AccuBioTech SUREPETTE SPA-10 User Manual

AccuBioTech SUREPETTE SPA-10 User Manual

Adjustable volume pipette, fixed volume pipette

Advertisement

Quick Links

SUREPETTE
®
USER'S GUIDE
Adjustable Volume Pipette
Fixed Volume Pipette
..._....,
AccuBioTech Co., Ltd.
•·
Accurate, Reliable, Cast Effective

Advertisement

Table of Contents
loading

Summary of Contents for AccuBioTech SUREPETTE SPA-10

  • Page 1 SUREPETTE ® USER'S GUIDE Adjustable Volume Pipette Fixed Volume Pipette A® ..._.., AccuBioTech Co., Ltd. •· Accurate, Reliable, Cast Effective...
  • Page 3: Table Of Contents

    CONTENTS PAGE 1. INTRODUCTION 2. PARTS CHECK LIST 3. DESCRIPTION 4. OPERATING RANGES 5. SETTING THE VOLUME 6. PIPETTING 7. GUIDELINES FOR GOOD PIPETTING 8. RECALIBRATION 9. MAINTENANCE 10. TROUBLE SHOOTING 11. STORAGE 12. SPECIFICATIONS...
  • Page 4: Introduction

    1. INTRODUCTION Thanks for your trust in our SUREPETTE ® Series Pipettes. Let's share the following benefits from SUREPETTE ® 1 ). SUREPETTE ® Series Pipettes include Adjust­ able Volume Pipettes and Fixed Volume Pipettes, covering a volume range from 0.5 µ L to 5 ml;...
  • Page 5: Description

    3. DESCRIPTION - - - - - - - - - - - Push Button - - - - - - - - - - - - - P ush Button Rod - - - - - - - - - - Tip E j e c tor Button - - - - - - - - - - Handl e '°...
  • Page 6: Setting The Volume

    SPA-1000 100- 1000 µ L - 5 ML SPA-5000 Fixed Volume Pipette Model Volume SPF-5 µL SPF-10 µL SPF-20 µL SPF-25 µL SPF-50 µL SPF-100 100 µL SPF-200 200 µL SPF-250 250 µL SPF-500 500 µL SPF-1000 1000 µL SPF-2000 2000 µL SPF-5000 5000 µL...
  • Page 7: Pipetting

    Example for each pipette: SPA-10 (Adjustable Unit Volume 0.1 µ L) 1 1 l 1 ° 1 l 1 °- 1 ° 1 0.5µL 1µL 10µL SPA-20 (Adjustable Unit Volume 0.5 µ L) 1 °- 1 ° 1 l 2 · 2.5µL 10µL 10.5µL...
  • Page 8 2). Pre-rinse the tip by aspirating the first volume of liquid and then dispensing it back or to waste; 3). Press the push button to the first stop. This part of the strake is the volume displayed on the volumeter; 4 ).
  • Page 9: Guidelines For Good Pipetting

    ill l � j � j � j ' , 1 ' , 1 ' , 1 ' , J ' , 1 ' � ' � '� ' , 1 ' , I � � 7. GUIDELINES FOR GOOD PIPETTING 1 ).
  • Page 10: Recalibration

    6). Liquid should never enter the tip holder; to prevent this: - press and release the push button slowly and smoothly; - never turn the pipette upside down; - never lay the pipette on its side when there is liquid in the tip. 7).
  • Page 11: Maintenance

    3). Pipette the distilled water into the beaker fol­ lowing the pipetting procedures and measure with an analytical balance. Repeat the mea­ surement six times. The results should be con­ sistent with the specifications. Otherwise, cali­ brate and test again until the results are consis­ tent with the specifications.
  • Page 12: Trouble Shooting

    5). Wipe the piston, the 0-ring and the tip holder with ethanol and a lint-free cloth. Note: Same models have a fixed 0-ring lo­ cated inside the tip holder. In this case, the 0- ring can not be accessed for maintenance. 6).
  • Page 13 for repair. lf the answer to these questions is YES, continue as follows: - 1s the tip ejector straight? - 1s the tip ejector free of chemical damage? - Remove tip ejector; is the tip holder undamaged and free of chemical damage? - Are the 0-ring and seal undamaged and free of chemical damage? lf the answer is NO, clean or replace the damaged...
  • Page 14: Storage

    11. STORAGE For convenience and safety, always keep your SurePette vertically on its own holder provided when not in use. When installing the holder, please follow the below instructions: 1 ). Clean the shelf surface with ethanol; 2). Remove the protective paper from the adhe­ sive tape of the holder;...
  • Page 15 Adjustable Volume Pipettes Normal Accuracy Precision Model & Volume (mean error) (repeatability) Volume Rang SPA- 10 ± 12 ,;;;;4_0 0.5~10 ±6.4 .;;; 3.0 µI ±3.2 .;;; 2.0 SPA -20 ± 10 .;;; 3.0 2~20 ±3.2 .;;; 2.0 µI ±3.0 .;;; 1.5 SPA- 50 ±6.4 .;;;...
  • Page 16 A• AccuBioTech Co., Ltd. •• Accurate, Reliable, Cost Effective...

Table of Contents