In order to use this device safely and correctly, please read this manual carefully before
operating it.
Please register promptly as 10.4 in user's manual after installation.
It is V1.2 for the version number of this user's manual.
Note:
1) Please check the power supply of adapter before charging the battery. Stop using once
the voltage of the power exceeds the scope printed on the adapter.
2) Please check the power cable of charger before charging the battery. Broken or cracked
power cable may cause a fire or electric shock which may lead to casualties.
3) Please unplug the power cable and take out the battery after charging completed.
Incomplete blackout may result in the occurrence of unusual or unsafe (Such as:
thunder strike and static electric fire, etc.)
4) Attentions before turning on the device:
a. Make sure the probe connect tightly with mainframe;
b.
Make sure you have good battery, put it into battery case and lock the cover.
5) Turn off the power after finishing using.
6) Do not be away from the device while it is being used.
7) Please calibrate the date and time before first time using (or not using this device for a
long time).
8) The system will enter the energy-saving mode without any operation on the device for
about 5 minutes if not in the pre-scan mode or touch screen calibration interface. The
system will automatically shut down without any operation on the device for 10
minutes.
9) The battery should be taken out from the device if not using for a long time.
10) Stop using the device and turn it off in stormy weather; do not use the charger and cut
off the power, take out the battery in such weather.
11) Please stop charging battery and cut off power, when charging battery or using device,
if any irregularities occur as follows:
a. Abnormal noise;
b. Sudden fume or burnt odor
c. Brightness of the display screen is too high.
PBS V5.1 User's Manual
1

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the PBSV5.1 and is the answer not in the manual?

Questions and answers

Deb Heiser
April 21, 2025

I need the 4 number code

Summary of Contents for Meike PBSV5.1

  • Page 1 PBS V5.1 User’s Manual In order to use this device safely and correctly, please read this manual carefully before operating it. Please register promptly as 10.4 in user’s manual after installation. It is V1.2 for the version number of this user's manual. Note: 1) Please check the power supply of adapter before charging the battery.
  • Page 2 14) The user needs to install and use the device according to electromagnetic compatibility information provided by the accompanying documents. 15) Portable and mobile RF communication devices may affect the performance of PBSV5.1 Palm bladder scanner. Please avoid these strong electromagnetic interference such as mobile phones and microwave ovens, etc.
  • Page 3 7) The device is not allowed to use together with high-frequency surgical equipment. 8) Don’t pull out the probe while using the device. 9) PBSV5.1 Palm bladder scanner should not be close to or stacked with other equipment. If adjacent or stacked use is necessary, it should be observed to verify normal operation in the configuration.
  • Page 4 This manual includes user’s manual and technical instruction Thanks for your attention Mianyang Meike Electronic Equipment Co., Ltd. Registered Add.: A214#, No. 133, Mianxing East Road, High-tech District, Mianyang Manufacturing Add.: No.63, Yinping Road, Longmen Town, Fucheng District Mianyang Sichuan, CHINA Contact Add.: No.63, Yinping Road, Longmen Town, Fucheng District, Mianyang Sichuan, CHINA...
  • Page 5 PBS V5.1 User’s Manual Revision History Version Date Description V1.0 2017.08 First Edition Add the illustration for not configuring wireless V1.1 2018.10 related hardware To change the background icon into "ready" icon of V1.2 2019.03 the schematic diagram for the main interface on Fig 7-3.
  • Page 6: Table Of Contents

    PBS V5.1 User’s Manual Contents CHAPTER 1 OVERVIEW ............................1 ................................1 EATURE ............................... 1 TRUCTURE ......................错误!未定义书签。 COPE OF APPLICATION ............................1 ONTRAINDICATION ................... 2 LLUSTRATION ABOUT SIGN OF INTERFACE ON THE DEVICE : ........................ 2 ESCRIPTION OF THE ACOUSTIC OUTPUT CHAPTER 2 TECHNICAL SPECIFICATION ......................
  • Page 7 PBS V5.1 User’s Manual ..........................17 WITCH ON OFF THE DEVICE ............................. 17 OGIN NTERFACE ..........................18 YSTEM NTERFACE ......................21 SCAN INTERFACE AND CAN INTERFACE ............................ 22 CAN RESULT INTERFACE ..............24 CANNING RESULTS TWO COLOR INTERFACE AND GRAYSCALE INTERFACE .........................
  • Page 8 PBS V5.1 User’s Manual 10.1 ..........................57 AINTENANCE FOR PROBE 10.2 ..........................57 AINTENANCE FOR DEVICE 10.3 ..............................57 ERVICE GUIDE 10.4 ............................58 EGISTER METHOD 10.5 ............................58 RODUCTION DATE CHAPTER 11 SIMPLE PROBLEM ELIMINATION ....................59 APPENDIX A PROBE CALIBRATION ........................60 APPENDIX B VOLUME CALIBRATION........................
  • Page 9: Chapter 1 Overview

