CONTROLLED ENVIRONMENT
CHAMBER
Model 5506-00 & 5506-11
Plus
Packages D & E
Operating Manual
3/07
www.electrotechsystems.com

Advertisement

Table of Contents

Troubleshooting

loading
Need help?

Need help?

Do you have a question about the 5506-00 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for ETS 5506-00

  • Page 1 CONTROLLED ENVIRONMENT CHAMBER Model 5506-00 & 5506-11 Plus Packages D & E Operating Manual 3/07 www.electrotechsystems.com...
  • Page 2: Table Of Contents

    TABLE OF CONTENTS Page GENERAL Basic Chambers Standard Packages 1.2.1 Chamber Specifications: 1.2.2 Pkg. D: 1.2.3 Pkg. E Unpacking CHAMBER Description Installation PACKAGE SYSTEMS Package D 3.1.1 Description 3.1.2 System Function 3.1.3 Installation 3.1.4 Operation 3.1.5 Calibration 3.1.6 Troubleshooting 3.1.6 Specifications Package E 3.2.1 Description 3.2.2 System Function...
  • Page 3: General

    The following charts show the time typically required to decrease and increase humidity (Blue = RH, Red = T °C) plus humidity gradients using an ETS Controller in conjunction with the Model 5461 Desiccant/Pump Dehumidification System.
  • Page 4: Chamber Specifications

    This system, utilizing the ETS Model 5311 Controlled Gas Dehumidification System, shown in Figure 1.2-1, is primarily used for storage or single use applications.
  • Page 5: Pkg. E

    Figure 1.2-1: Package D Controlled Valve System 5506 None 1.2.3 Package E: This configuration incorporates an adjustable dehumidification control system, shown in Figure 1.2-2, utilizing the ETS Model 5112 Dehumidifier Controller and the Model 5461 Desiccant/Pump Dehumidification System. The controller can measure and display humidity over the entire range, but controls humidity only from 0 to 55% RH.
  • Page 6 1. Chamber with glove ports, 8” (203mm) gloves, 12”x12” (30x30cm) door with gasket seal, inlet/outlet ports for accessories. 2. Pliable sealer 3. Operating Manual Model 5506-00, Package D 1. Chamber 2. Model 5311 Controlled Valve assembly 3. Pliable sealer 4. Operating Manual Model 5506-00, Package E 1.
  • Page 7: Chamber

    2.0 CHAMBER Description The Model 5506 is a 9.0 cubic foot (0.26 cu. m), sealed chamber fabricated from 0.25” (6mm) and 0.375” (9.5mm) clear and white acrylic and measures 36”Wx24”Dx18”H (92x61x46 cm). Access to the chamber is through a 12” (30.5cm) square opening. A heavy-duty, 0.5” (13mm) clear acrylic door, secured by a single ½-turn latch, along with a compression resistant gasket ensures an airtight seal when the door is locked.
  • Page 8: Installation

    Installation Place the chamber on an appropriate support surface such as a bench or desk. Open the latch by turning the handle counterclockwise until the cam turns away from the chamber wall and then open the access door. Place the object(s)to be stored or tested inside the chamber and feed any cables through the cable pass-through located at the rear of the left panel.
  • Page 9: Description

    The standard factory setting is 5% RH. Other fixed RH set points are available as an option. Contact ETS to order. The sensor is mounted to the wall of the chamber using the ½” NPT compression fitting, shown in Figure 3.1-1.
  • Page 10: System Function

    Figure 3.1-2: Control/Solenoid/Flow Regulator unit The Control Unit is designed as a complete assembly. There are no user controls except the flow regulator adjust knob. The only indicator is a Green LED that indicates the solenoid is on and supplying gas to the chamber. 3.1.2 System Function The Humidity Sensor measures the relative humidity inside the chamber.
  • Page 11 unit. Then plug the power module into the appropriate power outlet (MAINS). This will automatically power up the Control unit. There is no power ON/OFF switch. 3. Wait 3 minutes and check the Green LED mounted on top of the unit. If the ambient humidity is above the fixed set point the solenoid will be activated and the LED should light.
  • Page 12: Operation

    NOTE: If returning to the factory for repair or replacement, first obtain a RMA number from ETS by calling 215-887-2196 Ext. 220. 3.1.6 Troubleshooting The Model 5311 should provide trouble-free service. If a problem with the system is suspected, it is recommended the fault be initially isolated to either the Sensor, Control Unit or solenoid.
  • Page 13: Specifications

