  • Page 1 Pumps ® v1.5 C L I N I C I A N M A N U A L ® MORE THAN PAIN RELIEF... SUPERIOR PAIN CONTROL IP22 IP22...
  • Page 2 Become familiar with the ambIT ® pump Solution container Cassette Pump display RUN/PAUSE button Tubing BOLUS clamp button Tubing clamp ON/OFF switch ambIT ® Pump Remote BOLUS switch (optional) To solution container (Side view of cassette) Cassette shaft To patient...
  • Page 3: Table Of Contents

    1.1.2 D efinition of Symbols....................2 1.2 Warnings ..........................6 1.2.1 Sterile, Disposable (Single-Use) Administration Set ..........7 1.2.2 Protection From Air Infusion ..................8 1.2.3 Protection From Unintended Bolus ................8 1.2.4 Use of ambIT Pumps in MRI Environment ..........8 ® v1.5 1.3 Indications for Use.........................8 1.4 Product Description ......................9...
  • Page 4 SECTION 6 - PATIENT LOCKOUT ....................50 6.1 Accessing Lockout Mode....................50 6.2 To Lock the Pump........................50 6.3 To Unlock the Pump ......................51 SECTION 7 - ALARMS AND SIGNALS ..................52 SECTION 8 - TROUBLESHOOTING .....................54 SECTION 9 - SPECIFICATIONS .....................55 SECTION 10 - DELIVERY RATE ACCURACY ................57 10.1 Data from Volumetric Accuracy Testing .................58 10.2 Factors That May Affect Volumetric Accuracy ...............58 SECTION 11 - GENERAL CARE INSTRUCTIONS ..............63...
  • Page 5: Section 1 - Introduction

    BOLUS button or the optional remote BOLUS switch. A bolus is sometime referred to as a PCA bolus. The bolus has units of milliliters (ml). BOLUS/DOSE FLOW RATE: The flow rate at which a bolus or dose is delivered. It is a higher flow rate than the basal flow rate. In the ambIT pumps, the bolus flow rate ® v1.5 varies from 210 ml/hr to 125 ml/hr, depending on the energy left in the battery. Flow rate accuracy is not affected by the bolus flow rate. See Section 1.4.1 for an explanation of how the...
  • Page 6: D Efinition Of Symbols

    Warnings are found at Section 1.2, as well as throughout this document emphasized with grey shading. Failure to observe a warning message is potentially life threatening. 1.1.2 D efinition of Symbols Consult instructions for use. International symbol meaning “Attention, consult accompanying documents.” IEC symbol for “Type BF Applied Part.” (IEC Classification: Internally powered.) The ambIT pump complies with ES 60601-1:2012, 1st Edition, ® including Amend. 1; CSA C22.2 NO. 60601-1:2014, 3rd Edition; IEC ® 60601-1:2005, including Corr. 2:2007 and Amend. 1:2012; IEC 60601-1- 6:2013, Edition 3.1; IEC 60601-1-8:2012, Edition 2.1; IEC 60601-1-11:2015; 60601-2-24:2012, Edition 2; IEC 62304:2006, 1st Edition; IEC 62366:2007, 1st Edition, and Amend. 1:2014; IEC 60601-1-2:2014, 4th Edition, 2014-...
  • Page 7 Se c t io n 1 • I n tr od u c t io n Indicates the medical device manufacturer, as defined in EU Directives 90/385/EEC, 93/42/EEC and 98/79/EC. Indicates the date after which the medical device is not to be used. Indicates the manufacturer’s batch code so that the batch or lot can be identified.
  • Page 8 Se ct i on 1 • In tr o d uctio n Indicates which tubing connects to To solution container the solution container (side view of cassette) Indicates which tubing connects to the patient Cassette shaft To patient Bolus Program lockout Bolus lockout time in hours:minutes Bolus volume in milliliters Volume in milliliters...
  • Page 9 Se c t io n 1 • I n tr od u c t io n Low/dead battery indicator Alarm indicator ambIT pump display ® Pump power on Pump power off Battery orientation PIb mode PCA mode P+P mode Program lockout code...
  • Page 10: Warnings

    Se ct i on 1 • In tr o d uctio n Infusion history report Pump infusion history cleared 1.2 Warnings Federal (U.S.A.) law restricts this device to sale by or on the order of a physician. Read instructions before use. The pump must be used strictly in accordance with these instructions.
  • Page 11: Sterile, Disposable (Single-Use) Administration Set

