This Technical Brief provides instructions for using the
Robotic Arm Alignment Utility and Robotic Arm Cycle Test.
®
In Virtua
Version 3.1.1 and later software, these tools are
available on the Utilities screen in the Robotic Arm tab.
léÉê~íáçå~ä=aÉëÅêáéíáçå
The Robotic Arm Utility is used to help assure that the
Virtua Medical Disc Publisher's robotic arm is properly
calibrated for regular transport of discs from the optical disc
bin(s) to the optical drive(s) to the print tray and then the
output bin. It is strongly recommended that the Robotic Arm
Cycle Test (see page 7) be run upon initial Virtua system
installation or after major system repair/transport. If it does
not complete successfully, the calibration process should be
completed.
NOTE: This Utility should only be used by those
personnel familiar in its use, since mis-calibration could
cause poor performance or damage to the system. The
Utility will only be available to users with the Admin
role. Contact Codonics Technical Support for additional
assistance.
mêÉêÉèìáëáíÉë
The following items are required to complete these service
instructions:
T10 Torx driver (torque driver recommended)
Felt-tip pen
Up to 50 optical discs
3.1.1 software or later is installed on Virtua.
NOTE: This utility cannot be used on Virtua ECO
systems that run software versions 3.1.1 through 4.0.1.
This utility will not run on Virtua-1 systems.
NOTE: Check the trays of the optical disc drives for 4
clear Teflon stickers on the drive tray where the optical
disc would be placed. If they are not present, contact
Codonics Technical Support.
`~äáÄê~íáåÖ=íÜÉ=oçÄçíáÅ=^êã
1. Wait for Virtua to finish processing jobs.
2. Open the top cover and place a folded piece of paper or
card stock in the optical sensor as shown in the following
image. This will allow access to the drives and print tray
while the cover is open.
CAUTION: The robotic arm will move with the top
cover sensor disabled. Keep hands, hair, and loose
clothing clear of robotic arm when in motion.
3. Go to the Utilities tab and then the Robotic Arm tab.
4. Press the Start button in the Robotic Arm Alignment
section to enter calibration mode. Jobs subsequently sent
to Virtua will accumulate in the background and will be
processed after calibration mode is exited.
qÉÅÜåáÅ~ä=_êáÉÑ
®

Advertisement

loading
Need help?

Need help?

Do you have a question about the Virtua Robotic Arm Utility and is the answer not in the manual?

Questions and answers

Summary of Contents for Codonics Virtua Robotic Arm Utility

  • Page 1 The Utility will only be available to users with the Admin role. Contact Codonics Technical Support for additional 3. Go to the Utilities tab and then the Robotic Arm tab. assistance.
  • Page 2 NOTE: Number values within the Robotic Arm Utility Pick. described here are different than for the external Recorder Calibration Utility (reference Codonics Part 2. Press the right arrow button or left arrow button until No. 901-218-001). Bottom Drive Tray appears as the Component.
  • Page 3 3. Press the Drop button to have the robotic arm place the 8. Remove all optical discs and input bins. Shut down power to the Virtua system. Disconnect the USB cable optical disc in the bottom optical drive tray. and eSATA cables from the Recorder and switch off the 4.
  • Page 4 Please contact 2. Press the Move button to have the robotic arm hold the Codonics Technical Support for further instruction. optical disc at the top of the bin. 3. If the optical disc position is too far left or right, adjust the value to move the robotic arm right and left.
  • Page 5 -20 is not recommended. Press the Pick button. Watch the robotic arm body and Contact Codonics Technical Support if a lower value is note how much it is lifting up on the top guide rail. The required.
  • Page 6 2. Select the Stack Height component. 7. Repeat steps 2 through 5 substituting “Top Drive Height” for “Bottom Drive Height” and “Print Tray Height”. NOTE: If the unit has been incorrectly calibrated without a disc in the tray, it can no longer be calibrated from the User Interface.
  • Page 7 Specifications subject to change without notice. Patent 7,375,737 and +1.440.243.1198 Changning Dist., Shanghai Minato-ku, Tokyo, others pending. +1.440.243.1334 Fax P.R. China, 200051 108-0023 JAPAN Email info@codonics.com 86-21-62787701 81-3-5730-2297 www.codonics.com 86-21-62787719 Fax 81-3-5730-2295 Fax Copyright © 2017 by Codonics, Inc. Printed in the U.S.A. Part No. 901-222-003.02...