Summary of Contents for USFilter DEPOLOX 3 PLUS

  • Page 1 DEPOLOX 3 PLUS ® RESIDUAL ANALYZER FOR MEASURING CHLORINE, CHLORINE DIOXIDE, OZONE, AND/OR pH OR FLUORIDE BOOK NO. IM 50.560CA UA ISSUE B...
  • Page 2 DEPOLOX 3 PLUS RESIDUAL ANALYZER ® EQUIPMENT SERIAL NO. _________________ DATE OF START-UP _____________________ START-UP BY ___________________________ Prompt service available from nationwide authorized service contractors. ORDERING INFORMATION In order for us to fill your order immediately and correctly, please order material by description and part number, as shown in this book.
  • Page 3 3 PLUS RESIDUAL ANALYZER ® INTRODUCTION The Depolox 3 plus Residual Analyzer is a low, cost economical analyzer for potable water that continuously measures either free or total chlorine, chlorine dioxide or ozone residuals using proven amperometric measure- ment technology. There is also an arrangement available for measuring pH or flouride.
  • Page 4 DEPOLOX 3 PLUS RESIDUAL ANALYZER ® Table A - Analyzer Type Selection Sheet Using Package Numbers f I ( f f u l a t c t i a l i i f e a l l f I ( IM 50.560CA UA (B/4-03) Introd.
  • Page 5 EN 50082-2 (immunity) • EN 61000 -4- 2, 3, 4, 5, 6, 8, 11 DEPOLOX 3 plus devices are designed for use in industry, as well as the domestic, business, and trade sectors. Declaration of Conformity DEPOLOX 3 plus devices comply with the requirements of the EU direc- tives 89/392/EC “Machinery Directive”, 73/23/EC “Low Voltage Direc-...
  • Page 6 DEPOLOX 3 PLUS RESIDUAL ANALYZER ® WARNING: TO AVOID INJURY TO PERSONS CAUSED BY ELECTRICITY ONLY AUTHORIZED AND QUALIFIED ELEC- TRICIANS MAY INSTALL THE DEVICE AND OPEN THE HOUS- ING. THE DEVICE MAY ONLY BE OPERATED WHEN THE HOUSING IS CLOSED AND MUST BE CONNECTED TO THE PROTECTIVE CONDUCTOR.
  • Page 7 DO NOT DISCARD THIS INSTRUCTION BOOK UPON COMPLETION OF INSTALLATION. INFORMATION PROVIDED IS ESSENTIAL TO PROPER AND SAFE OPERATION AND MAIN- TENANCE. ADDITIONAL OR REPLACEMENT COPIES OF THIS BOOK ARE AVAILABLE FROM: USFILTER’S WALLACE & TIERNAN PRODUCTS 1901 WEST GARDEN ROAD VINELAND, NEW JERSEY 08360 PHONE: (856) 507-9000 FAX: (856) 507-4125 IM 50.560CA UA (B/4-03)
  • Page 8 DEPOLOX 3 PLUS RESIDUAL ANALYZER ® VERY IMPORTANT SAFETY PRECAUTIONS (CONT'D) NOTE Minor part number changes may be incorporated into USF/W&T products from time to time that are not immediately reflected in the instruction book. If such a change has apparently been made in your equipment and does not appear to be reflected in your instruction book, contact your local USF/W&T sales representative for information.
  • Page 9 Direct any questions concerning this equipment that are not answered in the instruction book to the Reseller from whom the equipment was purchased. If the equipment was purchased directly from USFilter's Wallace & Tiernan Products (USF/W&T), contact the office indicated below. UNITED STATES...
  • Page 10 DEPOLOX 3 PLUS RESIDUAL ANALYZER ® IM 50.560CA UA (B/4-03)
  • Page 11 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER SECTION 1 - TECHNICAL DATA List of Contents PARA. NO DEPOLOX 3 plus ............. 1.1 Sensor Feature List ........... 1.2 pH Sensor Kit ............1.3 Fluoride Sensor Kit ........... 1.4 IM 50.560CA UA (B/4-03)
  • Page 12: Depolox 3 Plus