    2. Green LED does not come on even though the humidity level inside the chamber is known to be at least several percent above the fixed set point of the Model 5311. Return unit to factory. 3. Obviously incorrect humidity control – If sensor was removed from its connector, check to see that it has been replaced properly.
  • Page 14: Description

    Package E utilizes the ETS Model 5461 Desiccant/Pump Dehumidification System to dry the chamber. As an option, the ETS Model 5465 Dry Gas Dehumidification System may be substituted for the Model 5461. This system uses a solenoid valve to...
  • Page 15 The sensor is mounted to the wall of the chamber using a ½” NPT compression fitting installed in the chamber wall. Other mounting configurations can also be used to meet the user’s specific requirement. The Control Unit provides the necessary indicators and controls to set the desired humidity level and to measure the actual humidity level in the chamber.
  • Page 16 The system Power switch and the controlled North American 3-prong grounded outlet are located on the rear panel as shown in Figure 3.2-3. Figure 3.2-3: Control Unit rear panel The Power switch, Fuse and AC line cord are separate components that control the power input to the Model 5112 Controller.
  • Page 17 NOTE: A 115 VAC pump or solenoid must be used when operating at 115 VAC and a 230 VAC pump or solenoid must be used when operating at 230 VAC. This 18” (46cm) hard wired cable provides the input/output interconnect between the Controller and the Humidity Sensor.
  • Page 18: System Function

    removed from the drying column and spread evenly, one granule deep on a tray. The desiccant should then be heated for approximately one (1) hour at about 400ºF (200º C). It should be allowed to cool in an airtight container before refilling the acrylic drying column.
  • Page 19: Installation

    DEHUMIDIFIER switch is in the OFF (out) position. Plug the Dehumidification System into the AC outlet. NOTE: All ETS operating systems utilize the standard 3-prong North American plug. A 230V/50Hz system will be marked as 230V, but will still have the same 3-prong plug 3.
  • Page 20 is reached. It will continue to cycle to maintain the set point level. 3.2.3.2 Installation The Humidity Sensor is normally mounted in the bulkhead using the compression fitting. However, other mounting locations may be used to meet user requirements. CAUTION: The sensing element is sealed within the sensor assembly protected against mechanical damage by a slotted housing cover.
  • Page 21: Operation

    4. Insert the remaining section of tube into the OUTLET or pressure side of the pump. 5. Cut the tube and insert the other end into the lower hose barb on the desiccator. 6. Cut an appropriate length of tubing and connect the desiccator upper hose barb to the upper hose barb on the chamber.
  • Page 22: Calibration

    NOTE: If returning to the factory for recalibration or repair, first obtain a RMA number from ETS by calling 215-887-2196 Ext. 220. Calibration of the Model 5112 is performed as a complete unit. Turn on the power and allow at least 5 minutes for the instrument to warm up.
  • Page 23: Troubleshooting

    to it. If the Sensor is to be calibrated outside the chamber, remove the Sensor from its installed position. 2. Turn on the power. Allow at least 5 minutes for the system to warm up. The Sensor and reference unit must be in a constant temperature environment of 73°F/23°C, ±5°F/3°C and humidity environment least...
  • Page 24 The following troubleshooting guide will assist the user in identifying various possible problems: 1. No Power. - Check power at the wall outlet. 2. No indication on LCD display, but Green Dehumidifier LED is on. - LCD display defective. Return unit to factory. 3.
  • Page 25: Specifications

    Figure 3.2-5: Front panel removal If the fuse/relay procedure does not fix the problem then replace U1 (MC1458) and recheck the system. If the system still fails to operate, return the unit to the factory for repair. 6. Unit reads either 0 or over 100% with sensor installed. - Sensor is defective.
  • Page 26: Model 5465 Dry Gas Dehumidification System (Optional)

    3.2.7.3 Drying Unit Column: Molded acrylic plastic. Dimensions: 2 5/8" O.D. x 11 3/8". Anodized aluminum cap: Fitted with "O-Ring" gasket. Safe working pressures: 90 psig max. Desiccant supports and coil spring: Cadmium plated steel. DRIERITE (Calcium Sulfate): Held firmly in place between felt filters. Connections: Plastic or rubber flexible tubing.
  • Page 27: Replacement Parts

    The Series 5400 Operating Systems should operate reliably for many years if maintained properly. Typical service includes gloves and desiccant. Listed are common replacement parts for the various operating systems and control units along with the corresponding ETS part number. ITEM ETS Part # 1.
  • Page 28: Warranty

    An RMA (Return Material Authorization) must be obtained from ETS prior to return of equipment. Equipment should be shipped prepaid and insured in the original packaging. If...