Operating Manual
Monitored Incubator Option 2 (MIO™)
6 Positions
4 Positions
MIO2 Operating Manual, 394934, en, V3.0

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the MIO2 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for tecan MIO2

  • Page 1 Operating Manual Monitored Incubator Option 2 (MIO™) 6 Positions 4 Positions MIO2 Operating Manual, 394934, en, V3.0...
  • Page 2 2017-08-31 Corrected shutters’ description in chapter “Part Locations” Added chapter “Customer Support” MIO2 is available for Freedom EVO and Fluent “Product Safety Signs” added © 2017, Tecan Trading AG, Switzerland, all rights reserved. Information contained in this document is subject to change without notice.
  • Page 3: Table Of Contents

    Part Locations ......... . . MIO2 Operating Manual, 394934, en, V3.0...
  • Page 4 Operating the MIO2 ........
  • Page 5: Scope Of This Manual

    CH-8708 Männedorf Switzerland 1.3 Intended Use The MIO2 is used on Tecan instruments to incubate and / or shake substances in microplates. The MIO2 is an instrument component. The intended use is defined for the instrument where it is used on. Refer to the corresponding Operating Manual of ...
  • Page 6: Improper Use

    1.4 Improper Use Improper use may prejudice the MIO2 safety concept.  The MIO2 must not be used with options or components that are not approved by Tecan.  The MIO2 is not explosion-proof and should not be installed in locations where there is a hazard of explosion.
  • Page 7: Instrument Manuals

    1-1 refers to the page number, whereas the first number stands for the chapter number (chapter 1 - page 1)  Note: For safety conventions and symbols refer to chapter 2 “Safety”, 2-1. MIO2 Operating Manual, 394934, en, V3.0 1 - 3...
  • Page 8: Abbreviations

    1 - About This Manual Abbreviations 1.10 Abbreviations Field service engineer Monitored Incubator Option 1 - 4 MIO2 Operating Manual, 394934, en, V3.0...
  • Page 9: Safety Message Conventions

    2 - Safety Safety Message Conventions 2 Safety This chapter describes the safety concept of MIO2, provides general rules of correct behavior, and warnings concerning hazards associated with the use of the MIO2. 2.1 Safety Message Conventions 2.1.1 Signal Words WARNING indicates the possibility of personal injury or even loss of life if the instructions are not followed.
  • Page 10: General Safety Information

    FSE certified for the repair and upgrading of the MIO2. Tecan will reject any claim resulting from unauthorized modifications.  Fire hazard caused by the improper use of the MIO2. The MIO2 should not be installed in locations where there is a hazard of explosion. ...
  • Page 11: Operating Company

    Let the MIO2 cool down hand-hot before removal. 2.3 Operating Company The operating company must ensure that the MIO2 and in particular the safety features, function properly and that all the personnel in contact with the instrument are adequately trained.
  • Page 12: Operator

    User Qualification  and have read and understood the instructions in the Operating Manual. Tecan recommends that the operator attends an operator training course. Please ask the Tecan Customer Service about available courses. Refer to section  10-1. 10 “Customer Support”, 2.4.1 Operator...
  • Page 13: Product Safety Signs

    Contact your local service organization to obtain this form and refer to section 7.1 “Decontamination”, 2 7-1. Note: Tecan reserves the right to refuse to deal with any MIO2 or its parts or accessories that is not accompanied by the Decontamination Declaration form. MIO2 Operating Manual, 394934, en, V3.0...
  • Page 14 2 - Safety Decontamination Declaration 2 - 6 MIO2 Operating Manual, 394934, en, V3.0...
  • Page 15: Overview

    MIO2 Incubator with 6 positions (slots) Incubator with 4 positions (slots) Configurations The MIO2 is provided in a 4 and a 6 slot configuration with the following features: Standard  The standard configuration does not contain the shaking option. Shaking Incubator ...
  • Page 16: Product Identification And Appendix Labels

    3 - Technical Data Product Identification and Appendix Labels Area of The MIO2 is designated for all Fluent and Freedom EVO instruments—except for Freedom EVO 75 that is incompatible with MIO2 with 4 positions. Application Delivery It is recommended that theMIO2 is only delivered and installed by Tecan authorized field service engineers.
  • Page 17 Fuse: T8A 250V Manufacturer’s name and address: Tecan Schweiz AG, CH-8708 Männedorf Barcode Barcode: Serial number and product type Calibration The calibration label (B) states the temperature the MIO2 was calibrated and  3-2. tested for Fig. 3-2, Label ...
  • Page 18: Dimensions And Weights

