Page 3
Terminal settings (TERMINAL)............22 Configuring interfaces (COMMUNICATION).......... 23 Diagnosis and printing out of the menu settings (DIAGNOS)....26 Interface description................27 OHAUS interface commands ............. 27 SICS interface commands..............29 TOLEDO Continuous mode ............... 32 Event and error messages ...............34 Technical data and accessories............36 Technical data ................
Page 18
INST.PRN * Options not Available at this time...
Page 23
Manual Data Output (Stable and Unstable) to the printer with print. INST.PRN...
Designation Order Number In Use Cover for RD..R.. 21203719 In Use Cover for RD..L.. 21203720 RS232 Cable for PC 80500525 RS232 Cable for Second Scale 80500526 Anti-theft Device 80850000 Carrying Case for RD..R.. 80850083 Carrying Case for RD..L.. 80850084 PRT CABLE,CBM910,MB35/45,RC,RD,RP,TXXP 80252571 PRT CABLE,STP103,MB35/45,RC,RD,RP,TXXP 80252581...
Ranger RD Appendix Legal For Trade 7.5.1 Calibration Perform user calibration as explained in section 3.3. 7.5.2 Technician Menu The menu of your Ranger includes special menu items dedicated to the service technician that must be accessed for Legal For Trade setup.
Page 47
Ranger RD Appendix Technician Menu Overview Scale NO APPr Select appropriate setting MEtrOLO OIML (after all other settings are complete) ntEP Scale1 Scale2 Select Scale1 rAMP rNP xx % SCAL.bLd SCAL.tyP SINGLE.r Select from table 2MULt.IN 2MULt.rN 3MULt.IN 3MULt.rN bAS.UNIt...
Page 49
Ranger RD Appendix The following tables are Ohaus suggested settings for OIML and NTEP approval. Check with local weights and measures officials for ctual requirements. Recommended Settings for Ranger Approved Scales NTEP - MFR RD3RM RD6RM RD15LM RD35LM SCAL.bLd SCAL.tyP SINGLE.R...
Page 50
Appendix Ranger RD OIML - MFR RD3SM RD6SM RD6RM RD15LM RD35LM SCAL.bLd SCAL.tyP 2NULt.RN 2NULt.RN 2NULt.RN 2NULt.RN 2NULt.RN bAS.UNIT VEr.iNt e = d e = d e = d e = d e = d SCL.CAP1 000000.6 000001.2 000001.2 000003.5 000007.0...
Page 51
Ranger RD Appendix 7.5.4 Verification Before this product can be used in a trade approved application, it must be inspected in accordance with local weights and measures regulations. It is the responsibility of the purchaser to ensure that all pertinent legal requirements are met. Please contact your local weights and measures office for further details.