STEMCELL ThawSTAR CFT2 Quick Start Manual

STEMCELL ThawSTAR CFT2 Quick Start Manual

Advertisement

ThawSTAR

Quick Start

Upon power-up of instrument, please allow 5 minutes for instrument warm-up.
Instrument is warmed up and in "ready" state.
Insert a clean, tightly capped frozen
1.8 - 2.0 mL cryogenic vial (below -70°C)
containing up to 1.8 mL volume into the
opening on the top of the instrument.
Beginning of solid to liquid phase change is
detected by instrument.
Upon thaw completion, the vial is raised
approximately 1 cm (0.4 in).
CFT2
®
Push the top of the vial down gently until
the instrument emits a deep dual tone.
The vial will now remain in a depressed
position.
Countdown of the last 10 seconds of the
thawing process begins.
Remove vial immediately and for a few
seconds, manually agitate the vial gently
from side to side to equilibrate the liquid
temperature until no ice can be seen within
the vial. Begin downstream protocol.
Quick Start Guide

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ThawSTAR CFT2 and is the answer not in the manual?

Questions and answers

Summary of Contents for STEMCELL ThawSTAR CFT2

  • Page 1 ThawSTAR CFT2 ® Quick Start Guide Quick Start Upon power-up of instrument, please allow 5 minutes for instrument warm-up. Instrument is warmed up and in “ready” state. Push the top of the vial down gently until Insert a clean, tightly capped frozen the instrument emits a deep dual tone.
  • Page 2 Error Codes Error Code 2: Solution: Failure to warm up properly; or oth- Check power source connection at er mechanical fault in the heating back of instrument, cable connection system. to power adapter and the plug con- nection to the electrical outlet. Turn the instrument off and back on.