    ® DEPOLOX 3 PLUS RESIDUAL ANALYZER DEPOLOX 3 plus Electronics ± ± i s i l / g l / g i t c ) y l c i r l l a t a l i t c h t i...
  • Page 13: Sensor Feature List

    ® DEPOLOX 3 PLUS RESIDUAL ANALYZER Sensor Feature List i t n c i t r t c i t n c i t r t c i t n c i t r t c r t e , c i r t e , c i r t e...
  • Page 14 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER Sensor Feature List (Cont’d) i t n c i t r t c i t n c i t r t c c i r c i r c i l r e t h t i r e t h t i...
  • Page 15 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER pH Sensor Kit (AAB5386) ° ° r e t i v i > S µ NOTE: When used for pH compensated Free chlorine measurement, the pH range must be between 6 and 9. NOTE: Where the standard 1.5-meter sensor cable is not long enough (for special installations up to 50-meter cable length), an impedance transformer is used to convert the very high-resistance electrode sig- nal of the pH to a low-resistance signal.
  • Page 16 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER IM 50.560CA UA (B/4-03)
  • Page 17: Table Of Contents

    Unpacking .............. 2.1.1 Check Voltages ............2.1.2 Adjust Line Voltage/Change Fuse ......2.1.3 Select Chlorine Cell ..........2.1.4 Installing DEPOLOX 3 plus Module ....2.1.5 Electrical Connections ........... 2.1.6 Digital Input ............2.1.7 Start-Up ..............2.1.8 Free Chlorine (Bare Electrode) Sensor Kit ....2.2 Membrane Sensor Kit ..........
  • Page 18: Check Voltages

    2.1.2 Check Voltages The line voltage of the DEPOLOX 3 plus is set at the factory in accordance with the instructions given at the time of the order (230 V or 115 V).
  • Page 19 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER connection of cables. The device may be opened and adjusted by an electrician only. For connection to power with two hot wires (L1, L2), replace the metal pin on the F2 socket by a fuse corresponding to the voltage (see Figure 2.1).
  • Page 20 Select the sensor at the ten-position DIP switch S3 on the board (factory setting is for Bare Electrode). WARNING: WRONG SETTING CAN CAUSE DAMAGE TO THE SENSOR AND TO THE DEPOLOX 3 PLUS. Figure 2.2 - DIP Switch 3-Elect. “Depolox 3 plus”...
  • Page 21: Electrical Connections

    (only if sensor current >200µA to max. 1000 µA): The other positions remain as before. 2.1.5 Installing the DEPOLOX 3 plus Electronics Module The DEPOLOX 3 plus electronics module is built in a wall-mounted housing and should be mounted near the flow-through assembly. 2.1.6 Electrical Connections Connect the sensor cables, alarm signal lines and the power cable accord- ing to the wiring diagrams (see Dwgs.
  • Page 22: Start-Up

    DEPOLOX 3 PLUS RESIDUAL ANALYZER 2.1.8 Start-Up WARNING: THE DEPOLOX 3 PLUS MODULE HAS NO POWER SWITCH OF ITS OWN AND IS IN OPERATION AS SOON AS THERE IS A CONNECTION TO THE POWER SUPPLY. CON- NECTED DEVICES HAVE TO BE SWITCHED OFF DURING IN- PUT OF OPERATING DATA IN ORDER TO PREVENT UNCON- TROLLED START-UP OR MALFUNCTIONS.
  • Page 23 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER Bare Electrode (Free Chlorine) Sensor Kit The flow block assembly with the sensors should be installed in a frost- protected location and, where possible, in a heated room. The sample water should provide a minimum of approximately 2 psi (0.2 bar) water gauge pressure at the measuring cell inlet.
  • Page 24: Ph Sensor Kit

    Under these conditions the sensor should give proper readings for three to six months. For connection to the DEPOLOX 3 plus electronics, refer to Dwgs. 50.560.155.020 and 50.560.155.040. The sensor requires a one to two hour run in time before a first calibration can be performed.
  • Page 25: Flow Switch

