Advertisement

Quick Links

Biosense Inc.
2616 Peartree Lane
San Jose, CA 95121
800-941-2844
RECIRCULATOR II (Part No. 9431)
User and Service Manual
th
4
Edition, Revised 3/20

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the RECIRCULATOR II and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Biosense RECIRCULATOR II

  • Page 1 Biosense Inc. 2616 Peartree Lane San Jose, CA 95121 800-941-2844 RECIRCULATOR II (Part No. 9431) User and Service Manual Edition, Revised 3/20...
  • Page 2 Recirculator II User and Service Manual...
  • Page 3: Table Of Contents

    Recirculator II User and Service Manual Table of Contents Introduction to the Recirculator II ....................3 Unpacking the Unit ........................3 Testing the Unit ........................... 3 Using the Recirculator II ......................4 Dual Speed Operation ........................4 Negative Pressure Adapter (Airflow Diverter) ................4 Maintaining the Recirculator II ....................
  • Page 4: Introduction To The Recirculator Ii

    Recirculator II User and Service Manual Introduction to the Recirculator II The Recirculator II is an air purifying device designed to trap and retain airborne particulates. Using the latest HEPA (High Efficiency Pariculate Air) filter technology, this unit is able to remove 99.97% of all particulate nuclei which exceed 0.3 microns in size.
  • Page 5: Using The Recirculator Ii

    Replacement” on page 3). Dual Speed Operation The Recirculator II is equipped with a two speed blower controlled by a dual position switch (High-Off-Low). At the High speed (upper position), the unit will purify about 575 cfm with new, clean filters installed. At the Low speed (lower position), the unit will purify about 475 cfm with new, clean filters installed.
  • Page 6: Servicing The Motor

    (Refer to diagram at the end of this manual for more information.) To service the motor: 1. Remove the metal grate on the bottom of the Recirculator II and oil the motor bearings. Consult the attached exploded view of the motor that illustrates the points at which to apply the oil.
  • Page 7: Changing The Filters

    The standard filters used in the Recirculator II are an extruded aluminum frame, enhanced performance HEPA filter designed to provide approximately 575 cfm when new, at a pressure differential of about 0.60-0.65 inches w.c.
  • Page 8: Hepa Filter Replacement

    0.3 microns whenever you open the top of the Recirculator II. 3. Open the Recirculator II only in an area with a negative pressure relationship to adjacent areas, so that contamination of such adjacent areas cannot occur as a result of these procedures.
  • Page 9: Pre-Filter Replacement

    Velometer or Balometer. Cleaning the Recirculator II The Recirculator II has a tough baked enamel finish and can be cleaned with most germicidal cleaners. Avoid abrasive or petroleum cleaners, as these can damage the finish. Over time, and in the normal course of use, some dust may collect inside the unit and possibly on the performance indicator.
  • Page 10: Appendix A: Negative Pressure Adapter (Part No. 9405)

    The NPA 270 is an adapter, which is designed to be fitted onto the grillwork face of the Recirculator series manufactured by Biosense, Inc. It consists of a rectangular chassis with an 8" diameter outlet sleeve to which exhaust ducting can be connected.
  • Page 11: Using The Npa 270

    It should be noted that once fitted with the NPA 270 connected to ductwork exhausting air out of the area, the Recirculator II will still return approximately 260 cfm into the room in which it is located. Although this will still provide about 9-10 air changes per hour (ACH) in a moderately sized room, it may be necessary to install a second Recirculator unit for local air treatment to maintain any required higher ACH rates.
  • Page 12: Product Service

    Prompt Disposition Biosense will make a good faith effort for prompt correction or other adjustment with respect to any product which proves to be defective within limited warranty. For any product believed to be defective within limited warranty, first write or call the dealer from who the product was purchased.