Page 3
Cerner authorized service personnel. The product must not be altered without the prior written approval of Cerner. The user of this Product shall have the sole responsibility for any malfunction, which results from improper use, faulty maintenance, im- proper repair, damage or alteration by anyone other than Cerner or Cerner authorized providers.
Table of Contents INTRODUCTION Disposal SAFETY Electrical Explosion Radio Communication Description of Warning Labels DESCRIPTION System Overview Special Cables Features Fuse Ratings INSTALLATION/OPERATINg INSTRUCTIONS SPECIFICATIONS MDC-2 MDA- MAINTENANCE AND CLEANINg Operator’s Manual MDC-2...
This manual provides information on the correct use of the Connectivity Engine (MDC-2). It is up to the user to ensure that any applicable regulations regarding installation and operation of the MDC-2 are observed, and the proper power line cord is used (if different than the North American cord supplied).
MDC-2 until the AC power source protective conductor is fully functional. WARNINg: Do not place the MDC-2 in any position that may cause it to fall on the patient. Do not lift the MDC-2 by the power supply cord or other cables.
WARNINg: When attached to other devices, insure that the total chassis leakage currents of all units do not exceed 300 µA. CAUTION: The MDC-2 is to be used on a flat surface only. CAUTION: The use of Accessory equipment not complying with the equivalent safety require- ment of this equipment may lead to a reduced level of safety of the resulting system.
Page 9
Cardiopulmonary Corp. could void the user’s authority to operate the device. This device also meets the requirements of the Canadian Interference-Causing Equipment Regulations. Cet appareil numérique de la Classe A respecte toutes les exigencies du Reglement dur le Matériel Brouilleur du Canada. Operator’s Manual MDC-2...
Attention, consult ACCOMPANYING DOCUMENTS. Equipment connected should comply with UL-60601-1 or IEC-950 with configuration to IEC-60601-1-1. Fuse type/rating. Mains Power On Mains Power Off Protective earth (ground) Functional Ground WEEE Compliance CE mark ETL Designation: Conforms to UL STD. 60606-01 Operator’s Manual MDC-2...
802.11 a/b/g compliant radio. The MDC-2 will be located within 5 meters/16.4 feet of any fixed or portable medical device. The MDC-2 is capable of connecting up to eight medical devices using USB cabling to the MDA-1 connector. The MDA-1 is powered through the USB connection to the MDC-2.
The fuses are located behind the cover of the power entry module. To replace the fuses, remove the cover and replace them only with same type and rating T1.6A 250V (Metric 5x20mm). Power Entry Module Fuses (Behind panel) Operator’s Manual MDC-2...
Healthcare Information Technology Systems provider for a supported list of devices. Connecting MDA-1 to MDC-2 Connect the USB cable from the MDA- to one of the eight ports on the rear panel of the MDC-2. Note the MDC-2 must be placed on a flat surface.
Page 16
Communications Eight external USB Type A connectors (or similar) One external RJ45 10/100 port One internal SD card adapter for non-volatile memory expansion, up to 4GB Mechanical Size: Height: 2.5in. (6.2cm) , Width: 7.5in. (18.75cm), Depth: 5.5in. (13.75cm) Weight: <2.5 lbs (<...
Page 17
EMC Directive (2004/108/EC): EN 55022:2006 EN 55024:1998, A1:2001, A2:2002 EN 61000-3-2:2005 EN 61000-3-3:1995, A1:2001, A2:2005 o R&TTE Directive (1999/05/EC): ETSI EN 301 489-1 v1.7.1 ETSI EN 301 489-17 v1.2.1 o CAN/CSA C22.2 No 601.1-M90 o CE mark Operator’s Manual MDC-2...
• Do not use petroleum based or acetones solutions, or other harsh solvents, to clean the MDC-2. Cables Do not autoclave the cables. Wipe the cables using a mild detergent solution. Never submerge the cables in any liquid or allow liquids to enter the electrical connections.