    In general, the flow switch is only connected to the digital input of the DEPOLOX 3 plus. In case of other applications, ensure that the rating of this Reed contact (100 V AC and 0.5 A) is not exceeded even for a short time.
  • Page 26: Depolox 3 Plus

    ® DEPOLOX 3 PLUS RESIDUAL ANALYZER DEPOLOX 3 PLUS ELECTRONIC MODULE - DIMENSIONS 50.560.100.010 ISSUE 0 4-00 IM 50.560CA UA (B/4-03)
  • Page 27 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER BARE ELECTRODE MEASURING CELL - DIMENSIONS 50.560.100.020 ISSUE 1 1-03 IM 50.560CA UA (B/4-03)
  • Page 28 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER MEMBRANE MEASUREMENT PACKAGE - DIMENSIONS Sensors & Flow Block Assembly 50.560.100.030 ISSUE 1 1-03 IM 50.560CA UA (B/4-03)
  • Page 29 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER DEPOLOX 3 PLUS WITH BARE ELECTRODE SENSOR KIT - WIRING 50.560.155.010 ISSUE 1 1-03 IM 50.560CA UA (B/4-03)
  • Page 30 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER DEPOLOX 3 PLUS WITH MEMBRANE SENSOR KIT - WIRING 50.560.155.020 ISSUE 1 1-03 IM 50.560CA UA (B/4-03)
  • Page 31 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER DEPOLOX 3 PLUS WITH BARE ELECTRODE SENSOR KIT AND pH OR FLUORIDE SENSOR KIT - WIRING 50.560.155.030 ISSUE 1 1-03 IM 50.560CA UA (B/4-03)
  • Page 32 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER DEPOLOX 3 PLUS WITH MEMBRANE SENSOR KIT AND pH OR FLUORIDE SENSOR KIT - WIRING 50.560.155.040 ISSUE 1 1-03 IM 50.560CA UA (B/4-03)
  • Page 33 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER DEPOLOX 3 PLUS WITH BARE TWO ELECTRODE SENSOR KIT - WIRING 50.560.155.045 ISSUE 0 1-03 IM 50.560CA UA (B/4-03)
  • Page 34 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER ALARM-RELAYS DISINFECTANT pH/FLUOR ALARM 1 ALARM 2 ALARM 1 ALARM 2 DEPOLOX 3 PLUS - ALARM RELAYS - CONNECTIONS 50.560.155.050 ISSUE 1 1-03 IM 50.560CA UA (B/4-03)
  • Page 35: List Of Contents

    High - Low Alarms ..........3.1.5 Alarm Relays ............3.1.6 Dosing Contact for Chlorine ......... 3.1.7 mA Output ............. 3.1.8 DEPOLOX 3 plus - Calibration ....... 3.2 Free Chlorine (Bare Electrode) Calibration ..3.2.1 Membrane Calibration ........... 3.2.2 pH Calibration ............3.2.3 Fluoride Calibration ..........
  • Page 36: Depolox 3 Plus - Operation

    3 PLUS RESIDUAL ANALYZER DEPOLOX 3 plus - Operation 3.1.1 General Description The DEPOLOX 3 plus module is a microprocessor-controlled electronic amplifier for the measurement of Disinfectants, and/or for pH value or fluoride in water. The built-in RS485 interface can be used to transfer the measured values and operating modes to a PC or SCADA system.
  • Page 37 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER Display next menu (jump from menu title to menu title) ENTER - change into change mode (“>” is displayed before the value), - save new setting. NOTE: Check that any alterations have been ENTERED before ex- iting the menu.
  • Page 38: Menu Summaries

