R e s p i r a t o r F i t T e s t i n g Model 8026 Particle Generator Operation and Service Manual 1980319, Revision E August 2005...
Page 3
Model 8026 Particle Generator Operation and Service Manual 1980319, Revision E August 2005 TSI Incorporated E-mail address: 500 Cardigan Road health.safety@tsi.com Shoreview, MN 55126 Web site: http://www.tsi.com U.S. INTERNATIONAL Sales and Customer Service: Sales and Customer Service: (800) 874-2811 /(651) 490-2811...
1. UNPACKING AND PARTS IDENTIFICATION ......1 Unpacking the Particle Generator..........1 Parts Identification for the Particle Generator......2 Replacement Parts and Supplies..........2 2. SETTING UP AND OPERATING THE MODEL 8026 PARTICLE GENERATOR ............3 Introduction ................3 Selecting a Fit Testing Location..........3 Setting up the Model 8026 Particle Generator......
Page 6
AVAILABLE APPLICATION NOTES To obtain any of the Application Notes, contact TSI at: U.S. (800) 874-2811/(651) 490-2811, Fax: (651) 490-3824 International (001 651) 490-2811, Fax: (001 651) 490-3824 These Application Notes can also be found under TSI’s web site: http://www.tsi.com...
SAFETY INFORMATION The Model 8026 Particle Generator includes a high-temperature safety switch. If the generator overheats, this switch shuts off the unit permanently, and it must be returned to TSI for servicing.
2918035 Model 8026 Particle Generator 1980319 Operation and Service Manual Note: The Model 8026 Particle Generator is also sold as an accessory to the Model 8095 N95-Companion to the P In this case, ORTA OUNT. the items listed in Table 1-1 are also listed in the N95-Companion...
Parts Identification for the Particle Generator Figure 1-1 identifies the parts of the Model 8026 Particle Generator. Please become familiar with these components before proceeding. Particle Generator Power Cord Reservoir Figure 1-1 Replacement Parts and Supplies The following items are sold by TSI as replacement parts and supplies for the Model 8026 Particle Generator.
Some general notes on alternate uses for the Model 8026 are found at the end of this chapter. Before setting up the Model 8026 Particle Generator, it is important to select a suitable location for fit testing.
Setting up the Model 8026 Particle Generator The particle generator supplements the naturally occurring particles in the room environment. Follow the steps below to set up the particle generator: 1. Fill the reservoir jar with clean tap water to the fill line. It is helpful use hot water to help the salt tablet dissolve more quickly.
Operating the Model 8026 Particle Generator The Model 8026 particle generator operates without intervention once you turn on the power. However, you can adjust the output to increase particle concentration.
Page 14
Adjusting the Particle Generator Output The Model 8026 is set in the factory to a normal operating pressure. This will produce an aerosol output concentration that is appropriate for most environments. If higher counts are needed, you can increase the output by turning the output adjustment screw 1/2 turn clockwise.
Chapter 3 Maintenance Routine Maintenance The Model 8026 Particle Generator requires very little maintenance. Cleanliness is the single most important maintenance requirement. To ensure trouble-free operation of the generator, do the following: • Use compressed air to blow out accumulated salt from inside the generator case.
Page 16
Retainer Figure 3-1 3. Use mild soap and warm water to clean the jet assembly parts. Examine the orifice for plugging. Rinse a second time if necessary. Either use clean compressed air to dry the parts, or air dry them. Note: The orifice is easily damaged.
Chapter 4 Troubleshooting This chapter describes the symptoms, possible causes and recommended solutions for common problems encountered with the Model 8026 Particle Generator. Some common problems you may encounter include: Symptom Possible Cause Corrective Action Low particle The particle generator is Turn on the power to the not running.
Appendix A Specifications The Model 8026 Particle Generator specifications are as follows. Specifications are subject to change without notice. 10 in. x 4 in. x 5.25 in. Size ......... (25.4 cm x 10.1 cm x 13.3 cm) 2.75 lb (1.25 kg) Weight........