CDI DIGITAL TORQUE TESTER DTT Operation Manual page 7

Table of Contents

Advertisement

sETUP
MOUNTING
Mount the DTT to a sturdy support before use. Torque
testing procedures may involve hundreds of pounds of
load. Be sure to consider counterweight of the mount
to accommodate torque testing operations and testing
fixtures. Allow enough room around and below the
mount for the wrenches or testing fixtures being used.
Vertical mounting on a wall or heavy pillar is preferred
for most wrench and tool testing as shown in the
illustration below. Use four 1/4" diameter bolts
(torqued to 10 ft.lb.) to insure a secure mount. The DTT
can also be mounted vertically to a horizontal surface
using a right angle bracket.
8.87"
Figure 3: Mounting, Dimensions and Right Angle Bracket
A - Mounting Screw (4)
B
B- Right Angle Bracket
0.00
D
8.87"
A
C
FT-LB
1.
Turn the power switch ON.
C
2.
In the MAIN MENU screen, select DTT SETTINGS.
3.
In the SYSTEM CONFIGURATION screen, the display
reads:
DATE:
TIME:
BRIGHTNESS:
4.
To increment or decrement the DATE/TIME/
BRIGHTNESS elements, select the element and
repeatedly press the UP or DOWN arrows.
NOTE: Press and hold the UP or DOWN arrows to
change the display quickly. The scrolling DATE/
TIME elements roll over at both ends for quick
and convenient adjustment of settings.
5.
Press NEXT to update DATE/TIME/BRIGHTNESS and
return to the MAIN MENU.
B
SW VERSION: 1.X.X
DATE:
TIME:
BRIGHTNESS:
TGT TORQUE
SERIAL NUM
OPERATOR ID :
TEsTER sETUP
MONTH / DAY / YEAR
HOUR/MINS
30-100
SYSTEM CONFIGURATION
Nov
03
2009
HR
11
07
60
NEXT
Figure 4: System Configuration screen
DIGITAL TORQUE TESTER
TEST SETUP
:
250 FT-LB
:
110900MAW
TECH512
NEXT
NEXT
A
D
CHECKSUM: A1XX
MIN
CHANGE
CHANGE
CHANGE

Advertisement

Table of Contents
loading

Table of Contents