MOBILE DECONTAMINATION CABINET
OPERATING and MAINTENANCE INSTRUCTIONS
Model no.
CGMH73A

HOW INSTALL UNIT:

1. Place cabinet on level floor, in well ventilated area.
2. Unpack and install parts for hangers or racks.
3. Plug cord into wall outlet.

HOW TO RUN THE DECONTAMINATION CYCLE:

1. Read all instructions before operating the Decontamination Cabinet.
2. Turn on the device by pressing the ON-OFF Switch. The switch light will come on.
3. The unit starts heating to the previous cycle settings. This is standby mode. The cycle light will be off.
The Temp Control will display "LOTPr" until 140°F is reached. Then the setpoint will be displayed.
4. To begin the decontamination cycle, press the "CYCLE" button. The display will show cycle "rcP01"
and the "CYCLE LIGHT" will be flashing. Cycle "rcP01" setpoints are 185°F at 65% humidity for
15 minutes.
5. Press the DIAL to set the cycle.
6. The unit will start preheating and the Temp Control will display "PrHEt" until the cabinet temperature setpoint is
reached. Then "PUSH StArt" will be displayed. Do not press the start button at this time.
7. The cycle conditions are preset by the factory and are locked to prevent modification and ensure repeatability.
8. Do not run the unit if the low water indicator light is ON. The light will turn off once water is refilled.
9. Use the "DISPLAY" button to check for actual temperature and humidity as described under the
"Digital Control Notes" section on page 2.
Volts
Watts
120
2000
Phone: 440-350-1100 Ext: 268 • www.crescor-hs.com
Amps
Hertz
16.7
60

FIRST TIME OPERATION:

1. Wipe the inside clean with detergent and hot water.
2. Fill water pan with 4 gallons of hot water.
MOBILE
DECONTAMINATION
Rev. 0 (7/21)
Page 1 of 6
Model #CGMH73A
Phase
NEMA
1
5-20P
FL-2424

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the CGMH73A and is the answer not in the manual?

Questions and answers

Summary of Contents for Cres Cor CGMH73A

  • Page 1 MOBILE FL-2424 DECONTAMINATION 5925 Heisley Road • Mentor, OH 44060-1833 Rev. 0 (7/21) Page 1 of 6 MOBILE DECONTAMINATION CABINET OPERATING and MAINTENANCE INSTRUCTIONS Model #CGMH73A Model no. Volts Watts Amps Hertz Phase NEMA CGMH73A 2000 16.7 5-20P HOW INSTALL UNIT: FIRST TIME OPERATION: 1.
  • Page 2 MOBILE DECONTAMINATION CABINET OPERATING and MAINTENANCE INSTRUCTIONS Model #CGMH73A NOTE: This cabinet operates with air temperature priority. Humidity will only start after cabinet is fully preheated. If cold or room temp water is used, humidity build-up time will be longer.
  • Page 3 5925 Heisley Road • Mentor, OH 44060-1833 Rev. 0 (7/21) Page 3 of 6 MOBILE DECONTAMINATION CABINET OPERATING and MAINTENANCE INSTRUCTIONS Model #CGMH73A DAILY MAINTENANCE: WATER PAN Drain, wipe and fill water pan daily. (Clear vinyl drain Delime or descale water pan parts as hose is provided.)
  • Page 4 5925 Heisley Road • Mentor, OH 44060-1833 Rev. 0 (7/21) Page 4 of 6 MOBILE DECONTAMINATION CABINET OPERATING and MAINTENANCE INSTRUCTIONS Model #CGMH73A WARNING RISK OF FIRE OR ELECTRIC SHOCK DO NOT OPEN WARNING: TO REDUCE THE RISK OF FIRE OR ELECTRIC SHOCK,...
  • Page 5 MOBILE FL-2424 DECONTAMINATION 5925 Heisley Road • Mentor, OH 44060-1833 Rev. 0 (7/21) Page 5 of 6 MOBILE DECONTAMINATION CABINET OPERATING and MAINTENANCE INSTRUCTIONS Model #CGMH73A REPLACEMENT PARTS: 120V 208V 240V Blower Kit 0769-180-SS-K 0769-182-SS-K 0769-182-SS-K Heater Kit, Air 2000W...
  • Page 6 MOBILE FL-2424 DECONTAMINATION 5925 Heisley Road • Mentor, OH 44060-1833 Rev. 0 (7/21) Page 6 of 6 MOBILE DECONTAMINATION CABINET OPERATING and MAINTENANCE INSTRUCTIONS Model #CGMH73A WIRING DIAGRAM: L1 T1 A2(-) A1(+) L1 T1 A2(-) A1(+) Phone: 440-350-1100 Ext: 268 • www.crescor-hs.com...

Table of Contents