MULTIVAP
TM
Nitrogen Evaporation System
Model 118
DIGITAL & TIME CONTROL SYSTEM
INSTRUCTION MANUAL
O
A
I
RGANOMATION
SSOCIATES
NC.
266 RIVER ROAD WEST
BERLIN, MA 01503
U. S. A.
Tel: 888-838-7300
Fax: 978-838-2786

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the MULTIVAP 118 and is the answer not in the manual?

Questions and answers

Summary of Contents for Organomation Associates MULTIVAP 118

  • Page 1 MULTIVAP Nitrogen Evaporation System Model 118 DIGITAL & TIME CONTROL SYSTEM INSTRUCTION MANUAL RGANOMATION SSOCIATES 266 RIVER ROAD WEST BERLIN, MA 01503 U. S. A. Tel: 888-838-7300 Fax: 978-838-2786...
  • Page 3: Table Of Contents

    Specifications UL - Underwriters Laboratories Warranty Wiring Diagrams Z-Purge NFPA Guidelines Accessories Other Products and Services Contacting Organomation and its Representatives Product designs are subject to change without notice © 1999 Organomation Associates Inc. M99.0 Printed in U. S. A.
  • Page 4: Introduction

    Contact Organomation Associates Inc. immediately if any damage or discrepancies are found. Your shipment should contain one or more of the instruments shown below. Option codes are listed on the next page.
  • Page 5 MULTIVAP 118 INTRODUCTION Option Codes and additional items shipped The following list contains option codes and items which may have been shipped in conjunction with the standard parts shown on the previous page. Please check your packing list and order information carefully to determine if these items are included in your shipment.
  • Page 6: Instrument Description

    MULTIVAP 118 INTRODUCTION Instrument Description The MULTIVAP Nitrogen Evaporation System is designed for general evaporation and / or concentration of analytical or biological samples in a variety of test tube based glassware under controlled and reproducible conditions. The complete instrument is shown below.
  • Page 7 MULTIVAP 118 INTRODUCTION PARTS LIST Description Item Part Name Cat # Hoist Assembly Assists lifting of instrument. Connects gas tube for swing away movement. Delivers gas to manifold. Toggle Gas Valve On/Off gas control for each manifold row. Lift Knob Rotate to raise and lower manifold.
  • Page 8 MULTIVAP 118 INSTALLATION Location The MULTIVAP Evaporator System should be located on a bench top or in a chemical fume hood if hazardous or flammable materials and solvents are to be used. The location should provide the necessary support services for the instrument. These include electrical power (required for water bath) and a clean inert gas source (Air or Nitrogen).
  • Page 9: Installation

    MULTIVAP 118 INSTALLATION Dry Bath Setup (Continued) Type-Z Purge Intrinsically Safe Bath option - If you do not have this option, please proceed to the next section. Procedures for operating this system may be found in the Operation Section. Quick start instructions are posted on the front of the bath. A copy of the NFPA (National Fire Protection Assoc.) guidelines for intrinsically safe equip-...
  • Page 10 MULTIVAP 118 INSTALLATION Dry Bath Setup (Continued) Flow meter Assembly - Provided with all N-EVAP Systems. If an OA-SYS bath was purchased without an N-EVAP instrument, proceed to the next section. See figure 4 for parts identification and position. Attach the flow meter and bracket to the top left rear corner of the Control Box with the two 10/32 x 1/2”...
  • Page 11: Instrument Setup

    MULTIVAP 118 INSTALLATION Instrument Setup Place the rod into the block located on the back of the heating unit. Place the manifold assembly onto the top of the rod and slide downwards until the top of the rod is secured in the hoist assembly.
  • Page 12 MULTIVAP 118 INSTALLATION Instrument Setup (Continued) Needles and Pipets - SS Luer Lok Needles are supplied with the standard MULTIVAP system. Optional Pasteur Pipet Adapters allow the use of glass pipets. These fittings replace SS Needles on the MULTIVAP. Pasteur Pipet Adapters may be purchased separately with the standard MULTIVAP, allowing the use of both needles and pipets.
  • Page 13 MULTIVAP 118 SAFETY Safety Considerations READ THIS SECTION BEFORE EQUIPMENT OPERATION! This equipment is designed for use in the Analytical or Environmental Laboratory by trained laboratory personnel for evaporative applications. Use of this equipment beyond its stated intended purpose and operating parameters is not recommended and will be the sole responsibility of the user.
  • Page 14: Operation

    MULTIVAP 118 OPERATION Instrument Controls - Main Control Panel Digital Controller - Controls bath temperature Timer Controller - Controls timed functions when selected. Used for bath preheat or automated control of services. Time Switch - Starts or resets timed operation, momentary switch.
  • Page 15 MULTIVAP 118 OPERATION Type Z-Purge Bath Operation - Optional What It Does The Type-Z Purged system prevents ignition of flammable materials caused by contact with electrical components inside the heating unit. How It Works The concept behind this purge system is to create a small positive pressure gradient inside the bath case.
  • Page 16: Planning And Preparation

    Under no circumstances should this equipment be used with materials capable of auto ignition below 150 Degrees Centigrade including materials containing peroxides. Please Contact Organomation Associates Technical Support if you have any questions concerning the use of TYPE-Z Purged equipment or questionable materials in OAI equipment.
  • Page 17: Bath Operation

