PHYSICIAN SHORT INSTRUCTION
BUTTONS
PUmP ACTIVATION
• The screen will show all symbols.
• Syringe size selection.
/
• Pusher self-adjust.
• Pump switches OFF.
ReVeRSING OF THe PUmP PISTON
BUTTONS
NUmBeR OF INFUSIONS deLIVeRed
and
• Reversing of the pump piston to the start position.
depress simultaneously
• Reading the number of infusions delivered.
depressed for 7 seconds
• Reset the number of boluses delivered (L0).
and
depress simultaneously
Confirm the reset of boluses delivered.
keY-PAd LOCk-OUT
• Press for 7 seconds.
• Key-pad lock level selection.
press without releasing
button
/
• Selection of the lock level.
• Pump switches ON.
SCReeN
10
10
20
20
10
10
10
20
20
20
10
20
10
20
20
SCReeN
10
20
20
10
20
10
20
20
20
10
20
20
20
10
10
20
20
20
20
PUmP ON
BUTTONS
SeTTINGS
Operations feasible only with key-pad unlocked.
• Setting flow rate
1 st depressing
• Setting bolus dose
2 nd depressing
• Setting bolus time limitations
3 rd depressing
• Setting bolus number limitations
4 th depressing
• Setting partial volume
5 th depressing
/
• Decrease/Increase above parameters.
BUTTONS
PRIme
Priming possible: only at start of a new infusion or start of a new partial infusion.
• Access to PRIME phase.
• PRIME
(each depressing administers 0.5 ml - max 1.5 ml).
• PRIME phase finish, infusion start.
and
• PRIME interruption (if in execution).
depress simultaneously
and
• Device switch OFF (if PRIME carried out).
depress simultaneously
In case there is a need to prime one or more times during an already
started infusion, the clinician bolus can be used
SCReeN
10
20
20
1 0
20
20
10
20
20
10
20
20
1 0
20
20
10
20
SCReeN
10
20
20
10
20
20
1 0
20
20
10
20
20