Operating Manual
Add this manual to your Laetus system file!
Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10

Advertisement

Table of Contents
loading

Summary of Contents for Laetus iBox COSI 221

  • Page 1 English iBox COSI 221 Operating Manual Add this manual to your Laetus system file! Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 2 The information in this manual has been compiled and edited with the utmost care. Laetus reserves the right to make changes to this manual and, in the course of technological progress, to make improvements on the devices without prior notice.
  • Page 3 Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 4 Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 5: Access Rights For The Users Accounts

    -Installation (Setup 1, modification settings. Setup 2, Configuration, allowed, Identification) -Identification view view. -Fault Readings delete. -Reset Counters ADMIN full access full access full access full access **** Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 6: Table Of Contents

    5.2.1 Current Data......... . . 57 Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 7 8. Error Messages and Remedies......68 Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 8 Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 9: Introduction

    PHARMA-CODE on various different products. The iBox-COSI 221 is connected with the evaluation unit (e.g. COMMANDER 12-01) via the Ethernet interface. Fig.1 COSI 221 - Barcode reader Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 10 The device is started using the machine. You do not have to switch on the COSI 221 and the iBox separately. Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 11: Selecting The Ibox-Cosi 221 In The "Laetus Navigator

    Selecting the iBox-COSI 221 in the "Laetus NAVIGATOR" 2. Selecting the iBox-COSI 221 in the "Laetus NAVIGATOR" Press the wt-Devices button in the "Main Menu" of the "Laetus NAVIGATOR". The "Discovery of wt-Devices" menu (in the NAVIGATOR) appears. Fig. 3...
  • Page 12 The Main Menu of the iBox-COSI 221 appears. Fig. 4 Main Menu of the wt-device – iBox-COSI 221 Press the field with the name of COSI 221. Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 13 In the status bar of the main menu is this description also used in the iBox-COSI 221 for the tabular display of a menu containing information on the session number, IP- address and the logged in user. Fig. 5 Active NAVIGATOR-Sessions Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 14: Service - Operation

    Auto logout time (in the "NAVIGATOR") has elapsed. The operating elements become inactive. Back The Back button returns you to the "Main Menu". Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 15: Service" Menu In Detail

    A thin red line is produced on the code level. This positioning aid allows you to adjust the sensor head parallel to the code and to center it on the code. Fig. 7 Switch off the position LED Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 16 Service – Operation If you would like to return the iBox-COSI221 to the Laetus shipping status, click on the Factory Settings button. The following menu appears: Fig. 8 Question to load the factory settings If you select Yes, the factory settings will be loaded.
  • Page 17: Setup - Product Format Parameters

    Select gate polarity Select result pulse width Select reading direction Identification View data: sensor name, sensor type, serial number, batch number, software version, hardware version. Save Save changed data. Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 18 As long as the save is still in progress, you can still interrupt it. To do so, press the Cancel button in the save window. The process is interrupted and the system reverts to the previously used data. Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 19 Close this window by pressing OK. The system only becomes active after you perform this confirmation. If you close the window and if you press the Back button, the following question appears: Fig. 12 Reminder for saving changed data Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 20: Setup" Menu In Detail - Setup1

    The "Setup – Setup 1" menu is opened. Fig. 13 Setup – Setup 1 menu Here you can select and edit the: Evaluation Status Code Type Reference Code Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 21: Evaluation Status - Active / Bypassed

    Use OK to confirm your selection. Use Cancel to discard your selection. If you press the Save button, the changes will be stored and also be applied in future. Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 22: Code Type

    Select this setting, if the code to be controlled corresponds to Mini - PHARMA- CODE. More information on code design is available in the Laetus "PHARMA-CODE Guide". Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 23 Make the appropriate selection and confirm with OK, or close the window without accepting any changes by pressing Cancel. If you press the Save button, the changes will be stored and also be applied in future. Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 24: Reference Code

    If you enter a reference code using the bar keys, you risk making entry errors, if the code is taken from a wrong product. Preferably, you should enter PHARMA-CODE numerically. Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 25: Setup Menu In Detail - Setup 2

