Advertisement

Quick Links

Version : 2020
Ultrasonic Cleaner
USA-UC-30L

Advertisement

Table of Contents
loading

Summary of Contents for USA Lab USA-UC-30L

  • Page 1 Version : 2020 Ultrasonic Cleaner USA-UC-30L...
  • Page 2: Table Of Contents

    TABLE OF CONTENTS Keep this manual on-hand so it can be used by all operators of the unit. Use the unit only in the way described in this manual. Failure to follow the instruction in this manual may cause wrong operation. 1.
  • Page 3: Principle Of Ultrasonic Cleaning

    1. Principle of ultrasonic cleaning Ultrasonic cleaner generates ultrasonic frequency electricity, which is converted by energy converter into high-frequency mechanical oscillation that is transferred to cleaning fluid. Ultrasound wave is beamed forwards unevenly in the cleaning fluid to make the fluid vibrate and generate tens of thousands of micro bubbles, which form and grow in the negative pressure area where the ultrasound wave travels longitudinally and close (crush out) in positive pressure area.
  • Page 4: Product Features

    • Before changing or discharging the cleaning fluid, make sure the fluid is at normal temperature and the ultrasonic wave generator is shut down and the power supply is disconnected. • Perform any maintenance operation only after the power supply is disconnected to avoid any safety accident.
  • Page 5: Specification

    4. Specification Model USA-UC-30L Tank Capacity Frequency 40KHz Ultrasonic Power 600W Heating Power 500W Time Setting 1~99 Minutes Temperature Setting 0~80℃ Power AC 110V 60Hz Inside Dimensions 500x300x200mm/ (LxWxH) 19.69”x11.81”x7.87” External Dimensions 560x320x385mm/ (LxWxH) 22.05”x12.60”x15.16” Gross Weight 16kg 5. Scope of application...
  • Page 6: Installation And Preparation

    6. Installation and preparation ① Place the machine horizontally at a well-ventilated and dry location. ② Select appropriate detergent, cleaning water or other solution based on properties of work piece to be cleaned. ③ Properly and firmly connect the plug to a three-pin power socket which is grounded reliably to ensure safety.
  • Page 7: Factors Influencing Performance

    Operation Instructions: ① Load cleaning fluid into the tank and connect the power supply (which shall be grounded reliably). After power-on, current actual temperature is displayed and the default temperature is 20℃. The ultrasound wave operating time is 99 minutes by default. ② Adjustment of ultrasound operating duration: Press time setting button, one push to increase one minute;...
  • Page 8: Troubleshooting

    9. Troubleshooting Symptoms Troubleshooting method ① Check whether the power cable has been connected properly (by inspecting operation of power indicator lamp). Failure of Ultrasonic wave ② Check whether PCB has been connected properly to the power socket. ③ Check whether the fuse tube is broken. ① Check whether the power voltage is normal.
  • Page 9: Maintenance

    10. Maintenance This machine shall be used, maintained by dedicated personnel and a full check on the equipment shall be performed regularly. ① Check the power supply and line connections for any looseness, overheating, damp absorption or poor contact. ② Inspect sealing of the bottom board of cleaning tank for any leakage. Check the joint between drain pipe and cleaning tank for leakage.
  • Page 10 Memo...
  • Page 11 Memo...
  • Page 12 11730 Globe St. Livonia MI 48150 Call us at 734-855-4890...

Table of Contents