The indicator comes with a power cable attached to the indicator. When the power cable is plugged in, the built-in battery is being recharged. The “recharge” light will be on. We strongly recommend recharging the indicator for at least 12 hours before the first-time use. - 3 - V2015.05 PS-IN202SS Indicator...
To save a new selection, press the [NET/GROSS] (Set) key .To exit without saving, press the [UNITS] (left) key to return to the current “P” heading. e. Repeat Steps 1 through 4 until the Setup Menu is programmed. - 4 - V2015.05 PS-IN202SS Indicator...
[NET/GROSS] to switch between Channel 1 and 2. When it is on Channel 2 mode, there is a “2” shown on the right upper corner of the screen. 2.4 USER (“S”) MENU a. Enter the Setup (“P”) menu. - 5 - V2015.05 PS-IN202SS Indicator...
To save a new selection, press the [NET/GROSS] (Set) key .To exit without saving, press the [UNITS] (left) key to return to the current “S” heading. g. Repeat Steps 2 through 5 until the Communication Menu is programmed. FIG 5 2.5 User Menu Chart FIG 6 - 6 - V2015.05 PS-IN202SS Indicator...
Scrolling down with the [ZERO] key one level begins the begin sequence procedure This sub menu will reset all parameters in the “P” and “S” menu to the Press [ZERO] key twice default setting. USE WITH CAUTION! to begin sequence - 7 - V2015.05 PS-IN202SS Indicator...
Page 9
Actuates the function that allows entry of a new ID No. Valid only when 0 - 999999 S6 is set to “1”. Pressing the [ZERO] key to scroll down one level begins 123456√ ID No. Entry the sequence. - 8 - V2015.05 PS-IN202SS Indicator...
Page 10
First set the Month/ Date/Year. Press up and down button to change the blinking numbers. Press Zero button to move forward. After the date is done, press Zero button to move to Time setting. When all done, press Zero button to save the changes. - 9 - V2015.05 PS-IN202SS Indicator...
[ZERO] key. Pressing the [PRINT] key or the [TARE] key will change the position of the flashing digit. d. After setting the exact value, press the [NET/GROSS] key to save the value. - 10 - V2015.05 PS-IN202SS Indicator...
Note: This procedure is intended for emergency use only in the case of non-volatile memory loss. A valid span calibration value, obtained from a successful P12 calibration procedure, must be used. a. While in the Setup mode, scroll to "P15", then scroll down once using the [ZERO] key. - 11 - V2015.05 PS-IN202SS Indicator...
Page 13
If the entered value is greater than zero, the display will show "E CAL 1" momentarily then revert back to P20. If a value of zero is entered, the indicator will briefly show "Err 5" then revert back to the screen described above in Step b. - 12 - V2015.05 PS-IN202SS Indicator...
Chapter 4 Operation 4.1 Display FIG 7 4.2 Display Details LED Enunciator Meaning Enunciator ZERO Centre Zero enunciator. This light is illuminated whenever the displayed weight is within ±0.25 divisions of true zero The indicator is displaying the net weight GROSS The indicator is displaying the gross weight TARE...
It will start with sampling and you will see “5 –“ on the screen. Keep pressing [UNITS] you will find the number changes to “10 –“, “20 – “…”100 -”. Place the - 14 - V2015.05 PS-IN202SS Indicator...
Page 16
You may press [ZERO] any time to return to the weighing mode. To clear the memory, press and hold [ACCU] then press [ZERO]. Customer Support: Toll Free: 844-735-5386 Office Direct: 818-885-5000 info@selletonscales.com www.selletonscales.com - 15 - V2015.05 PS-IN202SS Indicator...