    Fig. 18 Setup – Setup 2 menu Here you can select and edit the: Digital Level Sensitivity Result Pulse Output Tube Orientation Signal Orientation Parameter Code Print Decoder Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 26: Digital Level: 0 To 30

    In the "Digital Level" field, enter a numerical value using a keyboard (external or on-screen keyboard) and confirm with ENTER. If you press the Save button, the changes will be stored and also be applied in future. Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 27: Sensitivity - Normal / Medium / High

    To compensate for this, you can increase the sensitivity of the COSI 221 in two steps. If you press the Save button, the changes will be stored and also be applied in future. Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 28: Result Pulse Output - Immediately / End Of Gate

    Make the appropriate selection and confirm by pressing OK, or close the selection menu without making any changes by pressing Cancel. If you press the Save button, the changes will be stored and also be applied in future. Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 29: Tube Orientation Signal - Deactivated / By Length / By Bars / By Time

    If you select the setting "Deactivated", the "orientation parameter" will not be set. The corresponding field remains inactive. Fig. 22 Orientation signal - "Deactivated" Fig. 23 Selecting the tube orientation mode Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 30 Make the appropriate selection and confirm by pressing OK or discard any changes made by pressing Cancel. If you press the Save button, the changes will be stored and also be applied in future. Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 31: Orientation Parameter

    This is the most exact form of orientation. In this case, the number (1...16) of the bar to be used for orientation is entered under "Orientation parameter". Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 32 Setup – Product Format Parameters Orientation signal Length. Fig. 26 Entering the value for the length Orientation signal Time. Fig. 27 Entering the value for the time Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 33: Code Print - Normal / Inverse

    Make the appropriate selection and confirm by pressing OK or close the menu without accepting any changes by pressing Cancel. If you press the Save button, the changes will be stored and also be applied in future. Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 34: Decoder - Forwards / Backwards

    Make the appropriate selection and confirm by pressing OK or close the menu without accepting any changes by pressing Cancel. If you press the Save button, the changes will be stored and also be applied in future. Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 35: Setup - Configuration

    Press the Configuration button. The menu "Setup-Configuration" appears. Fig. 30 Setup – Configuration menu Here you can select and edit the following: Gate Gate Polarity Result Pulse Width Reading Direction Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 36: Gate - External / Internal

    This is where you define the gate type. If you select "External", the gate is created outside of the COSI 221, for example by the machine control. If you select "Internal", the gate is created by the sensor itself (only with Laetus tab guidance).
  • Page 37: Gate-Polarity - Negative / Positive

    Make the appropriate selection and confirm by pressing OK or close the menu without accepting any changes by pressing Cancel. If you press the Save button, the changes will be stored and also be applied in future. Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 38: Result Pulse Width: 1Ms

    The required impulse width depends on the machine control. If you press the Save button, the changes will be stored and also be applied in future. Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 39: Reading Direction - Forwards / Backwards

    Make the appropriate selection and confirm by pressing OK or close the menu without accepting any changes by pressing Cancel. If you press the Save button, the changes will be stored and also be applied in future. Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 40: Setup - Identification

    Software Version Hardware Version No entries can be made in this menu. You have successfully completed the installation process. Press the Back button. The "Main Menu" appears. Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 41: Setting Variable Data / Ibox-Cosi 221 As A Slave Device

    Press the Setup button in the ARGUS wt10 main menu in order to select a product and change its settings. Fig. 36 Main menu ARGUS wt10 Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 42 If the product cannot be loaded for the current grouping, the LED symbol is red. Mark the desired product whose settings you wish to change and Load it. Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 43 You can entering new codes: Fig. 39 Entering a new reference code Enter the reference code via numerical input or using "Enter PHARMA-CODE" and press the button Back. Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 44 If you press No, the "Batch Related Data" menu is displayed and you can select the Save button. The changed product data are logged in the ARGUS wt10. The ARGUS wt10 main menu appears. Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 45: Meaning Of The "Variable" Setting - Activated Or Deactivated

