Page 2
LIMITED WARRANTY AND LIMITATION OF LIABILITY Each Fluke product is warranted to be free from defects in material and workmanship under normal use and service. The warranty period is one year and begins on the date of shipment. Parts, product repairs, and services are warranted for 90 days.
Preparation This manual covers the P5514B-70M and P5514B-70M-EP. The Comparison Test Pump checks pressure-measuring instruments against Master Test Gauges. Note The terms “Master Test Gauge” and “Gauge” in this document refer to any pressure-measuring instrument such as Transfer Standards, Digital Calibrators, and Transducers.
Hydraulic Comparison Test Pump DUT Connection DUT Connection W Caution To prevent Product damage or damage to a gauge: • To prevent incorrect sealing that creates an unsafe condition, do not use thread- sealing tape or any other sealing method on the DUT or the adapters. The Gauge Adapter sealing system can be hand-tight sealed up to 70 MPa (10500 psi).
P5514B-70M/P5514B-70M-EP Users Manual Operation Priming Follow these steps: 1. Open the reservoir valve one turn counter clockwise and turn screw press fully in. 2. Close the valve and turn screw press fully out. 3. Open the valve and turn screw, press fully in.
Hydraulic Comparison Test Pump Ancillary Equipment W Caution Never release the system pressure by using the reservoir valve alone, as sudden depressurization will shock the system, which can damage the sensitive instruments attached to the system. Ancillary Equipment Pointer Remover/Punch, P5551 To remove and refit the pointer of a pressure gauge, use the pointer remover/punch instrument.
P5514B-70M/P5514B-70M-EP Users Manual Master Gauges See the associated instruction manuals for the gauges in Table for detailed lists of available ranges and specifications. Table 2. Gauges Model Nominal Specification 700G (See Figure 2) 0.05 % FS 700RG 0.04 % Reading + 0.01 % FS 2700G 0.02 % FS...