1. INTRODUCTION
The medTester 5000B is part of an equipment
management data base and testing system available
from DNI Nevada Inc.(now Fluke Biomedical Corp.). To
use Infutest with medTester, your medTester must
have the medCheck firmware option installed, plus you
must be able to use the equipment data base program
Sentinel, also by DNI Nevada.
2. SETTING UP
Use a Datrend Systems SSC-1 cable to connect
Infutest to COM2 of the medTester. Do not use a
null modem cable or adaptor.
Do not connect medTester to Infutest using the
medTester serial cable supplied by DNI Nevada.
This is a null modem cable for connecting COM1 of the
medTester to a PC.
Power up medTester and select UTIL from the main
menu level, followed by PORTS. Set COM1 to OFF
and COM2 to ON. Return to the UTIL menu.
Select RS232 and then COM2. Program COM2 as
follows:
Set Baud Rate to 9600
Set Stop Bits to 1
Set Length to 8
Set Parity and CTS to OFF
3. DESIGNING CHECKLISTS
A Checklist is a list of instructions or "check items"
which
medTester
performs
automatically, pausing occasionally for data input from
the user via the medTester keyboard, or from an
external device connected to medTester's COM2 port.
Checklists are created on a PC using the Sentinel
program, and then are downloaded from your PC into
medTester's memory.
To control Infutest with medTester, your Checklist
must included check items which begin with \M
followed by an Infutest remote control command. A
check item beginning with \M tells medTester to get
AN-17.2004: Page 1 of 4
Interface With medTester
sequentially
and
1-800-667-6557
©
Datrend Systems Inc.
APPLICATION NOTE
INFUTEST
Series C & Series D
data from COM2, instead of from the medTester
keyboard. In executing a Checklist with \M items, the
medTester will transmit whatever characters follow \M
to the Infutest via COM2, and then will wait for data to
be returned via COM2. Depending on the command
which follows the \M, Infutest may return up to 40
characters of data to the medTester.
\M is also necessary to introduce delays into the
Checklist so the Infutest has time to start the test and
collect some measurement data. The syntax for this
is:
\M pause
where pause is the number of seconds (1 to 32000)
you want medTester to wait until it continues with the
next check item.
For example, your Checklist might begin with some
PASS/FAIL items (\P items) for inputting information
about the condition of the pump under test, such as the
line cord and front panel keys and displays, and it might
end with AutoSequence 10 to test electrical safety of
the pump. In between, you want to conduct a test on
Channel A to check the flow at 100 ml/hr, and then
measure the occlusion pressure. To do the flow test,
your Checklist could include the following items:
\M RTA1
\M 180
\M STA
\M GSA1
This portion of a Checklist starts a Single Rate Test on
Channel A and runs the test for three minutes. The
flow test is stopped and medTester requests Infutest to
return the summary of the measurements made during
the test via the GSA1 command.
Now you want the medTester to tell Infutest to run the
Occlusion Pressure Test. To do that, the previous
check items would be succeeded by:
\M RTA4
\M 60
\M STA
\M GSA4
These check items start an Occlusion Pressure Test
www.datrend.com

Advertisement

loading

Summary of Contents for DATREND Systems medTester 5000B

  • Page 1 1. INTRODUCTION data from COM2, instead of from the medTester keyboard. In executing a Checklist with \M items, the The medTester 5000B is part of an equipment medTester will transmit whatever characters follow \M management data base and testing system available to the Infutest via COM2, and then will wait for data to from DNI Nevada Inc.(now Fluke Biomedical Corp.).
  • Page 2 5. medTester "BUGS" RECORD input box. This box allows you to enter new Check Items into Sentinel's Master List. For the DNI Nevada's medTester 5000B appears to have a command number of software bugs related to interface with externally connected devices. The type of bugs you...
  • Page 3 (e.g. cable, serial port setup, etc.). If the Infutest seems to respond to "run test" and "stop test" commands, but won't return test data to medTester, you are probably experiencing AN-17.2004: Page 3 of 4 1-800-667-6557 www.datrend.com © Datrend Systems Inc.
  • Page 4 For Run Sequence ("RS") commands, ensure that Infutest AutoSequence parameter "OUTPUT TO" is set to "Printer". Alternatively, AutoSequence AUTO-OUTPUT must be set to "Off". Do NOT use Infutest "DL" or "GR" commands with medTester. AN-17.2004: Page 4 of 4 1-800-667-6557 www.datrend.com © Datrend Systems Inc.