CALIBRATION PROCEDURE
PXIe-5111
This document contains the verification and adjustment procedures for the PXIe-5111. Refer to
ni.com/calibration
for more information about calibration solutions.
Contents
Required Software.....................................................................................................................1
Related Documentation.............................................................................................................2
Test Equipment..........................................................................................................................2
Test Conditions..........................................................................................................................4
Password................................................................................................................................... 5
Calibration Interval................................................................................................................... 5
As-Found and As-Left Limits................................................................................................... 5
Calibration Overview................................................................................................................ 5
Test System Characterization....................................................................................................6
Zeroing the Power Sensor................................................................................................. 6
Characterizing Power Splitter Amplitude Balance and Loss............................................6
Verification................................................................................................................................9
Verifying DC Accuracy...................................................................................................11
Verifying 50 Ω Bandwidth.............................................................................................. 17
Adjustment.............................................................................................................................. 22
Adjusting the PXIe-5111.................................................................................................22
Adjusting 1 MΩ Compensation Attenuator.................................................................... 22
Adjusting 1 MΩ DC Reference.......................................................................................23
Adjusting 50 Ω DC Reference........................................................................................ 24
Reverification..........................................................................................................................24
Updating Verification Date and Time..................................................................................... 25
Worldwide Support and Services............................................................................................ 25
Product Certifications and Declarations................................................................................. 25

Required Software

Calibrating the PXIe-5111 requires you to install the following software on the calibration
system:
Compatible version of the NI-SCOPE instrument driver. NI recommends you use the
latest version of NI-SCOPE for external calibration.
Supported application development environment (ADE): LabVIEW
You can download all required software from ni.com/downloads.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the PXIe-5111 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for NI PXIe-5111

  • Page 1: Table Of Contents

    CALIBRATION PROCEDURE PXIe-5111 This document contains the verification and adjustment procedures for the PXIe-5111. Refer to ni.com/calibration for more information about calibration solutions. Contents Required Software........................1 Related Documentation......................2 Test Equipment..........................2 Test Conditions..........................4 Password........................... 5 Calibration Interval........................5 As-Found and As-Left Limits....................5 Calibration Overview........................
  • Page 2: Related Documentation

    • Aberrations: <2% of peak for the first 500 ns DC generation: • Amplitude: ±5.0 V into 50 Ω, ±118.0 V into 1 M Ω • Accuracy: ±(0.025% of output + 25 µV) 2 | ni.com | PXIe-5111 Calibration Procedure...
  • Page 3 • 50 Ω • VSWR: <1.25 bandwidth • Impedance: 50 Ω 50 Ω BNC Fairview Test system terminator (f) Microwave ST3B-F characterization • Frequency: DC to 326 MHz • VSWR: <1.1 PXIe-5111 Calibration Procedure | © National Instruments | 3...
  • Page 4: Test Conditions

    • 50 Ω bandwidth Test Conditions The following setup and environmental conditions are required to ensure the PXIe-5111 meets published specifications: • Allow all test instruments a warm-up time of at least the amount of time stated in their specifications documents.
  • Page 5: Password

    2 years As-Found and As-Left Limits The as-found limits are the published specifications for the PXIe-5111. NI uses these limits to determine whether the PXIe-5111 meets the specifications when it is received for calibration. Use the as-found limits during initial verification.
  • Page 6: Test System Characterization

    SMA (f)-to-N (m) adapter and an SMA (m)-to-SMA (m) cable. Connect an SMA (m)-to-BNC (m) adapter to one of the power splitter output ports. Refer to this assembly as splitter output 1. 6 | ni.com | PXIe-5111 Calibration Procedure...
  • Page 7 Configure the signal generator to generate a sine waveform with the following characteristics: • Frequency: the Test Point Frequency value from the Power Splitter Characterization table • Amplitude level: 2.0 dBm PXIe-5111 Calibration Procedure | © National Instruments | 7...
  • Page 8 1. Signal generator 5. BNC (f)-to-N (f) adapter 2. SMA (f)-to-N (m) adapter 6. SMA (m)-to-BNC (m) adapter 3. SMA (m)-to-SMA (m) cable 7. Power splitter 8. 50 Ω BNC terminator (f) 4. Power sensor 8 | ni.com | PXIe-5111 Calibration Procedure...
  • Page 9: Verification

    Verification of the PXIe-5111 is complete only after you have successfully completed all tests in this section using the As-Found Limits. Refer to the following for the names, locations, and functions of the PXIe-5111 front panel connectors. PXIe-5111 Calibration Procedure | © National Instruments | 9...
  • Page 10 5 PFI 2 6 PFI 3 7 5 V square Probe compensation Generates a 5 V square wave for passive wave terminal probe compensation. 8 Ground Probe compensation Ground reference for passive probe terminal compensation. 10 | ni.com | PXIe-5111 Calibration Procedure...
  • Page 11: Verifying Dc Accuracy

