Page 1
AG-SELECT B1 Etatron GB Ltd, Newlin Business Park, Exchange Road, Lincoln, LN6 3AB...
Page 2
Pump Connections - AG-Select B1 If using 4-20mA output pump should be connected to J21...
Page 3
Manual - AG-Select B1 J21 = mA 1 output – J20 = mA 2 output...
Page 4
Pump Connections - AG-Select B1 If using 4-20mA output pump should be connected to J21 as shown above...
Page 5
Calibration pH - AG-Select B1 Please ensure you have to hand pH 7 and pH 4 buffer solutions and check use by dates...
Page 6
Calibration pH - AG-Select B1 From main screen select OK and press PLUS button X 3 or until “CAL” is shown Press OK button to go to 1 calibration point pH 7 Place clean pH electrode into the buffer solution,mix slowly and then wait until display stabilises after around 1 minute.
Page 7
Calibration mV - AG-Select B1 Please ensure you have to hand 650mV or 475mV buffer solutions and check use by dates...
Page 8
Calibration mV - AG-Select B1 From main screen select OK and press PLUS button X 3 or until “CAL” is shown Press OK button to go to 1 calibration point 650mV Place clean mV electrode into the buffer solution,mix slowly and then wait until display stabilises after around 1 minute.
Page 9
Factory Reset AG-Select B1 1. Power off instrument 2. Power instrument back up 3. Press OK button within 15 seconds 4. After screen has changed hold down the MINUS and PLUS buttons simultaneously 5. “rESEt” should be shown on screen, if you do not see “SET DEFAULTS”...