    Use of any remote BOLUS switch other than the approved remote BOLUS switch could result in an inadvertent bolus. Do not use any other administration set other than the approved ambIT cassettes. The pump ®...
  • Page 12: Protection From Air Infusion

    MRI environment. The pumps have not been tested for heating, migration, or image artifact in the MRI environment. The safety of the ambIT ® pumps in the MRI environment is unknown. Scanning a patient who has the v1.5...
  • Page 13: Product Description

    Se c t io n 1 • I n tr od u c t io n NOTE: S ummit Medical Products, Inc. does not recommend or endorse any one specific medication to be used with the ambIT pumps. The healthcare provider ® v1.5 is the sole individual who decides upon the prescribed medication, pump programmed parameters, method and location of infusion.
  • Page 14 The goal of the bolus or dose is to infuse the medication at an increased rate — mimicking an injection as much as possible. The purpose of the increased rate is generally to flood a greater area with analgesic medication or to increase the concentration of the medication. The flow rate at which the ambIT pumps infuse a bolus or dose is dependent on the energy state of ® v1.5 the batteries.
  • Page 15 Se c t io n 1 • I n tr od u c t io n Basal Time (Minutes) Figure 1-1 Basal flow rate only (PCA mode) infusion pattern For Figure 1-1, the basal flow rate is 10 ml/hr. The maximum available basal flow rate is 20 ml. Figure 1-2 Bolus only (PCA mode) infusion pattern The settings for Figure 1-2 are a bolus volume of 10 ml and a lockout time of 30 minutes. After a bolus is requested and delivered, a lockout time begins.
  • Page 16 Se ct i on 1 • In tr o d uctio n Figure 1-3 Basal flow rate and bolus (PCA mode) infusion pattern For Figure 1-3, the settings are a basal flow rate of 10 ml/hr, a bolus volume of 10 ml, and a lockout time of 30 minutes NOTE: T he basal flow rate is discontinued while a bolus is being delivered and resumes once the bolus has been completed. Figure 1-4 Dose only (PIb mode) infusion pattern The settings for Figure 1-4 are a dose volume of 25 ml and a dose interval of 90 minutes. An optional dose can be delivered at the start of an infusion by pushing the NOTE: BOLUS button within 60 seconds of the start of an infusion.
  • Page 17 Se c t io n 1 • I n tr od u c t io n Figure 1-5 Dose and bolus (P+P mode) infusion pattern (does not apply to the PIB-IL v1.5 Boluses are available after a lockout time. In the example used in Figure 1-5 and Figure 1-6, the dose volume is 20 ml, the dose interval is 90 minutes, the bolus volume is 10 ml, and the lockout time is 30 minutes.
  • Page 18 Se ct i on 1 • In tr o d uctio n Figure 1-6 Dose and bolus (P+P mode) infusion pattern of multiple dose delays (does not apply to the PIB-IL v1.5 Each time a dose is delayed, it is added to the previous delays. In Figure 1-6, the second dose is delayed by about 10 minutes (first dose delay), from 90 minutes to 100 minutes. This means that the third dose should have occurred at 190 minutes, but due to the second delay of a little over 5 minutes (second dose delay), the third dose occurs at a little over 195 minutes.
  • Page 19 Se c t io n 1 • I n tr od u c t io n Dose interval Dose interval Bolus delivery is stopped and Dose delivery is started because a new interval is started. Undelivered bolus volume is not made up. Lockout Time Lockout Time Lockout Time...
  • Page 20 Se ct i on 1 • In tr o d uctio n NOTE: The total volume is controlled by the IL over the dose interval. This means it is possible for the bolus and dose to be delivered back-to-back (i.e., the bolus at the end of one dose interval and the dose at the start of the next dose interval).
  • Page 21: Ambit ® Pump User Interface

    Se c t io n 1 • I n tr od u c t io n 1.4.2 a mbIT Pump User Interface ® The pump has two buttons: the RUN/PAUSE button and the BOLUS button. The RUN/PAUSE button is Cassette located just below the pump display. The BOLUS button is located just below the RUN/PAUSE button. See figure to the left. Pump display RUN/PAUSE An ON/OFF switch is part of the battery cap. See figure button BOLUS to the left.
  • Page 22: Section 2 - Set Up

    Accessories such as MediBag™ solution containers and carrying pouches may be added as required by the therapy. Contact Summit Medical Products, Inc. to obtain a complete listing of all optional accessories. WARNING: It is recommended that an air elimination filter be used for IV applications to protect them from air embolisms. The ambIT cassette can be ordered with or ® without an integrated air elimination filter.
  • Page 23: Ambit® Cassette

