3-1-2 TM-2430 Series Error Codes
ERROR CODE
E00
E03
E04
E05
E06
E07
E08, E10
E20
E21
E22
E23
E30, E31
E32
E50
E52
E53
E55, E56, E57
E60
E70, 71, 72, 73
E74
E75
POSSIBLE CAUSE
Clock parameters have not been
set.
An
error
code
is
without cuff inflation.
Measurement is stopped due to a
low battery.
Inflation pressure does not reach
the target pressure.
Pressurization greater than 320
mmHg.
Controlled stop using STOP key
Pulse can not be detected.
Pulse less than 35 beats or more
than 200 beats per minute.
Diastolic greater than 160 mmHg
or less than 40 mmHg.
Systolic greater than 280 mmHg
or less than 60 mmHg.
DSD greater than 150 mmHg or
less than 10 mmHg.
(DSD = difference between
systolic blood pressure and
diastolic blood pressure)
Air leaks in the cuff assembly.
Clock error
Pressure offset error
Memory error
Defective battery contact
Exhaust error
Interval setting error
RS-232C error
Voltage
decreases
communication.
Protocol error due to external
equipment
CORRECTION
Enter clock parameters.
displayed
Exhaust the air from the cuff completely.
Replace with new batteries. Auto mode
is quit. Restart the Auto mode if you use
it.
Wrap the cuff and connect to the
monitor. If the error persists, there may
be an air leak and repair is necessary.
Do not move and try to relax during the
measurement. If you can not clear the
error, repair is necessary.
Do not press the STOP key when
unnecessary.
Do not move and try to relax during the
measurement.
Measure the blood pressure by other
methods.
Measure the blood pressure by other
methods.
Measure the blood pressure by other
methods.
Measure the blood pressure by other
methods.
Repair is necessary.
If the error persists, repair is necessary.
Repair is necessary.
Repair is necessary.
Replace batteries correctly. If the error
persists, repair is necessary.
Relax and do not move during the
measurement.
frequently, repair is necessary.
Enter the interval parameters correctly.
Re-connect the communication cable. If
the error persists, repair is necessary.
during
Replace batteries with new ones and
restart communication.
Re-connect the communication cable. If
the error persists, repair is necessary.
50
If
this
error
occurs
Need help?