7823 Oxygen Analyser
FEATURES
High accuracy linear output
Configurable outputs:
4-20mA and 0-10
interface
Selectable output measurement ranges:
Standard ranges of 0-25% O
0-100% O
or fully adjustable via RS232 when
2
configured in 0-100% O
Externally triggered automatic or manual
calibration
Can be calibrated in normal air (20.7% O
any other known O
Cycling 3.3
V
monitoring of the O
diagnostic purposes
Selectable output filtering allows adaptive, fast
and dynamic or slow and stable output
SPECIFICATIONS
Maximum ratings
Supply voltage
Current consumption
Load
perature limits (electronics enclosure)
Storage
Operating
Temperature limits
(permissible gas temperature at sensor tip)
Operating
(Standard Temp)
(High Temp)
Gas flow rate
Weight
Incidental permissible acceleration
ive permissible acceleration
Sealing Rating
or RS232 comms
V
DC
and
2
mode
2
concentration
2
logic output allows direct
DC
sensor pump cycle for
2
24V
500mA max
@ 24V
DC
100 to 600Ω Tem-
-10 to 85C
-10 to 85C
-100 to 250C
-100 to 400C
0 to 10 m/s
<450g
30g Repetit-
5g
IP65
ECOline 7823 Oxygen Analyser
) or in
2
APPLICATIONS
Combustion control including oil, gas and
± 10%
DC
biomass boiler applications
Composting
4-20mA
Laboratory & building air quality monitoring
including confined space personnel safety
Industrial process control i.e. gas mixing for
welding and steel making
Oxygen generation systems
Medical
Scientific including respiratory studies of a
community or an organism, plants and
animals
Food and beverage packaging
Applications where low oxygen is key
including fermentation, rust and corrosion
prevention, inerting and purging
Page 1 of 12

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 7823 and is the answer not in the manual?

Questions and answers

Summary of Contents for ECOLINE 7823

  • Page 1 ECOline 7823 Oxygen Analyser E instruments 7823 Oxygen Analyser FEATURES  High accuracy linear output  Configurable outputs: 4-20mA and 0-10 or RS232 comms interface  Selectable output measurement ranges: Standard ranges of 0-25% O 0-100% O or fully adjustable via RS232 when...
  • Page 2 In order concentration in air or inert gas mixtures with temperatures to output an oxygen concentration (%) the 7823 must be of -100 to +400°C max. calibrated, or more specifically, re-referenced in a known gas concentration, typically normal air.
  • Page 3: Performance Characteristics

    ECOline 7823 Oxygen Analyser E instruments 7823 Oxygen Analyser PERFORMANCE CHARACTERISTICS Characteristics Min. Typ. Max. Unit Output inactive start up delay (heater warm up) Initial warm up time (till stable output) Measuring ranges 25% Configuration 100% Configuration (2) (3) Accuracy After Calibration...
  • Page 4 ECOline 7823 Oxygen Analyser E instruments 7823 Oxygen Analyser OUTLINE DRAWING AND MOUNTING INFORMATION Mounting holes accessible on removal of housing lid PCB LAYOUT (REMOVE HOUSING LID TO VIEW) Page 4 of 12...
  • Page 5: Electrical Connections

    ECOline 7823 Oxygen Analyser E instruments 7823 Oxygen Analyser ELECTRICAL CONNECTIONS Description ± 10% Calibrate Cycle 4-20mA/RS232 Tx 0-10V /RS232 Rx CENTRE Housing/Probe Earth Housing Connector: Amphenol Ecomate C016 30C006 100 12 Mating Connector: Binder 99-4218-00-07 (Mating Connector supplied with each product) Output pins 5 and 6 are both referenced to the supply GND (pin 2).
  • Page 6: Rs232 Communication Settings

    RS232 COMMUNICATION SETTINGS When connecting the 7823 via the RS232 connections ensure Tx goes to Rx of the PC and Rx goes to Tx of the The 7823 communicates via standard COM port settings that are default on most PCs and many other RS232 compatible devices.
  • Page 7 Continuous Data Streaming  On power up, after the initial 60s heater delay, the 7823 Power down the 7823 will automatically begin outputting the measured O  Remove the lid using a Philips screwdriver concentration and sensor Td as both an averaged and raw ...
  • Page 8 Analogue Outputs (4-20mA and 0-10V 20.7% O to allow simple calibration in normal air. The The 7823 is factory default to output a range of auto calibration value is factory set to 20.7% to take into 0-25% O via its two analogue outputs. This range can be account average humidity in the atmosphere.
  • Page 9: Calibration Procedures

    7823 Oxygen Analyser CALIBRATION PROCEDURES ERROR CONDITIONS Automatic Calibration If the oxygen sensor is damaged the 7823 will  Ensure the 7823 is configured for automatic highlight this by blinking the CYCLE output (pin 4) and calibration. See CONFIGURATION on Page 7.
  • Page 10 SENSOR OPERATING TIPS To get the best performance from the 7823 analyser it is important that the oxygen probe is installed and maintained in the correct manner. The following two pages outline some useful sensor operating tips and a list of gases and materials that must be avoided to ensure a long sensor life.
  • Page 11 ECOline 7823 Oxygen Analyser E instruments 7823 Oxygen Analyser SENSOR OPERATING TIPS continued Cross sensitivity with other gases: Gases or chemicals that will have an influence on the life of the sensor or on the measuring results are: Combustible Gases Small amounts of combustible gases will be burned at the hot Pt-electrode surfaces or Al filters of the sensor.
  • Page 12: Ordering Information

    ECOline 7823 Oxygen Analyser E instruments 7823 Oxygen Analyser ORDERING INFORMATION Part number Output Type Probe Length Measuring Range Gas Temperature 4 to 20mA and 0 to 10V or 0.1 to 25% Vol. O 7823 220mm -100 to 250C RS232 0.1 to 100% Vol.