User Manual Insu 10 Analog-Digital Insulation Tester Sifam Tinsley Instrumentation Inc. Version No. STIUS/2020/01 3105 Creekside Village Drive, Suite No. 801, Kennesaw, GA 30144 (USA) Contact No. : +1 404 736 4903 E-mail Id : psk@sifamtinsley.com Web : www.sifamtinsley.com...
Page 2
* Make sure that the new fuse makes good contact. Specification : Current Rating : 500 mA / 500V-6.3 mm x 32 mm with a breaking capacity Insu 10 of 10kA. Recommended fuse manufacturer : Siba Part No. : 70-065-63/0.5A Be absolutely sure that only the specified fuses are used !
13 Maintenance (1)Liquid Crystal Display (10) Symbols for displaying the (2)On/Off push button selected functions Attention ! (3)Push button (11) Display for selected function Disconnect the instrument from the measuring circuit before opening (4)Push button (12) Display for the unit of the instrument to replace the battery or the fuse ! (5)TEST Push button measured quantity...
You have selected an instrument which provides you with a high level of may be used) safety. The Insu 10 Analog-Digital Insulation Tester is manufactured in Service Life Typically 2500 x 5 second operations with compliance with safety regulations IEC 61010-1/EN 61010-1/VDE 0413 alkaline manganese cells part 1.
Repair, Parts Replacement and Calibration After opening the instrument, live parts may be exposed. Therefore, the instrument must be disconnected from the measuring circuit prior to opening its case for repair, replacement of parts or calibration. If repair or calibration cannot be avoided, unless the instrument is open and live, the 1300 work must be performed by a qualified person, who understands the 1000 V Range...
11.1 Description of Menu Items : 5. Insulation Resistance Measurement (M Ω 11.1.1 Setting the GoNoGo values : 5.1. Preparation for measurement (at all ranges except V ·Select the required "Test Voltage (50V/100V/250V/500V/1000V)" by changing the "Function Selector Switch". Operating Order : ·Connect the test probes firmly to the circuit under test.
Pre-selectable limit checks (Go/Nogo option) Storing the measured resistance value Minimum and Maximum value “MIN/MAX” storage xxxx xxxx = Time Menu structure Insu 10 = Recall Confirmation of selected menu item via ENTER key = Minutes Select next menu item = Seconds in the flow direction...
After 10 values are stored for the selected range the 11 value when stored will over-write the first value i.e. at any given point Insu 10 of time only ten readings per range can be stored. The readings are stored in a volatile memory. Therefore when the 2 9 2 instrument is switched off, the readings are lost.
Page 10
Continuity Testing and Resistance Measurement Continuity Test Attention ! Be absolutely certain that the device under test is Insu 10 voltage-free. 0 0 0 Continuity Testing Verify that device under test is electrically dead. Set the function selector switch to “...