    Fig. 3-4 Dimensions of the MIO2, 6 positions MIO2 standard MIO2 with shaker option Dimensions Tab. 3-2 Dimensions of the MIO2, 6 positions Dimensions Value Length of standard configuration 213 mm (8.38 in.) Length with shaking assembly / fan option 265 mm (10.43 in.)
  • Page 19 Fig. 3-5 Dimensions of the MIO2, 4 positions MIO2 standard configuration MIO2 with shaker option Dimensions Tab. 3-4 Dimensions of the MIO2, 4 positions Dimensions Value length of standard configuration 228 mm (8.98 in.) length with shaking assembly / fan option 278 mm (10.95 in.)
  • Page 20: Operating Range

    ±1°C at 46°C (max. ±1.8°F at 115°F) the chamber max. ±2°C at 60°C (max. ±3.6°F at 140°F) Warm-up time from max. 30 min. 20°C to 37°C 3 - 6 MIO2 Operating Manual, 394934, en, V3.0...
  • Page 21: Electrical Module Interface

    V 2.0 or later Application Software EVOware V 1.4 SP1 or later V 1.4 SP1 or later Application Software FluentControl V 1.0 or later Setup & Service FluentSetup V 1.0 or later MIO2 Operating Manual, 394934, en, V3.0 3 - 7...
  • Page 22: Environmental Conditions

    -20 to +60°C (-4 to 140°F) Transport humidity 20–80% relative (non condensing) MIO2 must be protected against dust and debris with a cover. Only store MIO2 in its original packaging. 3.8 Emissions Refer to the operating manual of the instrument. Refer to section 1.7 “Reference...
  • Page 23: Principles Of Operation

    4 - Description of Function Principles of Operation 4 Description of Function Purpose of This This chapter explains the basic principle of the MIO2, shows how it is structured and gives a functional description of the assemblies. Chapter 4.1 Principles of Operation The MIO2 contains the following features: ...
  • Page 24: Part Locations

    Slots The MIO2 is provided in a 6 and a 4 slot configuration providing the dispatching of 6 and 4 microplates, respectively, at the same time. Fixing devices prevent the microplates from dislocating.
  • Page 25: Unpacking

    5 - Installation Unpacking 5 Installation Purpose of This This chapter describes how the MIO2 is reinstalled after maintenance task have Chapter been performed. 5.1 Unpacking How to Unpack Before you completely unpack the MIO2, check the following: the Option ...
  • Page 26: Labels On The Packaging

    Ensure that the package will not be exposed to sunlight heat during transport and storage. Protect against strong sunlight. Do not stack Do not stack packages. The package is not designed to carry extra weight. 5 - 2 MIO2 Operating Manual, 394934, en, V3.0...
  • Page 27: Incubator Installation

    5 - Installation Incubator Installation 5.2 Incubator Installation Note: Please note that the MIO2 models shown in various illustrations in this manual do not necessarily correspond to the configuration installed onto the platform. Assuming that the MIO2 support plate for the respective instrument has already been installed proceed as follows to install the incubator.
  • Page 28: Connecting

    Incubator Installation 5.2.1 Connecting D-Sub Cable Note: One or two MIO2 can be connected in series (CAN) and have a device address switch each. Connect the interface cable D-sub connector (A) to the incubator and tighten the connector’s fixing screws.
  • Page 29: Multiple Incubators

     Activate the current limitation mode with the Instrument Software. 6 Positions MIO2 4 Positions MIO2 Fig. 5-3 Multiple MIO2, D-Sub cable and Loop cable connected MIO2 Operating Manual, 394934, en, V3.0 5 - 5...
  • Page 30: Leveling Incubator

    Use the special tool (D) to adjust the three adjustable incubator feet (E) Fig. 5-  5-6. 6 Positions 4 Positions Fig. 5-4 Special tool to adjust feet 5 - 6 MIO2 Operating Manual, 394934, en, V3.0...
  • Page 31 If more than one MIO option connect to the instrument, it is required to select individual CAN bus addresses for each MIO device. Connect the CAN cable of the instrument platform to the MIO2. Put the incubator in the designated position on the worktable.
  • Page 32 5 - Installation Incubator Installation 5 - 8 MIO2 Operating Manual, 394934, en, V3.0...
  • Page 33: Operating Elements

    PC. Depending on your application, refer to the relevant separate documentation. 6.2 Display and Operating Elements There are no display elements attached directly to the MIO2. The corresponding data is displayed on the PC screen. Operating The MIO2 has following operating modes: ...
  • Page 34: Operating The Mio2

