Flexible Endoscope Cleaning & Disinfection Guide
RHINO-LARYNGO VIDEOSCOPE
OLYMPUS ENF-VH2/V4
1
Wipe the insertion tube
Gently wipe the external
surfaces of the insertion tube
with a cloth moistened in water.
2
Attach the tester
Attach the leakage tester to the
venting connector on the
endoscope.
6
Clean the surfaces
Immerse the endoscope in
detergent solution. Using a
lint-free cloth, gently wipe all
debris from the endoscope's
external surfaces. Soak in
detergent solution for the
recommended time.
9
Disinfect the endoscope
Immerse the endoscope in
disinfectant solution. Con rm
that there are no air bubbles on
its surface. Soak for the time
speci ed by the manufacturer.
10
Rinse the surfaces
Remove the endoscope from
the disinfectant solution and
immerse in deionized water.
Using a sterile, lint-free cloth,
thoroughly rinse and gently
wipe all external surfaces.
RHINO-LARYNGO FIBERSCOPE
OLYMPUS ENF-GP2
• This guide is only a summary of the steps necessary to properly reprocess your
endoscope. Follow the detailed steps outlined in the ENDOSCOPE REPROCESSING
MANUAL that was included with your endoscope.
• Use detergent solution and disinfectant solution which feature validated processes in
accordance with national and local regulations and/or guidelines.
3
Apply pressure
When using MB-155 with MU-1
When using
or the light source
WA23080A
or
Connect MB-155 to the
Close the pressure release
maintenance unit (MU-1) or the
lever. Press the hand pump so
light source and turn the power
that the pointer stays within the
switch on.
green area.
7
8
Rinse the endoscope
Immerse the endoscope in
Remove the endoscope from
water.
water. Use a soft, lint-free cloth
to remove excess moisture
from the endoscope and
cleaning accessories in
preparation for disinfection.
11
Wipe the surfaces
Using a sterile, lint-free cloth, thoroughly and gently wipe the
external surfaces of the endoscope.
4
Observe
Immerse the entire endoscope in
water, observe for 30 seconds or
more while angulating the bending
section and con rm that there is
no continuous series of bubbles.
CAUTION: If a leak is detected, remove
from water and contact Olympus.
Wipe the surfaces
5
Detach the tester
Decompression
Remove the endoscope from the
water and release the air pressure in
the endoscope.*
When using WA23080A, open the
*
pressure release lever.
When using MB-155, turn the
*
maintenance unit or light source off.
Detach the leakage tester from the
air source and then the endoscope.