FOM 330 frying oil tester in 8 Steps
1
Preparation
Important! For optimum accuracy rest oil
for at least 20 minutes before measuring.
No food must be cooking at the time of
measurement.
5
Measurement
The measurement is finished, when the TPM-
value in the display stops blinking and the
NextStep-indicator suggests to stop stirring.
For service and calibration please contact:
VITO AG - Eltastr. 6 - 78532 Tuttlingen, Germany
P: +49 7461 96289-0 - F: +49 7461 96289-12 - info@VITO.ag - www.VITO.ag
2
Start
Press the ON / OFF-button. When -- appears
in the display, the instrument is ready for the
first measurement. The NextStep-indicator
suggests immersing the probe into the oil.
6
Evaluation
Good oil quality: signal lamp flashes green
Medium oil quality: signal lamp flashes yellow
Bad oil quality: signal lamp flashes red
3
Immersion
Immerse the probe into the oil between
the MIN / MAX markings on the probe. The
NextStep-indicator suggests to start stirring.
7
Cleaning
Take the probe out of the oil and clean the
probe.
4
Stirring
Stir
slightly
until
the
temperature
the display corresponds roughly to the
temperature of the oil. This can take about
20
seconds.
8
Restart
Immerse the probe into the oil between the
MIN / MAX markings on the probe and stir
slightly.
3
-
Ver. 1.0
in
8

Advertisement

loading
Need help?

Need help?

Do you have a question about the FOM 330 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for VITO FOM 330

  • Page 1 NextStep-indicator suggests to stop stirring. Bad oil quality: signal lamp flashes red slightly. For service and calibration please contact: Ver. 1.0 VITO AG - Eltastr. 6 - 78532 Tuttlingen, Germany P: +49 7461 96289-0 - F: +49 7461 96289-12 - info@VITO.ag - www.VITO.ag...
  • Page 2 Sie leicht um. umzurühren. blinkt rot. Für Service und Kalibrierung wenden Sie sich bitte an: Ver. 1.0 VITO AG - Eltastr. 6 - 78532 Tuttlingen, Deutschland T: +49 7461 96289-0 - F: +49 7461 96289-12 - info@VITO.ag - www.VITO.ag...
  • Page 3 Qualité d’huile mauvaise : le signal lumineux clignote en rouge Siège principal: Ver. 1.0 VITO AG - Eltastr. 6 - 78532 Tuttlingen, Allemagne T: +49 7461 96289-0 - F: +49 7461 96289-12 - info@VITO.ag - www.VITO.ag...
  • Page 4 Gemiddelde kwaliteit olie: het lampje knippert om te stoppen met roeren. geel Slechte kwaliteit olie: het lampje knippert rood Hoofdkwartier: Ver. 1.0 VITO AG - Eltastr. 6 - 78532 Tuttlingen, Duitsland T: +49 7461 96289-0 - F: +49 7461 96289-12 - info@VITO.ag - www.VITO.ag...
  • Page 5 Óleo de má qualidade: Indicador luminoso pisca em vermelho Sede principal: Ver. 1.0 VITO AG - Eltastr. 6 - 78532 Tuttlingen, Alemania T: +49 7461 96289-0 - F: +49 7461 96289-12 - info@VITO.ag - www.VITO.ag...
  • Page 6 Olio esausto: segnale lampeggiante rosso Sede principale: Ver. 1.0 VITO AG - Eltastr. 6 - 78532 Tuttlingen, Germania T: +49 7461 96289-0 - F: +49 7461 96289-12 - info@VITO.ag - www.VITO.ag...
  • Page 7 Aceite de media calidad: La señal destella amarilla. Aceite de mala calidad: La señal destella roja Para servicio y calibración diríjase a: Ver. 1.0 VITO AG - Eltastr. 6 - 78532 Tuttlingen, Germania T: +49 7461 96289-0 - F: +49 7461 96289-12 - info@VITO.ag - www.VITO.ag...