Page 1
Transcutaneous Electrical Nerve Stimulation (TENS) Model 2000-T Operation Manual Read this manual before operating your 2000-T. Save this manual for future use. The most current version of this manual can be found online at www.grahamfield.com. 2000-T-INS-LAB-RevC14...
Federal law (USA) restricts this device to sale by or on the order of a physician. Indications TENS is indicated to be used under a physician’s prescription for the symptomatic relief of chronic (long term) pain and for the treatment of postoperative or posttraumatic pain. 2000-T-INS-LAB-RevC14...
WARNING: TENS is not effective for pain of central origin (headaches). WARNING: If TENS treatment becomes ineffective or unpleasant, stimulation should be discontinued until evaluated by a physician. WARNING: Avoid adjusting controls while operating machinery or vehicles. WARNING: Always turn the TENS device OFF before applying or removing electrodes. 2000-T-INS-LAB-RevC14...
CAUTION: Isolated cases of skin irritation may occur at the site of electrode placement during long-term application. CAUTION: Effectiveness is highly dependent upon patient selection of a doctor qualified in the management of pain patients. CAUTION: Skin irritation and electrode burns are potential adverse reactions. 2000-T-INS-LAB-RevC14...
Pulse Width Control Mode Selector MODE Pulse Rate Control (uS) (Hz) PULSE WIDTH PULSE RATE Battery Compartment Battery Strip SYSTEM COMPONENTS Your device may include the following components or accessories: • TENS unit • Carrying case • Lead wires • Operation manual • Electrodes Also required (not included): One 9-Volt alkaline battery. 2000-T-INS-LAB-RevC14...
Modulation (M) pulse width decreases the pulse width down to 60% of the original width setting. This decreased pulse width is maintained for 1.5 seconds before returning to the original width setting, which is maintained for 3.5 seconds. The cycle is then repeated. The intensity and pulse rate are adjustable. 2000-T-INS-LAB-RevC14...
FDA. ELECTRODE SELECTION AND CARE Using Electrodes Use the electrodes as prescribed. Follow application procedures outlined in electrode packaging to maintain stimulation and prevent skin irritation. The electrode packaging provides instructions for care, maintenance, and proper storage of electrodes. 2000-T-INS-LAB-RevC14...
• Avoid excessive stretching of the skin when applying electrodes. Proper application is best accomplished by applying the electrode, then smoothly pressing it in place from the center outward. • When removing electrodes, always remove by pulling in the direction of hair growth. • It may be helpful to rub skin lotion on electrode placement area when not wearing electrodes. 2000-T-INS-LAB-RevC14...
Always start with the lowest setting and increase the intensity slowly. If the stimulation levels are uncomfortable or become uncomfortable, reduce the stimulation intensity to a comfortable level; or cease stimulation and contact your physician. 2000-T-INS-LAB-RevC14...
Remove the discharged battery from the device. Place new battery in the compartment. Note: Ensure the proper polarity (+ and –) markings match the markings in the device. Replace the front panel cover. Dispose of the old battery according to local guidelines and regulations. 2000-T-INS-LAB-RevC14...
Wipe lead wires with damp cloth moistened with soap and water. Do not immerse the lead wires. To properly store the device for an extended period of time, remove the battery from the unit. Place the unit and accessories in the carrying case provided and store in a cool, dry location. 2000-T-INS-LAB-RevC14...
If device still is not working, turn the unit off and contact your authorized GF Health Products, Inc. distributor. If there is any other problem, please contact an authorized GF Health Products, Inc. distributor. Do not try to repair a defective device. 2000-T-INS-LAB-RevC14...
2 bursts per Sec Continuous Adj. 0-80 Adj. 40-260 Adj. 2-120 Hz Modulation Adj. 0-80 Modulates Adj. 2-120 Hz 5 Sec total down from time preset width setting by 60% then back to original setting Note: All values have ±20% tolerance. 2000-T-INS-LAB-RevC14...
ENTIRE WARRANTY, EXCLUSIVE REMEDY AND CONSEQUENTIAL DAMAGES DISCLAIMER THIS WARRANTY IS GF’S ONLY WARRANTY AND IS IN LIEU OF ALL OTHER WARRANTIES,EXPRESS OR IMPLIED. GF MAKES NO IMPLIED WARRANTIES OF ANY KIND INCLUDING ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. IF ANY MODEL OR SAMPLE WAS SHOWN TO THE CUSTOMER, SUCH MODEL OR SAMPLE WAS USED MERELY TO ILLUSTRATE THE GENERAL TYPE AND QUALITY OF THE PRODUCT AND NOT TO REPRESENT THAT THE PRODUCT WOULD NECESSARILY CONFORM TO THE MODEL OR SAMPLE IN ALL RESPECTS. GF’S TOTAL LIABILITY FOR ANY PRODUCT OR SERVICE PROVIDED IS LIMITED TO THE COST OF THE PRODUCT GIVING RISE TO THE CLAIM. IN NO EVENT WHETHER IN CONTRACT, INDEMNITY, WARRANTY, NEGLIGENCE, STRICT LIABILITY OR OTHERWISE WILL GF BE LIABLE FOR ANY DIRECT, SPECIAL, INDIRECT, INCIDENTAL, EXEMPLARY OR CONSEQUENTIAL DAMAGES, INCLUDING, BUT NOT LIMITED TO: DAMAGES FOR L OSS OF PROFITS OR INCOME, LOSS OF USE, DOWNTIME, COVER, OR EMPLOYEE OR INDEPENDENT CONTRACTOR WAGES, PAYMENTS AND BENEFITS. This warranty gives you specific legal rights. You may have additional rights which vary from state to state. 2000-T-INS-LAB-RevC14...