    Se c t io n 1 • I n tr od u c t io n WARNING: The ambIT pump requires the use of an ambIT cassette. Use of cassettes ® ® not manufactured by Summit Medical Products, Inc. may cause the pump to malfunction and may cause an over-infusion or an under-infusion.
  • Page 24 Se ct i on 1 • In tr o d uctio n WARNING: Do not connect the cassette to the patient until the cassette has been primed, the cassette has been connected to the pump, and the pump has been programmed to the desired therapy and checked per hospital protocol.
  • Page 25 Se c t io n 1 • I n tr od u c t io n WARNING: Failure to properly snap the cassette bottom disc into the cassette body may result in incorrect flow rates or free-flow conditions. Placing the cassette on the pump will snap the cassette disc bottom in place, if it is not done prior.
  • Page 26: Attach Cassette To Pump

    Se ct i on 1 • In tr o d uctio n NOTE: T he filter bonded to the ambIT cassette tubing is an air-elimination filter ® (“AEF”). The AEF has two membranes. The larger membrane is a hydrophilic membrane and the smaller membrane is a hydrophobic membrane. The hydrophobic membrane will not allow water to flow through it. However, in conjunction with some other liquids, such as organic liquids, water may be permitted to flow through the hydrophobic membrane.
  • Page 27: Changing Fluid Reservoir

    Se c t io n 1 • I n tr od u c t io n 2.6 C hanging Fluid Reservoir WARNING: Not following your facility’s aseptic procedures to properly change the fluid reservoir may contaminate the fluid path. It is beyond the scope of this manual to provide the aseptic technique training. If you are unsure of how to perform this task, consult your facility or supervisor. WARNING: Do not use a syringe as a fluid reservoir, because under-infusion will occur.
  • Page 28: Battery Replacement

    Se ct i on 1 • In tr o d uctio n Step #2 Remove the cover and insert the batteries according to the illustrations at left. Step #3 Place the battery cover onto the pump as illustrated to the left (OFF symbol (O) on the cover slightly to the left of the ( I ) mark on the pump).
  • Page 29: Pump Power On And Off

    Se c t io n 1 • I n tr od u c t io n 2.8 P ump Power On and Off CAUTION: Always place the pump in pause mode prior to turning the pump off. Failure to do so may cause the therapy to be delayed and/or history to be lost. The pump will not sound an alarm if the pump is not placed in pause mode prior to being turned off.
  • Page 30: Section 3 - Programming Instructions

    Se ct i on 3 • P r o g r amm ing Inst r u c t ions SECTION 3 - PROGRAMMING INSTRUCTIONS 3.1 G eneral Information The ambIT pumps must be programmed and have the history cleared before ® v1.5 administering any medication or fluid. The pumps have two buttons: the RUN/PAUSE...
  • Page 31: Program Options

    Se c t ion 3 • Pr o g ra m m in g In s t ru c ti o n s 3.2 P rogram Options To enter program mode: Step #1 Place the pump in pause mode. If the pump is not on, turn it on (see Section 2.7). Step #2 Press and hold both the RUN/PAUSE button and the BOLUS button for two seconds until one beep is heard.
  • Page 32: Pca Mode

    BOLUS button or the optional remote BOLUS switch. A bolus is sometime referred to as a PCA bolus. The bolus has units of milliliters (ml). • Bolus/dose flow rate is the flow rate at which a bolus or dose is delivered. It is a higher flow rate than the basal flow rate. In the ambIT ® pumps, the bolus flow rate varies from 210 ml/hr to 125 ml/ v1.5 hr, depending on the energy left in the battery. Flow rate accuracy is not affected by the bolus flow rate. See Section 1.4.1 for an explanation...
  • Page 33 Se c t ion 3 • Pr o g ra m m in g In s t ru c ti o n s 3.2.1.2 P CA Mode Programming Steps NOTE: If at any point during programming the user needs to move backward in the display, refer to instructions found at 3.2.1.3.
  • Page 34 Se ct i on 3 • P r o g r amm ing Inst r u c t ions AUDIBLE ACTION VISUAL INDICATOR INDICATOR Set bolus volume: Letter “b,” numbers left of (A) Press and the RUN/PAUSE decimal point, “ml” and button to increase the number or bolus icon flash.
  • Page 35 Se c t ion 3 • Pr o g ra m m in g In s t ru c ti o n s AUDIBLE ACTION VISUAL INDICATOR INDICATOR (D) If during programming, you Each time the BOLUS need to move backwards in the button is pressed, the display to a previous setting, press display moves backward...
  • Page 36 Se ct i on 3 • P r o g r amm ing Inst r u c t ions 3.2.1.3 Moving Backward in Display While Programming The user can move backward in the display. This functions as follows: • Pushing and holding the BOLUS button held until a beep is heard causes the currently blinking value to be accepted (i.e., stored in memory) and the value to the left on the display to start to blink so that it can be changed.
  • Page 37: Pib Mode