    6 - Operation Operating the MIO2 6.3 Operating the MIO2 The MIO2 is operated with an application software and cannot be operated with other interfaces. 6.3.1 Safety Instructions WARNING Contamination risks through contamination of the slot.  Visually inspect all slots for possible spillage of hazardous liquids.
  • Page 35: Switching The Incubator On

    WARNING Injuries or contamination with hazardous materials possible if the maintenance is not carried out properly.  Make sure that the instrument worktable and the MIO2 are free from hazardous materials. Decontaminate them (or have them decontaminated) if necessary. ...
  • Page 36: Incubating

    Heating up The MIO2 is not suitable for warming up of frozen microplates. ATTENTION Risk of error occurrence. The MIO2 is unsuitable for warming up of frozen microplates.  Make sure samples and microplates have ambient temperature before operation.
  • Page 37: Decontamination

    2.6 “Decontamination Declaration”, WARNING Contamination! Substance residues on the MIO2 can cause personal injury and affect the integrity of the process.  Decontaminate the MIO2 and its parts and accessories before any interaction. The decontamination method must be defined by the key operator based on the type of contaminant and degree of contamination.
  • Page 38: General

     Only remove one incubator at the time if multiple incubators are installed to avoid muddling up. Lift and turn the incubator to remove the D-Sub cable connected to the MIO2. Completely remove the incubator. 7 - 2 MIO2 Operating Manual, 394934, en, V3.0...
  • Page 39: Cleaning

    Make sure no foreign objects get caught in the incubator slots. WARNING Fire hazard caused by the improper cleaning. Some cleaning agents are flammable liquids.  Let MIO2 cool down completely before cleaning.  Avoid the formation and accumulation of flammable vapors.  Avoid the spillage of flammable material.
  • Page 40 7 - Maintenance Special Maintenance Tasks 7 - 4 MIO2 Operating Manual, 394934, en, V3.0...
  • Page 41 Which Errors The troubleshooting table below lists possible malfunctions and errors of the MIO2. The operator is enabled to correct some of those problems or errors by can the himself. For this, appropriate corrective measures are listed in the column Operator “Corrective measure”.
  • Page 42: Troubleshooting Table

    Incubator not levelled cor- Level incubator according to  rectly 5.2.3 “Leveling Incubator”, Incubator not properly Remove incubator and do installa- applied to adapter plate. tion procedure 5.2 “Incubator Instal-  lation”, 8 - 2 MIO2 Operating Manual, 394934, en, V3.0...
  • Page 43: Disposal

    Recycling in accordance with applicable legal regulations!  Observe the laws applicable in your country for recycling. For further information refer to the Operating Manual of the instrument. Refer to  1-2. section 1.7 “Reference Documents”, MIO2 Operating Manual, 394934, en, V3.0 9 - 1...
  • Page 44 9 - Disposal 9 - 2 MIO2 Operating Manual, 394934, en, V3.0...
  • Page 45: Contacts

    This chapter informs you how to contact us in case help is needed. It lists addresses and telephone numbers of the manufacturer’s representatives. Chapter How to get Help Tecan and its representatives maintain a fully trained staff of technical specialists around the world. For any technical question, contact the nearest Tecan representative. Feedback on If you have any comments on this Operating Manual or suggestions for improvement, please send them by e-mail to docfeedback@tecan.com.
  • Page 46 +1 919 361 5200 9401 Globe Center Drive, Suite 140, +1 919 361 5201 Morrisville, NC 27560 Phone Toll Free in the US: +1 800 TECAN US or +1 800 832 2687 E-mail helpdesk-us@tecan.com 10 - 2 MIO2 Operating Manual, 394934, en, V3.0...
  • Page 47 10 - Customer Support Contacts Tecan Systems, Inc. Phone +1 408 953 3100 (Tecan Systems) 2450 Zanker Road Toll Free: San Jose, CA 95131 +1 800 231 0711 +1 408 953 3101 E-mail tecan-sy@tecan.com MIO2 Operating Manual, 394934, en, V3.0 10 - 3...
  • Page 48 10 - Customer Support Contacts 10 - 4 MIO2 Operating Manual, 394934, en, V3.0...
  • Page 49: Index

    ..... . . Barcode Temperature Range ..Calibration Label MIO2 Operating Manual, 394934, en, V3.0 11 - 1...
  • Page 50 11 - Index 11 - 2 MIO2 Operating Manual, 394934, en, V3.0...