    Verifying DC Accuracy This procedure verifies the DC accuracy of the PXIe-5111 by comparing the voltage measured by the device to the value sourced by the voltage standard. Caution Avoid touching the connections when generating a high voltage from the calibrator.
  • Page 12 -4.975 ±37.7 ±17.3525 -4.5 -4.975 ±37.7 ±17.8275 0.95 ±39.2 ±19.17 0.95 ±39.2 ±19.835 -0.95 ±39.2 ±19.17 -0.95 ±39.2 ±19.835 4.95 ±55.2 ±26.37 4.95 ±55.2 ±27.035 -4.95 ±55.2 ±26.37 -4.95 ±55.2 ±27.035 ±78.2 ±38.72 12 | ni.com | PXIe-5111 Calibration Procedure...
  • Page 13 ±1.2975 0.1 V -0.0475 ±2.15 ±1.2975 0.1 V -0.0475 ±2.15 ±1.2975 0.1 V 5.0475 ±22.15 ±10.2975 0.1 V 5.0475 ±22.15 ±10.2975 0.1 V -5.0475 ±22.15 ±10.2975 0.1 V -5.0475 ±22.15 ±10.2975 PXIe-5111 Calibration Procedure | © National Instruments | 13...
  • Page 14 ±28 ±13.67 0.4 V -5.19 ±28 ±13.67 0.4 V -5.19 ±28 ±13.67 0.475 ±19.7 ±11.675 0.475 ±19.7 ±11.675 -0.475 ±19.7 ±11.675 -0.475 ±19.7 ±11.675 20.475 ±99.7 ±47.675 20.475 ±99.7 ±47.675 -20.475 ±99.7 ±47.675 14 | ni.com | PXIe-5111 Calibration Procedure...
  • Page 15 ±82.7 10 V 4.75 ±195.2 ±116.75 10 V 4.75 ±195.2 ±116.75 10 V -4.75 ±195.2 ±116.75 10 V -4.75 ±195.2 ±116.75 10 V 104.75 ±595.2 ±296.75 10 V 104.75 ±595.2 ±296.75 PXIe-5111 Calibration Procedure | © National Instruments | 15...
  • Page 16 ±1880.2 ±917 40 V -100 -119 ±1880.2 ±917 Connect the calibrator test head to channel 0 of the PXIe-5111. Configure the PXIe-5111 with the following settings: • Input impedance: the Input Impedance value from the DC Accuracy Verification table. •...
  • Page 17: Verifying 50 Ω Bandwidth

    Repeat steps 2 through 8 for each configuration listed in the DC Accuracy Verification table. 10. Disable the calibrator output. 11. Connect the calibrator test head to channel 1 of the PXIe-5111 and repeat steps 2 through 9 for each configuration listed in the DC Accuracy Verification table.
  • Page 18 0.05 325.1 -3.0 to 1.3 -2.6 to 1.3 1.5 G 0.05 1.5 G 325.1 -3.0 to 1.3 -2.6 to 1.3 0.05 325.1 -3.0 to 1.3 -2.6 to 1.3 1.5 G 10 V 0.05 18 | ni.com | PXIe-5111 Calibration Procedure...
  • Page 19 1 and the signal generator is connected to the input port of the power splitter. The following figure illustrates the hardware setup. PXIe-5111 Calibration Procedure | © National Instruments | 19...
  • Page 20 8. Signal generator Configure the power sensor with the following settings: • Power measurement: continuous average • Path selection: automatic • Averaging: automatic • Averaging resolution: 4 (0.001 dB) • Aperture: 20 ms 20 | ni.com | PXIe-5111 Calibration Procedure...
  • Page 21 Test System Characterization section for the current Test Point Frequency. Use the PXIe-5111 to acquire and measure the power using the Extract Single Tone Information VI, converting the result from Vpk to dBm. Record the result as device input power.
  • Page 22: Adjustment

    16. Connect splitter output 2 of the power sensor assembly to channel 1 of the PXIe-5111 and repeat steps 2 through 14 for each configuration listed in the 50 Ω Bandwidth Verification table. 17. Disable the signal generator output. Adjustment This section describes the steps needed to adjust the PXIe-5111 to meet published specifications.
  • Page 23: Adjusting 1 Mω Dc Reference

    Caution Avoid touching the connections when generating a high voltage from the calibrator. Connect the calibrator test head to the specified channel of the PXIe-5111. Call the ni5110 Ext Cal API.lvlib:DC Reference Cal Initialize VI with the following settings: •...
  • Page 24: Adjusting 50 Ω Dc Reference

    PXIe-5111. Note If any test fails reverification after performing an adjustment, verify that you have met the test conditions before returning your PXIe-5111 to NI. Refer to the Worldwide Support and Services section for information about support resources or service requests.
  • Page 25: Updating Verification Date And Time

    NI. NI corporate headquarters is located at 11500 North Mopac Expressway, Austin, Texas, 78759-3504. NI also has offices located around the world. For support in the United States, create your service request at ni.com/support...
  • Page 26 . You can find patents.txt ni.com/patents information about end-user license agreements (EULAs) and third-party legal notices in the readme file for your NI product. Refer to the Export Compliance Information at for the NI global trade compliance policy and how ni.com/legal/export-compliance...

Table of Contents