  • Page 1 Instruction Manual ExTox Biogas Analyser Bio-Compact...
  • Page 2 Page 2 of 12 BA_Bio-Compact_e_2015-10-22.doc...
  • Page 3 Instruction Manual ExTox Bio-Compact Foreword We thank you very much for your confidence in our products and us, the ExTox Gasmess- Systeme GmbH. The biogas analyser Bio-Compact as well as all other ExTox-Products and services stand for our high quality targets. Our business is the health protection of mankind, protection of the envi- ronment and installations.
  • Page 4: Table Of Contents

    Instruction Manual ExTox Bio-Compact Content Introduction Features of the Bio-Compact Indications and Facilities Configuration Description Functions 5.1.1 Parameter Settings 5.1.2 Pre-configured Relays and Digital Inputs 5.1.3 Calibration and Adjustment Transmitter-Module 5.2.1 Replacement of a Transmitter-Module Operation of Bio-Compact Flow rate monitoring...
  • Page 5: Introduction

    We kindly ask you not to repair the Bio-Compact or to perform any changes which go beyond the measures described herein. Otherwise you endanger your own safety and your warranty claims of merchantability.
  • Page 6: Indications And Facilities

    , and H S (only Bio-Compact-A) The modular construction and the use of the approved ExTox-Transmitters technology and con- trol units assure a cost effective acquirement and maintenance. The control units in combination with ExTox-Transmitters comply with EN 60079-29-1 and EN 45544 series for Gas Detection Systems.
  • Page 7: Description

    This phase is needed to transport the sampled gas from the process to the Bio-Compact using the gas flow. The gas suction pump is switched off two minutes before end and the gas flow is closed. Measurement in batch mode is prepared.
  • Page 8: Pre-Configured Relays And Digital Inputs

    60 min including the default settings which correspond to 24 measurements per day. Please keep in mind that the design of Bio-Compact is based on 24 measurements per day in maxi- mum. More frequent measurements can shorten the lifetime of sensors and components in the transmitter module.
  • Page 9 Instruction Manual ExTox Bio-Compact Calibration is done in five phases: Phase Indication on Display Course RINS AIR The system is flushed with air for three minutes READ ZERO The gas suction pump is switched off. The system waits one minute until stabilisation of measured values. Then the measured values are stored (zero point values for CH S or 20.9 % (v/v) for O...
  • Page 10: Transmitter-Module

    5.2.1 Replacement of a Transmitter Module On delivery of the Bio-Compact a new transmitter module is installed (Art. 290xxx). In case you need a new one you will get a replacement transmitter module (Art. 295xxx), on which all wear parts have been replaced. The module has completely been checked and adjusted. It has the same warranty as a new module.
  • Page 11: Operation Of Bio-Compact

    At the measuring point the difference in pressure in comparison to the environment may not exceed -100 hPa and + 50 hPa. Please ensure that the line from sampling till measured gas inlet of the Bio-Compact consists of suitable material and is protected against damage and leakages.
  • Page 12: Electrical Installation

    (KSF1, Art. 700304) can be installed before the line gets into the Bio-Compact. If necessary please contact your partner at ExTox. Operation in most of biogas compositions is reliably possible due to the construction of the Bio- Compact.