    ® DEPOLOX 3 PLUS RESIDUAL ANALYZER d. Store the changed setting by pressing the E key. Range DIS Figure 3.3 - Setting Changes As long as the new value has not been stored by pressing the E key, pressing the key will return to the former setting.
  • Page 39 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER IM 50.560CA UA (B/4-03)
  • Page 40 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER IM 50.560CA UA (B/4-03)
  • Page 41 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER 3.1.3.1 Root Menu l a t l i a : e l t i n l i a . r e c t i h t i o i t l l i r e t l / g s i h l l i...
  • Page 42 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER 3.1.3.2 Calibration Menu i l a r b i i t a n i t s i h . h t o l l i r u b i l o i t i l a r t c ) y l b i l...
  • Page 43 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER 3.1.3.3 Setup Menu i t t i r u l a r i t t . ) s l / g . l a l / g l / g l / g l / g l / g l / g .
  • Page 44 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER 3.1.3.4 High/Low Dinfection Alarm Menu s t l . . . l / g . . . l / g t n i . . . i t c r a l l e r l / g ) l / .
  • Page 45 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER 3.1.3.7 Alarm Relay 1, 2 (Always Assigned to Disinfectant) s t l n i f i t i o i t l e r i t a r a l l e r o l / l a r t l u r a l...
  • Page 46 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER 3.1.3.8 Alarm Relay 3, 4 (With pH Option Only) s t l n i f i t i o i t l e r i t a r a l l e r o l / l a r t l u r a l...
  • Page 47 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER 3.1.3.9 Alarm Relay 3, 4 (With Fluoride Option Only) s t l n i f i t i , r o o i t l e r i t a r a l l e r , r o o l / l a r...
  • Page 48 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER 3.1.3.10 Signal Diagnostics Menu µ µ °C l l e a t l i t p ) y l a t l i t p ) y l l u f i t p ) y l l u f i t p...
  • Page 49 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER 3.1.3.11 System Diagnostics Menu s t l o l l > " . " o l l w " " t r a t a l n i t , s t s " r a t "...
  • Page 50: Code Number

    SETUP menu and thus proved that you are an authorized operator. 3.1.5 High - Low Alarms The DEPOLOX 3 plus has independent alarms that can be allocated to the measured values for chlorine (and optionally for pH or fluoride). Menu Setting...
  • Page 51: Alarm Relays

    DEPOLOX 3 PLUS RESIDUAL ANALYZER Figure 3.4 - High / Low Alarms 3.1.6 Alarm Relays DEPOLOX 3 plus features the following alarm relays: • Two relays allocated to chlorine alarms. • Two relays allocated to pH or fluoride alarms (optional).
  • Page 52: Dosing Contact For Chlorine

    The signal that is output corresponds to the measured value. The con- nected impedance must not be higher than 1000 Ohm. Broken wires or open mA loops will be detected. DEPOLOX 3 plus - Calibration 3.2.1 Free Chlorine (Bare Electrode) Calibration...
  • Page 53 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER b. Close the shut-off valve at the flow block assembly, stopping all flow to unit. Wait until the displayed value ceases changing. c. Press the E key twice in order to set the display to “0.00”. d.
  • Page 54 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER • Span calibration a. After zero calibration wait at least two minutes. Then take a sample at the measuring cell. Determine the residual value of this sample. b. Press the key until the menu “DIS span” is reached. Press the E key to open the menu.
  • Page 55: Ph Calibration

    NOTE: The offset calibration is typically all that is recommended to calibrate pH when first starting the unit or replacing the sensor. The Depolox 3 plus zero and span values are factory set in order to prop- erly work with the sensor. Complete the zero and span calibrations, if there is a problem, only after completing the offset calibration.
  • Page 56 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER • pH Zero Calibration (Isothermal Intersection, pH7): a. Starting at the basic display press the F key until the “CALI- BRATION” menu is displayed. Press the key until the “pH zero” menu is found. b.
  • Page 57: Fluoride Calibration

    NOTE: The offset calibration is typically all that is recommended to calibrate fluoride when first starting the unit or replacing the sensor. The Depolox 3 plus zero and span values are factory set in order to properly work with the sensor. Complete the zero and span calibra- tions, if there is a problem, only after completing the offset calibra- tion.
  • Page 58 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER justed to “- 0.08 mg/l fluoride” because the value is to come down from 0.28 mg/l to 0.20 mg/l.) e. Return to the basic display by pressing the key twice. • Calibration at 0.20 mg/l: a.
  • Page 59: Free Chlorine (Bare Electrode) Sensor Kit

