Do you have a question about the ISOTRON 4416C and is the answer not in the manual?
Questions and answers
Summary of Contents for ENDEVCO ISOTRON 4416C
Page 1
® Instruction Manual 4416C ISOTRON ® Signal Conditioner REV: A IM4416C, Page 1...
Page 2
® SAFETY CONSIDERATIONS This equipment has been designed and tested in accordance with the following standards: EN61326-1:2013 Electrical equipment for measurement, control and laboratory use – Group1, Class B (Emissions) EN61326-1:2013 Electrical equipment for measurement, control and laboratory use – Industrial Environment (Immunity) CFR 47, Class A Code of Federal Regulations: Pt 15, Subpart B...
® 1. Basic Information The Model 4416C ISOTRON® Signal Conditioner is a portable/desk top low-noise signal Conditioner for use with integral electronics piezoelectric (IEPE) transducers or piezoelectric (PE) transducers when used with a remote charge converter (RCC). The unit provides the two wire constant current supply to the transducer/remote charge converter and signal amplification/filtering.
® either the input is open circuit, i.e. no transducer connected, or if the input is short circuit, i.e. a cable fault. During turn on, both green and red conditions are cycled before the steady state is reached. Output 2.2.1 Connection The output connector is a standard BNC located on the front panel with one side connected to signal ground.
® 2.3.1 Turn On The conditioner is turned on and off by a switch on the rear panel. To turn on, press the “1” and to turn off, press “0”. When turned on the Conditioner will cycle through the front panel indicators lighting the gain indicators in turn and lighting the filter indicator.
® Output Con nection Measuring Instrumentation, i.e. data acquisition, data logger etc., should be connected to the output socket using coaxial cables. Load impedance should be no lower than 100kΩ to meet all specifications. Ideally, the 4416C should be placed near the measuring instrumentation to minimize the effect of cable capacitance loading.
® Once 100% capacity has been achieved the charge indicator will extinguish. In the event of a fault with the battery charging circuit, or the battery pack itself, the charge indicator will be “Red”. In this event, please contact your local representative to arrange for the repair of the Conditioner.
® • Select a gain of 10. Verify the signal on the oscilloscope is undistorted and approximately 1V pk in amplitude. • Select a gain of 100. Verify the signal on the oscilloscope is undistorted and approximately 10V pk in amplitude. •...
Page 11
® board. A gasket goes with each panel. Place the entire circuit board assembly on a static free surface, to prevent potential static damage to components. See Photo 5. Unplug the battery from the socket on the circuit board. See photo 5. Locate the new battery and connect it to the socket on the circuit board.
Page 13
® Photo 6 5. Options and Accessories 5.1 Included Accessories QSG4416C Quick Start Guide included IM4416C Instruction Manual (Download from Endevco.com website) EHM2107 Universal power supply, with adaptors for USA, UK, EURO, Japan and Australia 5.2 Options EJ21 10-32 to BNC adapter...