Do you have a question about the CN16 and is the answer not in the manual?
Questions and answers
Summary of Contents for EMC-PARTNER CN16
Page 1
User Manual CN16 PN 103538 CN16 is an accessory to TRA3000 C and EXT-TRA3000 C SHORT Title: Coupling Network CN16 Date: 30.04.1999 General Manager: M. Lutz Quality Manager: R. Henz Revised: 14. May 2014 Common Mode - Coupling Network CN16...
Common mode coupling network CN16 Contents: IEC 61000-4-16 INFORMATION Schematic circuit for the type tests Test levels T-Network Fehler! Textmarke nicht definiert. Mechanical dimensions Fehler! Textmarke nicht definiert. Accessories, dimensions 1.5.1 Included articles, dimensions 1.5.2 Standard accessories SAFETY Safety standards...
Page 3
Common Mode Coupling Network CN16 SERVICE INFORMATION GLOSSARY 3/28...
1 IEC 61000-4-16 Information 1.1 Schematic circuit for the type tests 1.2 Test levels E-CN16-E-Manual 5/28...
Page 6
106962 DN16-1P16 Mains decoupling transformer single phase 230V/16A 50/60Hz. For tests according to IEC 61000-4-16: DC, 50/60Hz and sinusoidal up to 150kHz . Housing 19" 4 UH For coupling select CN16, CN16T or CN16DC 106961 DN16-1P6 Mains decoupling transformer single phase 230V/6A 50/60Hz. For tests according to IEC 61000-4-16: DC, 50/60Hz and sinusoidal up to 150kHz.
EMC compatibility chapter. The CN16 should be operated in a dry, clean room. If for any reason water condenses in the CN16, then no CN16 operation should be started before the tester is dry.
Touch no connectors of connection cable when a EMC test is in operation. The high voltage of the CN16 network and the power on the EUT must turned off before a manipulation on the EUT is carried out.
Use the three cables (red, yellow and yellow/green) supplied to connect the CN16 to the TRA3000 C For common mode tests connect the CN16’s ‘AC-L’ and ‘AC-N’ to the TRA3000 C ‘EUT Power L and ‘EUT Power-N’ (as shown on the following picture for example).
Common mode coupling network CN16 4 AC Test-Voltage 4.1 AC test generator; coupling on one line Connect the cables and bridges as shown on the following picture: 12/28...
Common Mode Coupling Network CN16 4.2 AC test generator; coupling on two lines Maximum line to line voltage (EUT-AT) a.c. 230V and d.c.300V Connect the cables and bridges as shown on the following picture: 13/28...
Common mode coupling network CN16 4.3 AC test generator; coupling on four lines Maximum line to line voltage (EUT-AT) a.c. 230V and d.c.300V Connect the cables and bridges as shown on the following picture: 14/28...
5 DC-Voltage Test 5.1 DC test generator; coupling on one line Maximum line to line voltage (EUT-AT) a.c. ¨48V and d.c.68V Connect the cables and bridges as shown on the following picture: E-CN16-E-Manual 15/28...
Common mode coupling network CN16 5.2 DC test generator; coupling on two lines Maximum line to line voltage (EUT-AT) a.c. ¨48V and d.c.68V Connect the cables and bridges as shown on the following picture: 16/28...
Common Mode Coupling Network CN16 5.3 DC test generator; coupling on four lines Maximum line to line voltage (EUT-AT) a.c. ¨48V and d.c.68V Connect the cables and bridges as shown on the following picture: 17/28...
7 Recycling / Disposal 7.1 RoHS directive 2002/95/EG The CN16 complies with the directive 2002/95/EG (RoHS - Restriction of certain Hazardous Substances). From December 2005, all EMC Partner products either hand soldered or by machine are produced using lead-free solder.
Page 22
Common mode coupling network CN16 22/28...
Page 23
Common Mode Coupling Network CN16 8 Service Information EMC PARTNER AG Baselstrasse 160 CH - 4242 Laufen Switzerland ++41 61 775 20 50 ++41 61 775 20 59 service@emc-partner.ch www.emc-partner.com 23/28...
Page 24
Common mode coupling network CN16 24/28...
Page 25
Maximum continuous sinusoidal RMS voltage which may be applied. Rated DC voltage Maximum continuous DC voltage which may be applied. Insulation test The voltage waveform is relevant. Energy test The current waveform is relevant. Combination test The voltage and current waveform is relevant. E-CN16-E-Manual 25/28...
Page 26
Common mode coupling network CN16 Used symbols: Direct current Alternating current Three phase alternating current Earth (ground) terminal Protective conductor terminal IEC 417, No. 5019 Caution, risk of electric shock ISO 3864, No. B.3.6 Caution (refer to accompanying documents) ISO 3864, No. B.3.1...
Need help?
Do you have a question about the CN16 and is the answer not in the manual?
Questions and answers