    MULTIVAP 118 OPERATION Bath Operation Press the reset button on the GFCI (if present). Turn the bath rocker switch on. Digital Electronic Control - Adjust the digital controller to the desired temperature set point. The controller set point may be adjusted by depressing the “*” on the front panel and depressing the up or down arrow keys to the desired temperature.
  • Page 18 MULTIVAP 118 OPERATION Bath Operation Timed Gas and Heat Control The Time control system allows several operating modes which can be set for different operations. The timer may be used to preheat the bath in the morning prior to use, to shut down the system after a specific period of time (unattended operation), or for non- dryness endpoint operations.
  • Page 19 MULTIVAP 118 OPERATION Bath Operation (continued) Timed Gas and Heat Control (Continued) Automatic shutdown of gas or heat services - unattended operation To shut down the bath services after a period of time during a sample evaporative run, ideal when each sample is to be taken to dryness and the operator will not be present at the completion of the run.
  • Page 20 MULTIVAP 118 OPERATION Bath Operation (continued) Timed Gas and Heat Control (Continued) Non-dryness automated endpoint - unattended operation The MULTIVAP system my be used for timed non-dryness endpoint evaporation. order for this operation to be successful the following parameters must be maintained.
  • Page 21: Instrument Operation

    MULTIVAP 118 OPERATION Instrument Operation Turn on the heating and gas control system. Adjust the digital temperature control to the desired temperature. For samples which are not heat sensitive, adjust the controller to a temperature which is two degrees below the boiling point of the sample.
  • Page 22 MULTIVAP 118 OPERATION Instrument Operation (continued) When using 19 gauge needles a minimum flow of 0.3 LPM per sample is recommended. Higher flows may be used as desired and is recommended for larger needle sizes. High flows will improve evaporation rate however needle tip to sample surface will have to be increased to avoid splashing.
  • Page 23: Maintenance And Cleaning

    MULTIVAP 118 MAINTENANCE Maintenance and Cleaning The MULTIVAP Evaporation system is manufactured from extremely durable materials and may last for years if operated and maintained properly. The following guidelines are recommended for use with MULTIVAP systems. Cleaning - The anodized aluminum components may be cleaned with a non-abrasive scouring pad followed by damp cloth wipe with clean water.
  • Page 24: Troubleshooting

    MULTIVAP 118 TROUBLESHOOTING SYMPTOMS CAUSES SOLUTIONS No Power to bath. Electrical outlet not energized. Energize electrical outlet. Bath power cord not plugged in. Plug in bath power cord. GFCI not reset. Reset light gray switch on GFCI. Internal electrical fault.
  • Page 25 MULTIVAP 118 WIRING DIAGRAMS POWER CORD 120 VOLT TIMER OCTAL NEUT. PIN BASE ON / OFF SWITCH SOLENOID DIGITAL CONTROLLER 1 2 3 4 5 6 7 8 SOLID STATE RELAY HIGH TEMPERATURE RED ALERT LIGHT PROTECTION SWITCH GREEN TIME LIGHT...
  • Page 26 MULTIVAP 118 WIRING DIAGRAMS POWER CORD 240 VOLT TIMER OCTAL NEUT. PIN BASE ON / OFF SWITCH SOLENOID DIGITAL CONTROLLER 1 2 3 4 5 6 7 8 SOLID STATE RELAY HIGH TEMPERATURE RED ALERT LIGHT PROTECTION SWITCH GREEN TIME LIGHT...
  • Page 27: Service And Returns

    Contact Organomation Technical Support Department Before returning any product to Organomation Associates for any reason, please contact the Technical Support Department, toll free at 888-838-7300. Support is available Monday through Friday from 8:30 AM to 5:00 PM EST. Support is available free of charge to customers of Organomation in good standing for all products manufactured by Organomation.
  • Page 28: Shipping - Claims For Damage Or Shortage

    TECHNICAL INFORMATION Shipping - Claims for damage and shortage Organomation Associates Inc. makes a sincere effort to ensure your purchase is properly packed and all items listed on the packing slip are in fact enclosed with the shipment. In the event that your purchase is damaged or if any items are missing, please follow the procedures below.
  • Page 29: Specifications

    MULTIVAP 118 TECHNICAL INFORMATION Specifications Electrical Requirements: 120 or 240 VAC single phase, 50 - 60 Hz. 3 wire grounded outlet required. GFCI (Ground Fault Circuit Interrupter) optional. Model 11830 & 11848 heating system W* Model 11860 & 11896 heating system 800 W* * 240V units divide wattage by 2.
  • Page 30: Warranty

    This warranty is subject to the provisions and limitations as set forth below. Organomation Associates, Inc. and its authorized representatives, here-in-after referred to as Supplier, provide the following terms and conditions to the original end user, here-in-after referred to as Customer.
  • Page 31 EC Declaration of Conformity Revised January 3, 2003 We, Organomation Associates Inc a corporation registered in Massachusetts, United States of America, declare under sole responsibility that the following equipment to which this declaration relates, meets the principal protection requirements and is in conformity with relevant sections of the applicable EC standards and other normative documents.