    ® DEPOLOX 3 PLUS RESIDUAL ANALYZER Free Chlorine (Bare Electrode) Sensor Kit 3.3.1 Start-Up of the Measuring Cell a. Unscrew nut (2) and remove cover (3) (see Dwg. 50.560.000.010 in Section 5). b. Hold electrode assembly together with electrolyte reservoir (4-11) at the cover (10) and pull it out of the cell block (13) without turning.
  • Page 60 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER c. Check flow and reassemble. d. Perform calibration. 3.3.3 Bare Electrode Theory of Operation The free chlorine that is measured in water for disinfection purposes comes in three forms. Free chlorine consists of elemental chlorine (Cl hypochlorous acid (HClO), and hypochlorite ions (CIO-).
  • Page 61 This current is then passed to the microprocessor-based electronic system DEPOLOX 3 plus for processing. A flow switch gives a signal to the DEPOLOX 3 plus if the flow is reduced (optional).
  • Page 62: Membrane Sensor Kit

    Membrane sensor, if necessary turn the cover of the flow block assembly. Do not tolerate any air bubbles at the membrane of the sensor. Start the DEPOLOX 3 plus module as described in paragraph 2.1.8, Start- 3.4.2 System Shut-Down •...
  • Page 63 Total Chlorine. An amplifier in the electrode shaft conditions this signal including temperature compensation and passes it to the micro- processor-based electronic system DEPOLOX 3 plus for processing. Air bubbles on the membrane may prevent chlorine from passing through the membrane, resulting in a false reading of the probe. A water jet, designed into the flow block assembly, is used to remove any potential air bubbles from the membrane.
  • Page 64: Ph Sensor Kit

    ® DEPOLOX 3 PLUS RESIDUAL ANALYZER • FC1, OZ7, CD7: The FC1 for Free Chlorine, the OZ7 for Ozone, and the CD7 for Chlorine Dioxide all work similarly, however, they utilize a silver/silver chloride reference electrode, a different elec- trolyte, and a different membrane to allow for the selectivity required for the different measurands.
  • Page 65: Fluoride Sensor Kit

    See instructions included with sensor. RS485 Interface 3.7.1 Printer Facility The DEPOLOX 3 plus is not able to send data directly to a printer. 3.7.2 Description of the RS485 Bus Interface The serial RS485 bus interface of the DEPOLOX 3 plus, built-in as a standard, is used for data transmission to a PC, to an external plant control or to the MF485 interface card (e.g., for connection to a printer).
  • Page 66 (large cross sections and short cables) to protection ground, to have optimum interference compatibility. 3.7.2.2 Interface Connection The bus cable for the communication with the DEPOLOX 3 plus should be connected to the following terminals: Bus line A: Terminal 28...
  • Page 67 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER Both ends of the bus cable must be connected to a moving load Rt (150 Ohm). Symmetry is assured at only one point of the bus. With a voltage supply of 5 V, the resistors Rd and Ru (390 Ohm each) connect to ground and 5 V.
  • Page 68: Specification Of Bus Interface

    ® DEPOLOX 3 PLUS RESIDUAL ANALYZER 3.7.3 Specification of Bus Interface 1 Synchronization mode: Asynchronous 2 Transmission rate: 19200 Baud 3 Data format (asynchronous): - Start bit: 1 Bit - Data bit: 8 Bit - Parity bit: even - Stop bit: 1 Bit 4 Signal polarity: Differential voltage interface...
  • Page 69 (slave address). The address can be a number be- tween 0 and 31 Dec. NOTE: The DEPOLOX 3 plus module uses up to two bus addresses because of the two-fold function of the module (Cl measurement, pH or fluoride measurement).
  • Page 70 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER 1101 ASCII 1110 Mixed data format 1111 Mixed data format Additional functions of the check byte (Bit 4 through 7): Bit (7654) Function 1000 Minimum value of the variable 0100 Maximum value of the variable 0010 Default value of the variable 0001...
  • Page 71 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER 3.7.4.2 Request - Frame This is necessary to read data or additional information from a slave. Format of the request frame: Byte: 0-2 —————————————————————— Name: SYN SB Byte: Name: occupied by: Synchronization bytes Start byte 10H Slave address destination address check byte...
  • Page 72 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER 3.7.4.3 Set - Frame The set frame is used for writing data to a slave. Format of the set frame: Byte: 9-X Y ————————————————————————— Name: SYN SB SA ZA KB AB FC DU DC EB Byte: Name: Function:...
  • Page 73 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER 3.7.4.4 Answer - Frame The answer frame is transmitted by the slave because of a request frame of the master. Format of the answer frame: Byte: 9-X Y ————————————————————————— Name: SYN SB SA ZA KB AB FC DU DC EB Byte: Name: Function:...
  • Page 74: Address Reference List

