Characteristics
· Compatible with electric Volatimeter GAB
· Automatic test tube swaping system
· It stops the heater at the end of the distillation without the user intervention
· Exact dosification through optical sensors: it guarantees result precision and its reliability
· Unattended operation
· Single button control
· It saves a notable amount of time and energy
· Drastically reduction of flasks breakage
· Power: 110-230 V / 50-60 Hz
Equipment configuration
·
1 inox structure
·
4 tenamatic test tubes 15x100
·
1 tenamatic conection cable
·
1 small silicone tube + 2 joints
·
1 instructions manual
Related products
· GAB volatimeter
Instructions
Assembly and previous operation
01
Place the Volatimeter over the Tenamatic (see backwards photo).
02
Join a silicone tube from the outflow of the cooler to the inox tube of the inox arc.
03
Connect with the bridge power cable from the left backward plug of the Tenamatic to the power
plug of the Volatimeter.
04
Connect the power cable from the right backward plug from the Tenamatic to the network.
Prepare the flask with 11ml of sample to analyze, 1-2 granules of siliconed pumice stone and
05
06
place it over the Volatimeter heater.
Open the water key to cool the Volatimeter.
07
Switch on the Tenamatic backward's connector and make sure that the one from Volatimeter is
also switched on (1 position).The front of the Tenamatic will shine in cyan color [0 state]
08
Change or place a clean and dry tube in the 3.2ml location.
09
Press the front button once, the front light will shine in dark blue colour
change or place a clean and dry tube in the 5.1ml location.
Press again the button and the front led will shine in red colour
10
previously prepared starts heating.
11
After few minutes the sample will boil and distillation will begin. When level of 5.1ml is reached
the system will turn automatically 90º clockwise and [state 3] will begin, f
and will center 3.2ml tube under the center of the arc.
GAB Sistemática Analítica S.L.
Sant Jordi, 30. Moja-Olèrdola 08734. Bcn [Spain]
ref. 3026212
ref. 36.760-180-BP
ref. 1010001
Tenamatic GAB (ref. 1010016)
[1 state]
[2 state]
.At this state the sample
ront led become purple
Tel. +34 938 171 842
Fax. +34 938 171 436
www.gabsystem.com
gab@gabsystem.com
. In this state