    Fig. 42 Product list in the ARGUS wt10 Mark the product in the list that you would like to edit and then press the Edit Product button. Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 46 Fig. 43 Edit product, first step In order to edit the selected product, press the Connect to button. A connection is then established to the COSI 221. Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 47 With this aid, the sensor head can be positioned parallel to and over the middle of the code. With Next and Previous, you can switch between the tab cards Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 48: Diagnosis

    Fig. 46 Diagnosis menu - Current Data In this menu, you can check the current reference code, the timing diagram and the contrast levels. The counter is reset. Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 49: Factory Settings

    If you press the Back button, the product data will be restored and you can decide whether these settings should continue to be saved. If you would like to return the iBox-COSI221 to the Laetus delivery status, click on the Factory Settings button.
  • Page 50 Setup 1" menu by putting a tick in the appropriate box. As a result, you then only have to enter the reference code when loading a product. Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 51 Once you have entered the reference code, press the Back button. The "Edit Product First Step" menu reappears: Press the Transfer button. The following question appears: Fig. 50 Question about saving a changed product Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 52 Save your selection. The following information appears: Fig. 52 Information about the changed product Press the OK button. In the "Setup Product" menu, Load the changed product. Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 53 You will need to enter a new reference code in order to continue editing the data. Change the reference code and press the Back button. Then save the change. The ARGUS wt10 main menu appears. Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 54 The main menu of the ARGUS wt10 server is displayed after loading. Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 55: Production - Control

    View: gate inactive, gate length, code length, result pulse width. Contrast Level Depict the contrasts of read code bars. Installation View, inspect settings of the installation process. Changes are not possible. Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 56 Inactive buttons have a gray symbol and gray writing. Entry fields are displayed in gray. This means that either the user does not have the corresponding access rights or the current operating mode does not permit this function. Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 57: Production" Menu In Detail

    This menu also appears if you press the Production button in the "Main Menu". Fig. 55 Production – Current Data menu Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 58: Fault Readings

    A 0 indicates: no code in the gate. Fig. 57 Delete Fault Readings Press the Delete the list of fault readings button in order to delete the last fault readings from the memory. Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 59: Timing Diagram

    (The last reading is displayed). The system continues running while displaying the timing diagram, i.e. the machine does not need to be stopped. Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 60 Gate length Beginning of code Beginning of result pulse End of code End of result pulse Code length Result pulse width Fig. 59 Explanation of signal edges Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 61: Contrast Level

    The higher the contrast values, the more reliable the code evaluation. Note Contrasts under 25 % cannot be read anymore. Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 62: Installation

    Identification" you can view, inspect settings of the setup process. No entries can be made in these menus. You have completed "Production – Control". The Back button returns you to the "Main Menu". Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 63: Administration

    Auto logout time (in the "NAVIGATOR") has elapsed. The operating elements become inactive. Back The Back button returns you to the "Main Menu". Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 64: Administration" Menu In Detail

    Place the cursor into the "Equipment Name" field and enter the name using a keyboard (external or on-screen keyboard). Confirm your entry by pressing ENTER. No other entries can be made in this menu. Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 65 Here, you can select the user language for the selected user account. Press the "..." button. The selection menu for the user language appears. Select the desired user language. Fig. 64 Selecting the user language Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 66 Fig. 65 Question on saving changed data Yes accepts the changes. You have completed the "Administration" process. Press the Back button in order to return to the "Main Menu". Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 67: Sample Configuration

    If the tube cannot be turned any further although orientation has not been completed, this tube is regarded as FAULT. Filling is suppressed and the tube is ejected. Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...
  • Page 68: Error Messages And Remedies

    The background color of the title bar is red. Remedy: Check the cable connection between the sensor (or iBox) and the evaluation unit. Check the power supply cable between the iBox and the sensor. Laetus web technology | 13 9 66 0802 | FirmwareVersion 1.1.x/2.0.x | 2009-10...