Advertisement

Quick Links

INSTALLATION & OPERATING INSTRUCTIONS
Slop Hopper - AGSP
5035886TM_04 • 02/2020

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the AGSP and is the answer not in the manual?

Questions and answers

Summary of Contents for Arjo AGSP

  • Page 1 INSTALLATION & OPERATING INSTRUCTIONS Slop Hopper - AGSP 5035886TM_04 • 02/2020...
  • Page 2 Design Policy and Copyright ® and ™ are trademarks belonging to the Arjo group of companies. © Arjo 2020. As our policy is one of continuous improvement, we reserve the right to modify designs without prior notice. The content of this publication may not be copied either whole or in part without the...
  • Page 3: Table Of Contents

    Contents Safety Warning ............4 Safety Regulations .
  • Page 4: Safety Warning

    SAFETY WARNING Safety Regulations This machine has been designed with a number of built-in safety devices. To avoid injury, it is highly important not to bypass or disable these safety devices. Operators and maintenance personnel must undergo safety training for the machine. All personnel who handle chemicals for washing and disinfection must understand the washing process, possible health hazards and ways of detecting leaks of toxic chemicals.
  • Page 5 “I” and “O” to identify the position of the switch. Environmental Protection Incorrect disposal of this equipment and its component parts, particularly batteries or other electrical components, may produce substances that are hazardous to the environment. To minimise these hazards, contact Arjo for information on correct disposal.
  • Page 6: Introduction

    If you have any difficulties using the Slop Hopper contact your local Arjo sales office, listed at the end of this manual. Intended Use The is designed to flush away liquids and...
  • Page 7: How To Use

    Cleaning The should be cleaned and disinfected Slop Hopper regularly by trained personnel. Most commercially Arjo available cleaning products can be used except chlorine based products. Caution Never use products containing chlorine, abrasives or steel wool for the cleaning since such products will lead to corrosion marks and scratches on the stainless steel.
  • Page 8: Components

    Components The Slop Hopper mainly consists of the following components. Conductor Label - attention high tension Connection rim flushing Label - rim flushing Sieve insert Plate - type Drainage tubing Hand button Control unit with cable set Pump Rim rinsing connection piece Water tank Flexible hose The ventilation openings in the base frame must not...
  • Page 9: Installation And Setup

    2. Installation and Setup Unpacking Remove the packaging. The folder containing documentation must accompany when it is delivered. Slop Hopper Before installing the , check it for Slop Hopper damage in transit. Installation Carry out the installation in the following order: 1.
  • Page 10 • Connect the to protective grounding Slop Hopper and to the supply voltage stated on the type plate. Check that the connection has suitable overcurrent protection. The correct fuse rating is stated on the type plate. • All cables must be classified for at least 90°C. 8.
  • Page 11: Maintenance

    3. Maintenance Servicing should be serviced by trained Slop Hopper Arjo personnel. Cleaning the Water The water filter is located at the inlet of the floating valve. Clean the water filter as follows: Filter • Turn off the water supply by using the upstream arranged shutoff unit •...
  • Page 12: Care

    • Replacement is the reverse of the above procedure. Clamp Siphon trap Care The has to be cleaned and disinfected Slop Hopper regularly. For removing spots, use commercially available cleaning products for chrome steel. Caution Never use products containing chlorine, abrasives or steel wool for the cleaning since such products will lead to corrosion marks and scratches on the stainless steel.
  • Page 13: Trouble Shooting

    4. Trouble Shooting Problem Cause Action Rim rinsing does not work. • Power supply interrupted. • Contact your local Arjo • Pump and/or control aftersales service office. system defective. Pump does not convey • Water supply turned off. • Open upstream shutoff •...
  • Page 14: Technical Description

    5. Technical Description Slop Hopper Model: AGSP Voltage: 230V Frequency 50 Hz Input rating Fuse protection: 230V - 50 Hz - 5A - FI 30mA Pump Power Rating: 300W Class protection: Kind of protection: IP22 Maximum water pressure: 1000 kPa Minimum supply water temperature: 3°C...
  • Page 15: Spare Parts

    6. Spare Parts Table 1: Standard structural components Item Qty. Part No. Description 467031600 Slop Hopper, Welded Frame (with Rim Flush) 467031601 Slop Hopper, Welded Frame (without Rim Flush) 463051000 Front Panel - Flat design (with Rim Flush) 463051001 Front Panel - Flat design (without Rim Flush) 4510008 Funnel shaped Slop Hopper Basin (Stainless) SPL1019...
  • Page 16 Table 2: Slop Hopper with rim flushing Item Qty. Part No. Description 81210573 Slop Hopper rebuilding kit for rim-flush 4670303 Pump assembly 4670302 Pump connection piece 4540076 Hand button Label rim flushing Plate - model 4550503 Flexible hose 4670301 Tank 4670300 Connection rim flushing 4670299...
  • Page 17 Table 3: Pump Item Qty. Part No. Description 4630515 Pump bracket 4630516 Pump casing 4530104 Booster pump 602618 Anti-vibration damper 600201 Locking nut...
  • Page 18 Table 4: Cable bushing Item Qty. Part No. Description AH1026695 Control unit 503607100 Conductor plate 4540077 Contact block 4540029 Floating switch - horizontal Label - Attention high voltage...
  • Page 19 Table 5: Control unit Qty. Part No. Description 4540074 Control unit with cable set...
  • Page 20 Table 6: Hand button Qty. Part No. Description 4540076 Hand button Table 7: Tank Item Qty. Part No. Description 4550051 Tank 4530112 Floating Valve 4550057 Elbow hose 4560079 Outlet bend with sealing 602622 Hose clamp 600401 Hexagon bolt 600204 Washer...
  • Page 21 Table 8: Floating valve Item Qty. Part No. Description 4530112 Floating valve Flow limiter Mudguard...
  • Page 22 Table 9: Pump connection piece Item Qty. Part No. Description 4560067 Quick release coupling 602622 Pipe clamp 4550505 Hose with cloth Sealing...
  • Page 23 Table 10: Connection rim flushing Item Qty. Part No. Description 4560066 Quick release coupling 4560048 Elbow union with sealing 4560149 Pipe union 4550504 Hose with cloth 602615 Hose clamp...
  • Page 24 Table 11: Drainage tubing Item Qty. Part No. Description 4640030 Drainage Tubing for Slop Hopper 4640031 Drainage Tubing for Slop Hopper (floor version) 4550046 Rapid Binder Normaconnect 4640102 Stainless Steel Pipe Table 12: Slop Hopper Conductor Item Qty. Part No. Description 4520016 Grate...
  • Page 25 Table 13: Sieve insert Item Qty. Part No. Description 4640089 Sieve insert...
  • Page 26 Table 14: Cover for Slop Hopper with hinges Item Qty. Part No. Description 89110040 Cover for Slop Hopper with hinges...
  • Page 27 AUSTRALIA FRANCE POLSKA Arjo Australia Arjo SAS Arjo Polska Sp. z o.o. Building B, Level 3 2 Avenue Alcide de Gasperi ul. Ks Piotra Wawrzyniaka 2 11 Talavera Road CS 70133 PL-62-052 KOMORNIKI (Pozna ) Macquarie Park, NSW, 2113, FR-59436 RONCQ CEDEX...
  • Page 28 At Arjo, we are committed to improving the everyday lives of people affected by reduced mobility and age-related health challenges. With products and solutions that ensure ergonomic patient handling, personal hygiene, disinfection, diagnostics, and the effective prevention of pressure ulcers and venous thromboembolism, we help professionals across care environments to continually raise...