Contents Before first use Positioning of freezer Transportation and moving of freezer Energy saving tips Important information Product information Product details Controller overview Viewing and resetting minimum and maximum temperatures Changing the factory set point Changing the temperature scale Changing the remote alarm contacts Alarms Remote alarm contacts Maintenance and Servicing...
Thank you for purchasing this Lec Medical product. Before you first use this laboratory freezer please carry out the following actions: Check that the freezer has not been damaged in any way during transportation. If any damage is found it must be reported to our customer services department immediately on 0844-815-3755 ...
Laboratory and is designed to store contents between -18°C and below. Please see below for specific product information: Model Height Width Depth Weight Refrigerant Ambient/Room Door (mm) (mm) (mm) (Kg) type working Type temperature LSF151 +10°C Solid R600a LSF232 1565 Solid +32°C...
Product details 1. Lock 2. Electronic Controller 3. Antimicrobial handle 4. Wire Shelf 5. Wire Basket 6. Adjustable foot (2 at front of appliance) There are also 2 roller castors at the rear of the appliance to ease product movement. Additional or replacement parts are available from our spare parts department.
Changing the factory set point (the set point is the point at which the freezer normally holds the temperature at; in some circumstances such as high/low room temperatures it may be necessary to adjust this slightly) Hold button(s) Function Press Example screen down for To change the set point...
Changing the remote alarm contacts Hold button(s) Function Press Example screen down for To change the alarm contact state 3 seconds followed by Three times followed by 1 second (current state) followed by Once followed by 1 second Note: NO = Normally Open, NC = Normally Closed...
Remote alarm contacts All Lec Medical laboratory freezers include remote alarm contacts and these terminal contacts enable the user to connect to a remote control station; these contacts can be found on the rear of the cabinet.
Page 11
In a normal condition (i.e. no alarm) the relay OPENS the contact for the remote alarm facility. In an alarm condition the relay CLOSES the contact for the remote alarm facility.
Maintenance and Service of your Lec Medical Freezer Defrosting All laboratory freezers require manual defrosting; before manually defrosting the cabinet, make sure the power lead is disconnected from the mains. Any loose frost found on the evaporator can be removed carefully using a wooden or plastic scrapper.
Reversing the door In order to reverse the door please follow the below instructions. This should be carried out by 2 people for safety purposes. 1. Tilt the freezer backwards by 45° and lean on a vertical, hard surface such as a wall. 2.
8. Insert the door into the top, left hand hinge (figure 8). 9. Install the bottom, left hand hinge (figure 9) and screw in the feet. Figure 8 Figure 9 10. Add the feet and base plate to the bottom, right hand side. 11.
Electrical Connection The appliance is intended for connection to alternating current. The connection values for voltage (V) and frequency (Hz) are declared on the name plate in the cabinet. Power must be connected via a wall socket with a switch. It is strongly recommended that the wall socket is easily accessible.
Disposal of your freezer When disposing of your freezer do not take it to regular municipal waste containers. Instead, take it to an electrical/electronic waste recycling and re-use centre. This freezer contains insulation gases and refrigerant and must be disposed of in the appropriate manner.
Troubleshooting PROBLEM REASON SOLUTION 1. Make sure the plug is properly inserted into the socket A gap in the electric system 2. Make sure the socket is not faulty The appliance does circuit 3. Make sure the power supply cord is not not work damaged The compressor hardly ever...