The Setup Mode allows the user to adjust the configuration of the meter. The Setup Mode
is entered by pressing The ON/SETUP key when the unit is on. The parameter and the
current value will alternately flash in the display. The following table indicates the
Parameters that are available, their meaning and available setting range:
PARAMETER
BAT
CONT
AOFF
BEEP
FCON
BAT – (MODEL 1300 ONLY) This parameter gives an indication of the remaining battery
power. The battery life is displayed as percent of life.
CONT – This parameter controls the contrast of the LCD display. A Higher setting will
cause the display to darken. A lower setting will cause the display to lighten.
AOFF – This parameter determines the period of inactivity before the meter is turned OFF.
A timer is started when the meter is turned ON and is reset each time a key is pressed.
When the timer reaches the value set in this parameter, the power is automatically turned
OFF.
NOTE: Setting this parameter to 0 disables the Auto Off timer. When running from
line power, the meter will not automatically shut off.
BEEP – This parameter controls the length of the audio feedback beep (click) that occurs
when a key is depressed. A Higher setting will cause a longer (louder) beep. If the value is
set to 0, the beep is eliminated.
FCON – This parameter controls the Filter Constant. The software has a Digital Filter that
averages the pressure readings to produce a stable display. This setting determines the
number of samples that are averaged in the digital filter. Increasing this setting will cause a
more stable display. However, it will also cause a slower response to small changes in
pressure. The best setting is the smallest number that provides a stable display.

SETUP MODE

DESCRIPTION
REMAINING BATTERY LIFE
CONTRAST
AUTO OFF TIMER
KEY BEEP LENGTH
DIGITAL FILTER CONSTANT
12
RANGE
0 - 100 %
0 - 15
0 - 30 MINUTES
0 - 15
1 - 255