    . In this manual, the term “dose” will be used when discussing either PIb or PIEb. The dose has units of milliliters (ml). • The bolus/dose rate is the flow rate at which a bolus or dose is delivered. It is a higher flow rate than the basal flow rate. In the ambIT ® pumps, the bolus flow rate varies from 210 ml/hr to 125 ml/hr, v1.5 depending on the energy left in the battery. Flow rate accuracy is not affected by the bolus/dose flow rate. See Section 1.4.1 for an explanation...
  • Page 38 Se ct i on 3 • P r o g r amm ing Inst r u c t ions 3.2.2.2 P Ib Mode Programming Steps NOTE: If at any point during programming the user needs to move backward in the display, refer to instructions found at 3.2.2.3. AUDIBLE ACTION VISUAL INDICATOR...
  • Page 39 Se c t ion 3 • Pr o g ra m m in g In s t ru c ti o n s AUDIBLE ACTION VISUAL INDICATOR INDICATOR Set dose interval time: “INT” is displayed for two seconds to indicate that the (A) Press and release the RUN/ pump is in dose interval PAUSE button to increase the dose mode.
  • Page 40 Se ct i on 3 • P r o g r amm ing Inst r u c t ions AUDIBLE ACTION VISUAL INDICATOR INDICATOR (C) Repeat steps 5(A) and 5(B) to Volume to be infused is set the digits in the tens and ones established.
  • Page 41: P+P Mode

    BOLUS button or the optional remote BOLUS switch. A bolus is sometime referred to as a PCA bolus. The bolus has units of milliliters (ml). • The bolus/dose flow rate is the flow rate at which a bolus or dose is delivered. It is a higher flow rate than the basal flow rate. In the ambIT ® pumps, the bolus flow rate varies from 210 ml/hr to 125 ml/ v1.5 hr, depending on the energy left in the battery. Flow rate accuracy is not affected by the bolus flow rate. See Section 1.4.1 for an explanation of how...
  • Page 42 Se ct i on 3 • P r o g r amm ing Inst r u c t ions NOTE: When programming P+P mode, if the tens and/or ones digits are already set to the desired setting, press the RUN/PAUSE and the BOLUS buttons at the same time until a beep is heard to accept the setting and move on to the next step. This option is not available when a parameter is programmed to its maximum value.
  • Page 43 Se c t ion 3 • Pr o g ra m m in g In s t ru c ti o n s AUDIBLE ACTION VISUAL INDICATOR INDICATOR Set dose volume: (A) Press and release the RUN/ Letter “d,” number left of PAUSE button to increase the decimal point, and “ml” flash.
  • Page 44 Se ct i on 3 • P r o g r amm ing Inst r u c t ions AUDIBLE ACTION VISUAL INDICATOR INDICATOR Set bolus volume: (A) Press and release the RUN/ Letter “b,” numbers left of PAUSE button to increase the decimal point, “ml”...
  • Page 45 Se c t ion 3 • Pr o g ra m m in g In s t ru c ti o n s AUDIBLE ACTION VISUAL INDICATOR INDICATOR (D) If during programming, you Each time the BOLUS button need to move backwards in the is pressed, the display moves display to a previous setting, press backward one screen to enable...
  • Page 46: 3.3 Program Review

    Se ct i on 3 • P r o g r amm ing Inst r u c t ions • If the leftmost value is blinking and the BOLUS button is pushed and held until a beep is heard, then the value is accepted (i.e., stored in memory) and the display changes the preceding parameter. For example if the left most value of the bolus lockout time (hours) is blinking and the BOLUS button is pushed and held held until a beep is heard, the hours value is accepted and the display changes to the previous parameter (bolus volume) and...
  • Page 47: Section 4 - Operating Instructions

    S ec t io n 4 • Op e ra t in g In s t ru c ti o ns SECTION 4 - OPERATING INSTRUCTIONS NOTE: All operating instructions are the same for the three modes of the ambIT ® pumps except “Start Infusion” (Section 4.1). Please read respective sections v1.5...
  • Page 48: P+P Mode Start Infusion

