U S E R M A N U A L OXYGEN PEDIATRIC FLOWMETER MODELS: 1MFA3001 (SHOWN) 4MFA1001 6MFA1001 Authorized EU Representative: EMERGO EUROPE, INC. Molenstraat 15 2513 BH The Hague NETHERlaNdS SAVE THESE INSTRUCTIONS Federal (USa) law restricts this device to CAUTION sale by or on the order of a physician.
Page 2
RECEIVING / INSPECTION Remove the Precision Medical, Inc. Flowmeter from the packaging and inspect for damage. If there is any damage, dO NOT USE and contact your Provider. INTENDED USE The Flowmeter is intended for use by physicians, respiratory therapists and other authorized hospital personnel to administer selected doses of medical oxygen to a patient.
Page 3
WARNING • Use this Flowmeter only for its “Intended Use” as described in this manual. • alWaYS confirm prescribed flow before administering to patient and monitor flow on a frequent basis. • This Flowmeter contains magnetic, ferrous material that may affect the results of an MRI.
Page 4
SPECIFICATIONS Model 6MFA1001 4MFA1001 1MFA3001 0-200 cc 0-1 lpm 0-3 lpm Flow Range Oxygen Oxygen Oxygen 20 cc .1 lpm .125 Increments (starts at 20 cc) (starts at .1 lpm) from .125 to 1 lpm .25 lpm from 1 to 3 lpm ±...
Page 5
Inspect the Flowmeter for visual damage before use, dO NOT USE if damaged. NOTE: Precision Medical, Inc. strongly recommends the use of kink proof Cannula. Turn Knob to the “OFF” position. Connect the Flowmeter to a 50 psi oxygen gas source. For international products, connect to appropriate oxygen source pressure.
Page 6
• Replace Body assembly RETURNS Returned products require a Returned Goods authorization (RGa) number. any product returned to Precision Medical, Inc. must be packaged in a sealed container to prevent damage. Precision Medical, Inc. will not be responsible for goods damaged in transit.
Page 7
International parts specifications and specific ratings are available upon request. International flowmeters are calibrated per specifications on flowtube. DECLARATION OF CONFORMITY Manufacturer: Precision Medical, Inc. 300 Held drive, Northampton, Pa 18067, USa CONTaCT: Quality Manager Phone: 610-262-6090 authorized European Representative: Emergo Europe...
Page 8
ORal STaTEMENTS dO NOT CONSTITUTE WaRRaNTIES. The representative of Precision Medical, Inc. or any retailers are not authorized to make oral warranties about the merchandise described in this contract, and any such statements shall not be relied upon and are not part of the contract for sale.