SLS Baseline 8900 Setup and Calibration Manual Document Number: SLS Baseline 8900 Setup and Calibration Manual 5-1-GL-GL-SUL-00028 Originators Cody Fillman Rev 1 Dec 2013 Approval Jose Ortiz No. of Pages: 25...
Page 2
Revision Tracking Date Notes Updated By Reviewed By Dec. 2013 Initial Release C. Fillman SLS Baseline 8900 Setup and Calibration Manual 5-1-GL-GL-SUL-00028 Page 1...
Page 3
Experience and other specialized training can complement the materials used in this Operating Manual. C. Weatherford makes no representation as to the accuracy or completeness of the materials in this Operating Manual. All materials are provided “AS IS” WITHOUT WARRANTY OF ANY KIND...
Failure to do so may cause damage to the instrument NOTE Do not power on the Baseline 8900 without the proper support gases connected. Doing so may cause damage to the instrument. SLS Baseline 8900 Setup and Calibration Manual 5-1-GL-GL-SUL-00028 Page 5...
Server) and a data acquisition card installed. Baseline 8900 A FID chromatograph made by Baseline Support Gases Other gases required in order for the detector to operate (Hydrogen and zero air) Page 6 5-1-GL-GL-SUL-00028 SLS Baseline 8900 Setup and Calibration Manual...
4) Flip the power switch on the back of the Baseline 8900 to the “On” position 5) Allow the Baseline unit approximately 20 minutes to warm up. The front screen on the instrument will indicate when the warm up time is over Page 8 5-1-GL-GL-SUL-00028 SLS Baseline 8900 Setup and Calibration Manual...
Connect the Baseline 8900 using null RS-232 serial cables as shown Figure 3: Serial Connection When opening the Skychrom software, it will prompt a username and password. Use: U/N: admin Password: admin Page 10 5-1-GL-GL-SUL-00028 SLS Baseline 8900 Setup and Calibration Manual...
Refer to Figure 4, for information on the purpose of each menu button. Figure 4: Skychrom Menus SLS Baseline 8900 Setup and Calibration Manual 5-1-GL-GL-SUL-00028 Page 11...
Page 13
If the alarm on the instrument is triggered, it may be acknowledged via the software by going to Instrument->Acknowledge Alarm, as seen in Figure 5. Figure 5: Acknowledge Alarm Page 12 5-1-GL-GL-SUL-00028 SLS Baseline 8900 Setup and Calibration Manual...
In Figure 7, The correct operational parameters are shown and, in addition, the captions indicate how to modify the parameters should any changes be required. For higher gas concentrations, a SLS Baseline 8900 Setup and Calibration Manual 5-1-GL-GL-SUL-00028 Page 13...
Page 15
This value may be adjusted if needed for a very low concentration, or in a low gain method, but for the standard medium gain method 300 counts ensures readability for low concentrations while still filtering out extra results caused by noise. Page 14 5-1-GL-GL-SUL-00028 SLS Baseline 8900 Setup and Calibration Manual...
Page 16
In Figure 9, an example of a spline may be seen along with details of how to add one to the method. This is useful for increasing accuracy as well as aiding the noise filter in reducing any unwanted noise peaks from being read in the software. SLS Baseline 8900 Setup and Calibration Manual 5-1-GL-GL-SUL-00028 Page 15...
All sequences must be saved in order to function. In order to do this, the sequence window must remain open in Skychrom and the user must navigate to the “Sequence” at the top menu. Select “Save as” and save the sequence. Figure 11 demonstrates this process. SLS Baseline 8900 Setup and Calibration Manual 5-1-GL-GL-SUL-00028 Page 17...
Page 19
Figure 11: Sequence Save Page 18 5-1-GL-GL-SUL-00028 SLS Baseline 8900 Setup and Calibration Manual...
Figure 12: Calibration Setup After the chromatograph has stopped running, go to “Edit” on the top menu and select “Curr. Method” from the drop down menu as shown in Figure 13. SLS Baseline 8900 Setup and Calibration Manual 5-1-GL-GL-SUL-00028 Page 19...
Page 21
“edit” button to save the change. When creating a method from Page 20 5-1-GL-GL-SUL-00028 SLS Baseline 8900 Setup and Calibration scratch, type in the gas that is being added into the “Calibration Substance” box and define the “R. Tim”. Manual...
Page 22
Start with nC5 and work backwards so as to avoid overlapping time readings which will cause errors in the software. Figure 14: Skychrom Calibration SLS Baseline 8900 Setup and Calibration Manual 5-1-GL-GL-SUL-00028 Page 21...
Peak’s Attribute Fields should have “Concentration” selected on the first drop down menu. Click OK at the bottom right corner of the window in order to save changes that have been made. Refer to Figure 16 for proper settings. Page 22 5-1-GL-GL-SUL-00028 SLS Baseline 8900 Setup and Calibration Manual...
4) Delete the hash tag before the 8900BL line and change the “serX” to the correct serial port (port “A” in Figure 3) 5) Type “WWreboot” to reboot WellWizard so that the changes will take place Page 24 5-1-GL-GL-SUL-00028 SLS Baseline 8900 Setup and Calibration Manual...