    Se ct i on 4 • Op er at in g In stru ct ion s If the BOLUS button is pushed within 60 seconds of placing the pump in run mode after changing the settings or clearing the history, the dose is delivered immediately instead of waiting for the dose interval to elapse.
  • Page 49: Bolus Activation

    S ec t io n 4 • Op e ra t in g In s t ru c ti o ns NOTE: In the case of an OCC alarm due to downstream pressure (occlusion alarm), the pump will display a visual alarm initially.
  • Page 50: Summary Of Operating Controls

    Se ct i on 4 • Op er at in g In stru ct ion s 4.6 S ummary of Operating Controls AUDIBLE ACTION STEPS TO TAKE VISUAL INDICATOR INDICATOR Review and verify While the pump is in One beep “RX” - display cycles through program setttings pause mode, press and each programmed parameter...
  • Page 51: Section 5 - Infusion History Reports

    S e c ti on 5 • I nfu s i o n H is t o ry R e p o rts SECTION 5 - INFUSION HISTORY REPORTS 5.1 P ump Infusion History To obtain the pump infusion history: Step #1 Place the pump in pause mode. Step #2 Press and hold the BOLUS button for two seconds until one beep is heard and “HX”...
  • Page 52: Pib Mode Infusion History

    Se ct i on 5 • I n f u sio n His tory Re po rt s 5.1.2 P Ib Mode Infusion History INFUSION HISTORY VISUAL INDICATOR Current delivery mode Total volume infused in ml “ml” icon Total number of doses delivered No.
  • Page 53: Clearing Pump Infusion History

    S e c ti on 5 • I nfu s i o n H is t o ry R e p o rts 5.2 C learing Pump Infusion History. CAUTION: Do not clear the infusion history unless the pump is being used for a new patient, as doing so will allow a bolus or dose immediately, resulting in over- infusion.
  • Page 54: Section 6 - Patient Lockout

    Se ct i on 6 • P at ien t L ocko ut SECTION 6 - PATIENT LOCKOUT The pump has been designed with patient lockout features: WARNING: The steps required for entering the pump’s program mode should not be revealed to the patient. This prevents unauthorized tampering with the infusion parameters, and prevents the infusion history from being cleared.
  • Page 55: To Unlock The Pump

    Se c t io n 6 • P a ti e n t L o ck o u t NOTE: The clinician may program one, two, three, or four digits into the pump for the lock code. If programming only one, two, or three digits into the pump for the lock code, when the desired code is programmed (e.g., 12 or 123), press and hold both the RUN/PAUSE button and the BOLUS button for two seconds until one beep is heard.
  • Page 56: Section 7 - Alarms And Signals

    Se ct i on 7 • Ala r m s and Sign als SECTION 7 - ALARMS AND SIGNALS VISUAL AUDIBLE PRIORITY STATUS ICON COMMENTS INDICATOR INDICATOR Green run light (inside the BOLUS Pump is Periodic movement button) blinks, infusing None of the cassette gears is “ml”...
  • Page 57 Se c t ion 7 • A l a rm s a n d S ig n a l s VISUAL AUDIBLE PRIORITY STATUS ICON COMMENTS INDICATOR INDICATOR One long tone followed The “infusion complete” Infusion Bag icon flashes in by three alarm will sound every High complete the display...
  • Page 58: Section 8 - Troubleshooting

    Se ct i on 8 • Tr o ub lesh oo ting SECTION 8 - TROUBLESHOOTING Problem Resolution Reference • Verify that all tubing clamps are open. • If a spike cassette is used, verify that the spike is Cassette will not prime completely inserted into the solution container.
  • Page 59: Section 9 - Specifications

    Se c t io n 9 • S p e c if ic a t io n s SECTION 9 - SPECIFICATIONS Specification Details Stroke volume bolus 50 microliters **Bolus/dose delivery rate 210 ml/hr Volumetric accuracy +/- 6% Interval limit (IL) 0 to 50 ml in 1 ml steps Basal flow rates 0 to 20 ml/hr in 0.1 ml/hr steps...
  • Page 60 Se ct i on 9 • S p ecif icat ions Specification Details RUN/PAUSE button BOLUS button Operating controls Remote BOLUS switch (optional) ON/OFF twist cap Volume infused Doses delivered History reports Boluses delivered Boluses requested Elapsed time PIb+PCA Delivery profiles Approximately 2 years with an infusion rate of one liter Estimated life span of pump per month.
  • Page 61: Section 10 - Delivery Rate Accuracy

