Thank you for choosing our product. This instruction is intended to facilitate correct operation, safe use, and taking full advantage of the recorder's functionalities. Before you start the device, please read and understand this instruction. In the event of any additional questions, please contact our technical adviser. CONTENTS 1.
1. INSTALLATION GUIDELINES The device is designed so as to ensure an appropriate level of immunity to most interferences that may occur in industrial environments. In environments of unknown level of interferences, it is recommended to implement the following measures so as to prevent potential interference with the operation of the device: use shielded sensor, and signal cables, whereby the earthing of the shield should be single-point and located as close to the device as possible;...
- an SD memory card (2 GB); - an SD/MMC card reader; - a stabilized power supply, 5 V/150 mA; NOTE: − before you start working with the recorder, read this instruction, perform the actions described in chapters 8, and then properly connect the sensor or the input signal and configure the parameters (chapter 9);...
Communication interface for work USB, drivers compatible with Windows XP/Vista/7/8/10 with a computer programmable 10 s to 24 hours Measurement and data recording interval Data storage memory (non-volatile): - internal 4MB, FLASH, file system FAT12, recording of up to 80,000 measurements - external SD/MMC card FAT16, FAT32, maximum size 2 GB, recommended size ≤...
...) and two exchangeable disks: internal memory 4 MB with the AR233 label and SD/MMC memory (available after the card is put in the “SD/MMC CARD” slot). In the internal memory the configuration text file AR233.cfg can be seen (chapter 9).
“csv” text file created in the internal memory in the recorder or on an SD/MMC card (chapter 11) The latest versions of the aforementioned programs are also available at the website (www.apar.pl). The detailed descriptions of the aforementioned applications can be found in the installation folders.
Table 9. List of configuration parameters User Name Description of parameter settings RTC (1) internal real time clock (chapter 9.1) RECORDING OPTIONS (chapter 9.2) recording start time Start recording end time Stop data recording interval MPer SDest selection of memory to save archive files (to record data) CONFIGURATION OF THE MEASUREMENT INPUT (chapter 9.3) type of measurement input rrtd...
− off-line configuration of the parameters of the device (from the AR233.cfg configuration file after the “COPY” button has been pressed, see chapter 9) − transfer of “csv” archive files from the internal memory onto the card (this is done after the “COPY” button has been pressed and the authorization has been successful or when the authorization demand is switched off;...
SD/MMC card. The file name contains the device type (AR233), the identification number 16:ID, and the date and time of creation of the file, e.g. „AR233_1_2009-11-30_10-57-16.csv”. The format of a single data record is the following: “sequential number of the event;date;time;identifier of the event;measurement;internal...
(internal memory or the SD/MMC card) and during operation of the ARSOFT-CFG WZ1 software. − keep a copy of the current configuration file (AR233.cfg) in external memory (SD/MMC card, computer disk, etc.) for the purpose of archiving and future off-line configuration.