SAFETY WARNING Safety Regulations This machine has been designed with a number of built-in safety devices. To avoid injury, it is highly important not to bypass or disable these safety devices. Operators and maintenance personnel must undergo safety training for the machine. All personnel who handle chemicals for washing and disinfection must understand the washing process, possible health hazards and ways of detecting leaks of toxic chemicals.
Page 5
“I” and “O” to identify the position of the switch. Environmental Protection Incorrect disposal of this equipment and its component parts, particularly batteries or other electrical components, may produce substances that are hazardous to the environment. To minimise these hazards, contact Arjo for information on correct disposal.
If you have any difficulties using the Slop Hopper contact your local Arjo sales office, listed at the end of this manual. Intended Use The is designed to flush away liquids and...
Cleaning The should be cleaned and disinfected Slop Hopper regularly by trained personnel. Most commercially Arjo available cleaning products can be used except chlorine based products. Caution Never use products containing chlorine, abrasives or steel wool for the cleaning since such products will lead to corrosion marks and scratches on the stainless steel.
Components The Slop Hopper mainly consists of the following components. Conductor Label - attention high tension Connection rim flushing Label - rim flushing Sieve insert Plate - type Drainage tubing Hand button Control unit with cable set Pump Rim rinsing connection piece Water tank Flexible hose The ventilation openings in the base frame must not...
2. Installation and Setup Unpacking Remove the packaging. The folder containing documentation must accompany when it is delivered. Slop Hopper Before installing the , check it for Slop Hopper damage in transit. Installation Carry out the installation in the following order: 1.
Page 10
• Connect the to protective grounding Slop Hopper and to the supply voltage stated on the type plate. Check that the connection has suitable overcurrent protection. The correct fuse rating is stated on the type plate. • All cables must be classified for at least 90°C. 8.
3. Maintenance Servicing should be serviced by trained Slop Hopper Arjo personnel. Cleaning the Water The water filter is located at the inlet of the floating valve. Clean the water filter as follows: Filter • Turn off the water supply by using the upstream arranged shutoff unit •...
• Replacement is the reverse of the above procedure. Clamp Siphon trap Care The has to be cleaned and disinfected Slop Hopper regularly. For removing spots, use commercially available cleaning products for chrome steel. Caution Never use products containing chlorine, abrasives or steel wool for the cleaning since such products will lead to corrosion marks and scratches on the stainless steel.
4. Trouble Shooting Problem Cause Action Rim rinsing does not work. • Power supply interrupted. • Contact your local Arjo • Pump and/or control aftersales service office. system defective. Pump does not convey • Water supply turned off. • Open upstream shutoff •...
Page 26
Table 14: Cover for Slop Hopper with hinges Item Qty. Part No. Description 89110040 Cover for Slop Hopper with hinges...
Page 27
AUSTRALIA FRANCE POLSKA Arjo Australia Arjo SAS Arjo Polska Sp. z o.o. Building B, Level 3 2 Avenue Alcide de Gasperi ul. Ks Piotra Wawrzyniaka 2 11 Talavera Road CS 70133 PL-62-052 KOMORNIKI (Pozna ) Macquarie Park, NSW, 2113, FR-59436 RONCQ CEDEX...
Page 28
At Arjo, we are committed to improving the everyday lives of people affected by reduced mobility and age-related health challenges. With products and solutions that ensure ergonomic patient handling, personal hygiene, disinfection, diagnostics, and the effective prevention of pressure ulcers and venous thromboembolism, we help professionals across care environments to continually raise...
Need help?
Do you have a question about the AGSP and is the answer not in the manual?