    SB SA ZA KB AB FC 3.7.5 Address Reference List NOTE: To each DEPOLOX 3 plus two addresses are allocated. Each of the two addresses lead to a corresponding address reference list (e.g., Adr00 for Cl measurement, Adr01 for pH or fluoride measurement).
  • Page 75 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER Table 3.4 - Address Reference List - Address 1 - Chlorine " " 5 " " r t n i 9 " " 4 " " S c i t l b i t n i t a l i g i l a t...
  • Page 76 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER Table 3.4 - Address Reference List - Address 1 - Chlorine (Cont’d) i f e t i n t l u i a f l / g . . . A " " 3 V "...
  • Page 77 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER Table 3.5 - Address Reference List - Address 2 - pH V " " 5 " " r t n i . . . 9 " " 4 " " H c i t l b i i g i l a t...
  • Page 78 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER Table 3.5 - Address Reference List - Address 2 - pH (Cont’d) i f e t i n t l u i a f A " " 3 V " " 1 IM 50.560CA UA (B/4-03)
  • Page 79 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER Table 3.6 - Address Reference List - Address 2 - Fluor " " 5 " " r t n i . . . 9 " " 4 " " l c i t l b i i g i l a t i t c...
  • Page 80 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER Table 3.6 - Address Reference List - Address 2 - Fluor (Cont’d) i f e t i n t l u i a f A " " 3 V " " 1 IM 50.560CA UA (B/4-03)
  • Page 81 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER BARE ELECTRODE SENSOR KIT - ASSEMBLY 50.560.160.010 ISSUE 1 1-03 IM 50.560CA UA (B/4-03)
  • Page 82: Membrane Sensor Kit

    ® DEPOLOX 3 PLUS RESIDUAL ANALYZER ELECTRODE SHAFT WITH AMPLIFIER COUNTER ELECTRODE REFERENCE ELECTRODE GOLD WORKING O-RING ELECTRODE MEMBRANE CAP ELASTOMER SEAL VENT HOLE MEMBRANE MEMBRANE SENSOR KIT - ASSEMBLY 50.560.160.020 ISSUE 1 1-03 IM 50.560CA UA (B/4-03)
  • Page 83 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER Reference electrode system Ag/AgCl Reference electrolyte (polymer gel) Ceramic diaphragm Ball membrane pH SENSOR KIT - ASSEMBLY 50.560.160.030 ISSUE 0 4-00 IM 50.560CA UA (B/4-03)
  • Page 84 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER IM 50.560CA UA (B/4-03)
  • Page 85 DEPOLOX 3 PLUS RESIDUAL ANALYZER SECTION 4 - SERVICE List of Contents PARA. NO DEPOLOX 3 plus ............. 4.1 Errors ..............4.1.1 Free Chlorine (Bare Electrode) Sensor Kit ....4.2 Maintenance ............4.2.1 Troubleshooting ............. 4.2.2 Membrane Sensor Kit ..........4.3 Maintenance ............
  • Page 86: Depolox 3 Plus