    S e c ti on 1 0 • D e li v e ry R a te A cc u r ac y SECTION 10 - DELIVERY RATE ACCURACY The ambIT pumps are pseudo-continuous, positive displacement pumps. This means ®...
  • Page 62: Data From Volumetric Accuracy Testing

    Based on the volumetric accuracy testing carried out, as described above, the ambIT ® v1.5 pumps met the volumetric accuracy specifications for flow rate and bolus volumes as shown in the table found at Section 10.1.
  • Page 63 The dose and bolus accuracy is shown at different volumes over a number of sequential bolus/dose events. The basal flow rate accuracy is shown as trumpet curves under different conditions of flow rate and delivery pressures. These graphs are applicable for all ambIT pump cassettes. ® The ambIT pumps use a rotary peristaltic mechanism to deliver fluid. Each rotation ® v1.5 of the peristaltic mechanism delivers 0.050 ml (50 μl) of fluid. During bolus/dose delivery, the pump delivers a prescribed volume at a flow rate of 210 ml/hr accuracy. The accuracy is shown as the percent error for the actual vs. prescribed bolus/dose volume. This error is measured sequentially over a number of bolus/dose events at a minimum bolus/dose volume (0.1 ml) and...
  • Page 64 Se ct i on 1 0 • D elive ry Ra te Acc ur a c y 20 ml/hr flow rate 25.0 20.0 15.0 10.0 -5.0 EP Max EP Min -10.0 -15.0 Observation window (min) 20 ml/hr + 100 mmHg back pressure 45.0 40.0 35.0 30.0 25.0 20.0 15.0 10.0 EP Max EP Min -5.0 0...
  • Page 65 S e c ti on 1 0 • D e li v e ry R a te A cc u r ac y 20 ml/hr - 100 mmHg back pressure 45.0 40.0 35.0 30.0 25.0 20.0 15.0 10.0 EP Max EP Min -5.0 Observation window (min) Supply container below pump 20 ml/hr -2.0 -4.0 -6.0...
  • Page 66 Se ct i on 1 0 • D elive ry Ra te Acc ur a c y 25 0.1 ml bolus/dose volumes 5.0% 4.5% 4.0% 3.5% 3.0% 2.5% 2.0% 1.5% 1.0% 0.5% 0.0% 10 11 12 13 14 15 16 17 18 1 9 20 21 22 23 24 25 Bolus number 25 20 ml bolus/dose volumes 3.5%...
  • Page 67: Section 11 - General Care Instructions

    11.1 W arranty Information Contact your local sales representative for warranty and extended warranty lengths. This warranty will not apply to ambIT pumps that have been, in the judgment of Summit ® Medical Products, Inc., damaged in whole or in part due to misuse, abuse, negligence, alteration or improper installation, or that have been dropped or used in a manner inconsistent with their labeling and packaging.
  • Page 68: Cleaning And Disinfecting Instructions

    Se ct i on 11 • Gen er al Ca re Ins t r uc t io ns 11.2 C leaning and Disinfecting Instructions Step #1 Dampen a clean rag or paper towel with any household cleaners such as: • A fresh solution of one (1) part household bleach to nine (9) parts water; • Rubbing alcohol (70% Isopropyl alcohol); •...
  • Page 69 Se c t ion 1 2 • El e c tr oma g n e ti c I m mu n i ty (E M C ) SECTION 12 - ELECTROMAGNETIC IMMUNITY (EMC) Mobile RF communications equipment can affect the operation of the ambIT pump. ®...
  • Page 70 Se ct i on 1 2 • E lec tr om agn etic Im m unit y ( EM C) Table 12-1. Electromagnetic immunity levels tested and passed by ambIT pump. ® Phenomenon Basic EMC Standard Or Test Immunity Test Levels Method ±2, 4, 6, 8 kV contact Electrostatic discharge IEC 61000-4-2 ±2, 4, 8, 15 kV air...
  • Page 71: Section 13 - Customer Assistance

    Se c t ion 1 3 • C u s t om e r A s s is t a nc e SECTION 13 - CUSTOMER ASSISTANCE For customer assistance, please contact your ambIT distributor, or Summit Medical Products, ®...
  • Page 72 Sect ion 1 3 • C ustomer Assis ta nc e ambIT Infusion Pump is a trademark of: ® Finding better ways to care For people™ 63290.00 (6/20)