Page 1
GP1-PBA1 Precision Bridge Adapter for GP1 Logger, SWT3/4/5 tensiometers and bridge sensors Quick Start Guide Version 1.0 Delta-T Devices Ltd...
Page 2
Installation Step 1: You can install one or two GP1-PBA1s in the GP1. Align the GP1-PBA1 with the CH1 or CH2 terminals so that one of its 5 pins overhangs the GP1 terminal block. Insert the remaining four pins into the four screw terminals of CH1 or CH2. Tighten the screws.
Page 3
1. Connect the GP1 to your PC and start DeltaLINK. 2. Select the Program window and click Change. 3. Double-click the channel with GP1-PBA1 installed. 4. In Voltage Channel Properties: • Enter a Label • Select Sensor Type : <custom>.
Page 4
In Voltage Channel Properties, enter appropriate Warmup, Units, Minimum, Maximum and Resolution values for your sensor. To record sensor readings in mV: enter a1 = 0.02 [the GP1-PBA1 gain is 50] For readings in ‘engineering units’, calculate a1 from sensitivity excitation voltage values provided by the manufacturer of the sensor.