    ® DEPOLOX 3 PLUS RESIDUAL ANALYZER DEPOLOX 3 plus Table 4.1 - Errors r a l r e t o l f < r e t l l e i r i s e l r b i i t a...
  • Page 87 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER Table 4.1 - Errors (Cont’d) i g i l a t a t i g i r , l a , . g r e t o l f s n i s n i i t c i t t a t l...
  • Page 88 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER Table 4.2 - Relay / Display Operation During Alarm Conditions i l b i l b t l u t l u i l b t l u t l u i l b i g i l a t i g i...
  • Page 89: Free Chlorine (Bare Electrode) Sensor Kit

    ® DEPOLOX 3 PLUS RESIDUAL ANALYZER Free Chlorine (Bare Electrode) Sensor Kit 4.2.1 Maintenance • Daily checks Each day check the complete unit for leaks, including all pipe unions and glands, and rectify all leaks immediately. NOTE: All air leaks must receive immediate attention as they will have a detrimental effect on the operation of the cell unit.
  • Page 90 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER • Monthly checks Remove and clean, if fitted, the detachable element of the strainer in the sample water supply line. Check the calibration of the cell (refer to paragraph 3.2.1, Free Chlorine (Bare Electrode) Calibration) •...
  • Page 91 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER b. Drain the cell by unscrewing the drain plug. c. When the cell block is empty, install the drain plug. d. Unscrew the electrolyte reservoir clamping nut (2) and remove the cell cover (3). e.
  • Page 92 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER b. Unscrew the cover (10) and remove the push-fit connections from each electrode (and the temperature sensor if fitted). c. Unscrew the nut (9) which clamps the electrolyte reservoir to the electrode housing (8) and separate the two items. d.
  • Page 93: Troubleshooting

    ® DEPOLOX 3 PLUS RESIDUAL ANALYZER o l l & Figure 4.1 m. Reassemble the cover (10) and cable gland (11) to the electrode housing. n. Refit the electrode/electrolyte reservoir assembly and secure with the clamping nut (2). o. Reconnect the water inlet line, outlet line and the cell drain line. p.
  • Page 94 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER Table 4.3 - Troubleshooting - Free Chlorine (Bare Electrode) Sensor Kit c i f r e t l l a r t s t c i l l a o l f t r i t i n l l e r i a...
  • Page 95: Membrane Sensor Kit

    ® DEPOLOX 3 PLUS RESIDUAL ANALYZER Membrane Sensor Kit 4.3.1 Maintenance Daily: Check the entire system including all screw fittings for leakage and repair any leaks immediately. Using the cell outlet (see Dwg. 50.560.100.030 in Section 2), carry out a manual measurement every day. NOTE: The same test method should always be used in order to obtain consistent comparisons between results.
  • Page 96: Ph Sensor Kit

    ® DEPOLOX 3 PLUS RESIDUAL ANALYZER Table 4.4 - Troubleshooting - Membrane Sensor Kit i f f e i c o l f r o f n i a g n i r e t h g i t r i l l e g n i l l e...
  • Page 97: Fluoride Sensor Kit

