC. TB-2082 Page 3 of 6 Employee Owned...
  • Page 4 The alarm will shut off approximately 8-10 seconds after the removal of the banana lead. TB-2082 Page 4 of 6 © 2009 DESCO INDUSTRIES, INC. Employee Owned...
  • Page 5 11.4 cm x 6.4 cm x 3.0 cm Pass Fail V Measure ~ 50 10 Mat 8 Megohms 12 Megohms 100 Mat 80 Megohms 120 Megohms 500 Mat 400 Megohms 600 Megohms © 2009 DESCO INDUSTRIES, INC. TB-2082 Page 5 of 6 Employee Owned...
  • Page 6 Limit of Liability In no event will Desco or any seller be responsible or liable for any injury, loss or damage, direct or consequential, arising out of the use of or the inability to use the product. Before using, users shall determine the suitability of the product for their intended use, and users assume all risk and liability whatsoever in connection therewith.