Digitimer NeuroLog NL405 Width/Delay User Manual

Advertisement

Users Manual

Introduction

The NL405 WIDTH/DELAY is a digitally controlled pulse duration generator and is used
to produce precise pulses of adjustable duration with high accuracy and repeatability.
The NL405 produces TTL logic output pulses (of both polarities) that have a duration set
using three decades of thumbwheels and a seven (7) position range switch. It also
features a press to start button, LED indication of an output and a separate logic output
signal at the end of the defined WIDTH time. It was designed as a replacement for the
NL401 and in most places where the NL403 would have been used.
The NL405 can be triggered off either positive or negative-going edges of the input pulses
and the NL405 is capable of 100% duty cycles. An external push to make switch can be
used when -ve edge triggering is selected.
In the "FIXED WIDTH" mode, the NL405 produces an Output pulse as specified by the
front panel controls; In the "RE-TRIGGER" mode, any valid Input trigger arriving whilst the
Output is active will restart the timing and therefore extending the Output pulse duration.
As long time periods can be generated (999 seconds is over 16 minutes) any change in
position of the rotary range switch will stop the current Output and inhibit further input
triggers for 450 ms.
The "DELAYED" Output socket provides a short duration pulse (set internally) that is
produced at the end of the timed WIDTH pulse. This provides a trigger-out after the set
Delay time.
The NL405 can be used for digital control of pulse duration or for introduction of a delay to
a trigger pulse. The logarithmic or exponential nature of the many biological responses
make the very wide continuous adjustment (from µs to minutes) particularly beneficial.

Specification Summary

Input:
Trigger edge
Manual
Output (Width)
Polarity
Range
Control
Multiplier
Accuracy
Delay from Trigger In
Indicator
Output (Delay)
Polarity
Duration
Output
Mode
Reset/Inhibit
Rear connections to the motherboard allow Input and Output interconnections between this and other logic
modules without the need of front panel cables.
Digitimer Ltd
E-mail: sales@digitimer.com
NeuroLog™ System
NL405 - Width/Delay
:
TTL pulse. Input is 1 TTL load.
:
Positive or Negative, selected from front panel
:
Push-button, operation on push.
:
TTL. Output will drive 10 loads.
:
Both active-high and active-low sockets
:
1 µs to 999 s
:
001 to 999
:
1 µs, 10 µs, 100 µs, 1 ms, 10 ms, 100 ms, 1 s.
:
±0.5%
:
<200 ns
:
Amber LED on for approximately 100 ms for each pulse
:
TTL. Output will drive 10 loads.
:
Active-high
:
0.5 µs, 1.5 µs, 5 µs
50 µs, 150 µs, 500 µs
:
Fixed Width or Re-Triggered, selected from front panel
:
450 ms ±20%, after change of rotary range control.
Page 1 of 5
Tel:+44 (0)1707 328347; Fax:...373153
±20%
±10%, by on-board jumpers.
Website: www.digitimer.com
NL405
Copyright © 2003

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the NeuroLog NL405 Width/Delay and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Digitimer NeuroLog NL405 Width/Delay

  • Page 1 450 ms ±20%, after change of rotary range control. Rear connections to the motherboard allow Input and Output interconnections between this and other logic modules without the need of front panel cables. Digitimer Ltd Page 1 of 5 Copyright © 2003 E-mail: sales@digitimer.com...
  • Page 2 ‘click’ position and then back to where it was set. The Outputs are inhibited for 450 ms after the switch becomes still. Note that a Delayed Output pulse is NOT generated at the end of a WIDTH that is terminated early. Digitimer Ltd Page 2 of 5 Copyright © 2003 E-mail: sales@digitimer.com...
  • Page 3 This may need to be found from trial and error but a value of 1µF would be a good starting point. Digitimer Ltd Page 3 of 5 Copyright ©...
  • Page 4 OUTPUT - the lower jumper selects which of the three Output signals is passed to the Input of the module to the immediate right, or it may be "parked" for "no-connection (N/C)". Digitimer Ltd Page 4 of 5 Copyright © 2003 E-mail: sales@digitimer.com...
  • Page 5 We reserve the right to alter specifications and price without prior notification. First Issued: December 19, 2002 Last Revision: March 26, 2003 Printed: March 26, 2003 File Reference: N:\Docs\Company\Manuals\NeuroLog \ NL405-02.lwp Digitimer Ltd Page 5 of 5 Copyright © 2003 E-mail: sales@digitimer.com Tel:+44 (0)1707 328347; Fax:...373153...