    ® DEPOLOX 3 PLUS RESIDUAL ANALYZER to be determined empirically, but it is recommended to clean or calibrate it every four weeks. To prevent incorrect measurement, both the glass diaphragm and the ceramic membrane have to be cleaned. Clean the glass diaphragm mechanically with a moistened, clean and soft napkin or cot- ton swab.
  • Page 98 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER IM 50.560CA UA (B/4-03)
  • Page 99 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER SECTION 5 - ILLUSTRATIONS List Of Contents DWG. NO. Parts Bare Electrode Sensor Kit ......... 50.560.000.010A&B Bare Electrode Residual Analyzer..... 50.560.000.020A-E Membrane Flow Block Assembly ..... 50.560.000.030A&B IM 50.560CA UA (B/4-03)
  • Page 100 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER NOTE: FOR PARTS LIST, SEE DWG. 50.560.000.010B. BARE ELECTRODE SENSOR KIT - PARTS 50.560.000.010A ISSUE 1 1-03 IM 50.560CA UA (B/4-03)
  • Page 101 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER & WHEN ORDERING MATERIAL, ALWAYS SPECIFY MODEL AND SERIAL NUMBER OF APPARATUS. BARE ELECTRODE SENSOR KIT - PARTS LIST 50.560.000.010B ISSUE 1 1-03 IM 50.560CA UA (B/4-03)
  • Page 102 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER NOTE: FOR PARTS LIST, SEE DWG. 50.560.000.020E. BARE ELECTRODE RESIDUAL ANALYZER - PARTS 50.560.000.020A ISSUE 2 1-03 IM 50.560CA UA (B/4-03)
  • Page 103 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER NOTE: FOR PARTS LIST, SEE DWG. 50.560.000.020E. BARE ELECTRODE RESIDUAL ANALYZER - PARTS 50.560.000.020B ISSUE 1 1-03 IM 50.560CA UA (B/4-03)
  • Page 104 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER NOTE: FOR PARTS LIST, SEE DWG. 50.560.000.020E. BARE ELECTRODE RESIDUAL ANALYZER - PARTS 50.560.000.020C ISSUE 1 1-03 IM 50.560CA UA (B/4-03)
  • Page 105 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER NOTE: FOR PARTS LIST, SEE DWG. 50.560.000.020E. BARE ELECTRODE RESIDUAL ANALYZER - PARTS 50.560.000.020D ISSUE 1 1-03 IM 50.560CA UA (B/4-03)
  • Page 106 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER NOTE: INCLUDED IN ELECTRODE HOUSING. ONLY FOR TRANSPORT. WHEN ORDERING MATERIAL, ALWAYS SPECIFY MODEL AND SERIAL NUMBER OF APPARATUS. BARE ELECTRODE RESIDUAL ANALYZER - PARTS LIST 50.560.000.020E ISSUE 1 1-03 IM 50.560CA UA (B/4-03)
  • Page 107 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER MEMBRANE FLOW BLOCK ASSEMBLY - PARTS NOTE: FOR PARTS LIST, SEE DWG. 50.560.000.030B. 50.560.000.030A ISSUE 1 4-03 IM 50.560CA UA (B/4-03)
  • Page 108 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER WHEN ORDERING MATERIAL, ALWAYS SPECIFY MODEL AND SERIAL NUMBER OF APPARATUS. MEMBRANE FLOW BLOCK ASSEMBLY - PARTS LIST 50.560.000.030B ISSUE 1 4-03 IM 50.560CA UA (B/4-03)
  • Page 109 ® DEPOLOX 3 PLUS RESIDUAL ANALYZER SECTION 6 - SPARE PARTS LIST List Of Contents PARA. NO. Free Chlorine (Bare Electrode) Sensor Kit ....6.1 Membrane Sensor Kit ..........6.2 pH Sensor Kit ............6.3 Fluoride Sensor Kit ........... 6.4 Electronics ..............
  • Page 110: Free Chlorine (Bare Electrode) Sensor Kit

    ® DEPOLOX 3 PLUS RESIDUAL ANALYZER Free Chlorine (Bare Electrode) Sensor Kit DESCRIPTION PART NO. Spare part set for two years operation UXA96819 Spare part set for five years operation UXB96819 Individual spare parts: Electrode housing (with working and counter electrode) U95827 Membrane (two required) U95641...
  • Page 111: Ph Sensor Kit

    ® DEPOLOX 3 PLUS RESIDUAL ANALYZER 15m Cable AAC4693 25m Cable AAC5812 50m Cable AAC5815 pH Sensor Kit DESCRIPTION PART NO. Complete pH Sensor Kit AAB5386 pH sensor U95690 cable for the pH sensor U93838 Impedance transformer (optional) U95607 Fluoride Sensor Kit DESCRIPTION PART NO.
  • Page 112: Electronics

    ® DEPOLOX 3 PLUS RESIDUAL ANALYZER Electronics DESCRIPTION PART NO. PCB Assembled - 115V Cl AAB1501 PCB Assembled - 230V Cl AAB1513 PCB Assembled - 115V Cl + pH/Fluor AAB1504 PCB Assembled - 230V Cl + pH/Fluor AAB1516 Fuse cover P96994 160 mA fuse UXF92568...