INSTRUCTIONS
FOR USE
Please read these Instructions For Use
carefully before use, and file for future
reference.
A-dec 301
MADE IN JAPAN
Air Motor
OM-DM0906EN
000

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the A-dec 301 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Atdec A-dec 301

  • Page 1 Air Motor A-dec 301 INSTRUCTIONS FOR USE Please read these Instructions For Use carefully before use, and file for future MADE IN JAPAN OM-DM0906EN reference.
  • Page 2 User and Indications for Use Precautions for Handling and Operation User: Qualified Professionals (Dentist/Dental hygienist) • Please read these precautions carefully and use only as Indications for Use: intended or instructed. Air Motor is intended for the following application(s): • Safety instructions are intended to avoid potential hazards that A connection equipment to transfer energy (rotation, air, water, could result in personal injury or damage to the device.
  • Page 3 CAUTION General product specification information NOTICE highlighted to avoid product malfunction and • Read this Instructions For Use before use to fully understand performance reduction. the product functions and file for future reference. • When operating the product, always consider the safety of the WARNING patient.
  • Page 4 CAUTION CAUTION • Should the motor function abnormally during use, stop using • U.S. Federal law restricts this device to sale by or on the order the motor immediately and contact your Authorized A-dec of a licensed physician. Dealer. NOTICE •...
  • Page 5: Preparation For Use

    Preparation for Use CAUTION • Do not use air contaminated by dust, moisture or oil. Setting of Air & Water Supply Pressure Connecting and Disconnecting the Motor Measure the supply pressure Chip Air Pressure Exhaust Air at the motor/hose connection Pressure Water 3-2-1 Connecting...
  • Page 6 Connecting and Disconnecting the Handpiece from the Motor 3-2-2 Disconnecting Hose Nut Loosen the hose nut and Connecting remove the motor from the hose (Fig. 4). 1) Insert the handpiece directly into the motor (Fig. Motor Fig. 4 2) Confirm that the handpiece is firmly Handpiece Motor...
  • Page 7 Forward / Reverse Rotation Disconnecting Hold the motor and the handpiece, then pull apart. (Fig. 5) 1) Forward Rotation: Turn the “F”: Forward Rotation FWD/REV “R”: Reverse Rotation FWD/REV Change Ring to “F”. CAUTION Change Ring (Fig. 6) • Do not connect or disconnect the handpiece until the motor 2) Reverse Rotation: Turn the has completely stopped.
  • Page 8: Post-Use Maintenance

    Check Before Treatment Post-use Maintenance Follow the check below before use. If any abnormalities are After each patient, maintain the product as follows. found, stop using the handpiece immediately and contact your WARNING Authorized A-dec Dealer. • To ensure cleaning and sterilization efficacy, use 1) Connect the handpiece and make sure that it is securely only the following procedures for reprocessing.
  • Page 9 2) Remove the bur. (Fig. 10) Cleaning at point-of use CAUTION • Do not use the following liquids to wipe, immerse, or clean the product: strong/super acid water, strong acid/alkaline chemicals, chlorine-containing solutions, solvents such as Fig. 10 benzine or thinner. •...
  • Page 10 Cleaning and Drying Use a state-sanctioned disinfectant with proven bactericidal, fungicidal, and virucidal properties. Perform cleaning and drying using the manual or automatic The following disinfectants can be used in the United method. States and Canada. Manual Method (Cleaning and Drying the Exterior) CaviCide , CaviWipes (manufactured by Metrex)
  • Page 11 When using disinfectant, follow the instructions given by the 3) Remove any moisture from the motor with compressed air manufacturer of the disinfectant. and wipe off any external moisture using a dry lint free Use a state-sanctioned disinfectant with proven bactericidal, cloth.
  • Page 12 Lubrication ■PANA SPRAY Plus (Lubricating the Interior) CAUTION 1) Attach a tip nozzle* to the Exhaust air port • Hold the spray can upright. nozzle of the PANA SPRAY • Firmly hold the motor to prevent it slipping when spray Drive air Plus.
  • Page 13 Packaging, Sterilizing, and Drying CAUTION • Use an FDA-approved steam sterilizer to perform 1) Insert the motor into an FDA-approved sterilization pouch sterilization. that conforms to ISO 11607-1 (EN ISO 11607-1), and seal the pouch. • Follow local rules, regulations, and guidelines regarding the reprocessing of devices.
  • Page 14 CAUTION Storage • Be sure to use sterilizers that can perform sterilization no Store the product in a dry, clean location. higher than 275°F (135°C). In some sterilizers, the CAUTION chamber temperature may exceed 275°F (135°C). Do not • After the sterilization and drying cycles are complete, use these sterilizers as failure of the motor could occur.
  • Page 15: Maintenance

    Maintenance Periodical Maintenance Checks Perform periodical maintenance checks every three months, If a deteriorated O-ring makes O-ring referring to the checklist below. If any abnormalities are found, it difficult to attach the contact your Authorized A-dec Dealer. handpiece, replace the O-ring. 54.0804.00 Using a needle or other pointed Points to check...
  • Page 16: Specifications

    Specifications Model Hose Connection Type ISO 9168 (EN ISO 9168) Type 2 (Midwest 4 hole) Handpiece Connection Type E Type ISO 3964 (EN ISO 3964) Rotation Speed 19,800 - 24,200rpm (36psi (0.25MPa)) Breakdown Torque ≥1.5N·cm (36psi (0.25MPa)) Drive Air Pressure 29 - 36psi (0.20 - 0.25MPa) Air Consumption <66NL/min (36psi (0.25MPa))
  • Page 17: Warranty

    Warranty Symbol IMPORTANT For warranty, electrical specifications, State of being fixed. identification of symbols, and other regulatory requirements, see the Regulatory Information, Specifications, and Warranty document (p/n 86.0221.00), which is available in the Resource Center at www.a-dec.com.
  • Page 18: Disposing Product

    Disposing product To avoid the health risks of operators handling the disposal of medical equipment, as well as the risks of environmental contamination caused thereof, a surgeon or a dentist is required to confirm the equipment is sterile. Ask specialist firms who are licensed to dispose of specially controlled industrial wastes, to dispose the product for you.
  • Page 20 Manufactured for A-dec 2601 Crestview Drive Newberg, Oregon 97132 United States Tel: 1.800.547.1883 within USA/CAN Tel: +1.503.538.7478 outside USA/CAN www.a-dec.com Specifications are subject to change without notice. Copyright 2021 A-dec, inc. All rights reserved. Issued Date: 2021...