    Overview 1.1 Feature PBSV5.1 Palm bladder scanner is used for measuring bladder volume with its ultrasonic echo. It consists of data processing of main frame and ultrasonic probe. Its display is color TFT-LCD with Pixel 800×600, and its measurement results can be printed out with thermal printer and also can be stored in the built-in flash-disk.
  • Page 10: Illustration About Sign Of Interface On The Device

    PBS V5.1 User’s Manual The device is not suitable for those who are allergic to ultrasound gel and the people with abdomen skin wound and skin disease. 1.6 Side effect and Complication 1.7 Illustration about sign of interface on the device Sign Meaning Sign...
  • Page 11 PBS V5.1 User’s Manual Fig 1-1 Technical Data Sheet for level of acoustic output 2)Device accords with requirements of IEC 60601-2-37:2007+A1:2015 “Part 2-37 : Particular requirements for the safety of ultrasonic medical diagnostic and monitoring equipment” 201.12.4.2 a) ,b) . Exempted from showing MI, TI. MI <1 TI<1...
  • Page 12: Chapter 2 Technical Specification

    Transducer: 2.6 MHz; 2.1.4 Detect depth: ≥140 mm; 2.1.5 Scope of Measurement volume: 0 ml~ 999 ml; 2.1.6 Measurement accuracy:±15%,±15 ml(on a Meike tissue-equivalent phantom) 2.1.7 The display unit of volume measurement results: ml 2.1.8 Structure form: clamshell 2.1.9 Power supply voltage adaption range.
  • Page 13 PBS V5.1 User’s Manual 6) Input interface: one probe interface; one bar code scanning interface; one charging interface. 7) Output interface: one USB interface. 2.2.2 Equipment software function Software function item Software normal switching machine. Login In Display mode (expert/easy) Printer status Probe connection status.
  • Page 14 PBS V5.1 User’s Manual Display mode Settings Scanning bit Settings. Network Settings Key tone switch Settings. System Delete all records Settings Scan mode Settings The probe calibration Volume calibration Time, date setting. Server configuration Touch screen calibration Equipment information display Equipment calibration function Temperature of overheating prompt.
  • Page 15: Chapter 3 System Principle

    PBS V5.1 User’s Manual Chapter 3 System Principle 3.1 Principle: Palm bladder scanner is a medical device with high performance combined with modern ultrasound technology and computer technology. The device consists of host and probe. It can speedily complete the detection of bladder area through scan of probe connected with the device, and transmit ultrasound echo signal detected to embedded computer system.
  • Page 16 PBS V5.1 User’s Manual Application interaction Signal Processing Motor Driver LVDS Driver QR/Bar Code Micro Printer 8 Chanel QSPI+CMDI LVDS Hi Pulse DHPF FPGA Driver Memory +5.0V/2A +3.3V/1A +AV5.0V/200mA Li-on Battery Power Switch -AV5.0V/200mA (Include System Protection) +11V/400mA +HV/20mA DC Input Li-on -HV/20mA 9V/2A...
  • Page 17: Pictures Of The Device

    PBS V5.1 User’s Manual 3.3 Pictures of the device 3.3.1 The picture of the device, see Figure 3-3 ① ② ④ ③ ⑤ ⑥ ⑦ ⑧ ⑨ Figure 3-3 Device picture ① Touch screen ② Control panel ③ Power indicator ④...
  • Page 18: Circuit Diagram And Components List Description

    PBS V5.1 User’s Manual 3.3.2 Host interface schematic diagram (See Figure 3-4 ) ② ① ④ ③ Figure 3-4 Host interface schematic diagram ① USB interface ② Probe interface ③ Recharging interface ④ The bar code scanning port 3.3.3 3D probe schematic diagram, see Figure 3-5 ②...
  • Page 19 PBS V5.1 User’s Manual In case of eligible technicians of user’s part have any need to repair device parts permitted and indicated by our company, may apply to our company or any offices for circuit diagram, parts list and other materials connected with the parts...
  • Page 20: Chapter 4 Control Panel

    PBS V5.1 User’s Manual Chapter 4 Control Panel 4.1 Control Panel ⑪ ⑩ ① ② ③ ④ ⑤ ⑥ ⑦ ⑧ ⑨ Figure 4-1 Control Panel ① Scan Button ② Print Button ③ Up Button ④ Left Button ⑤ Down Button ⑥...
  • Page 21: Chapter 5 Environmental Requirement

    PBS V5.1 User’s Manual Chapter 5 Environmental Requirement 5.1 Environmental requirement for device use 5.1.1 Environment temperature range: +10℃~+40℃; 5.1.2 Relative humidity range: 30%~75%; 5.1.3 Atmospheric pressure range: 700 hPa~1060 hPa; 5.1.4 Power supply of charger: AC 100V~240V 50Hz/60Hz; 5.1.5 Away from strong electrical field, strong magnetic field and high-voltage equipment.
  • Page 22: Chapter 6 Device Installation

    PBS V5.1 User’s Manual Chapter 6 Device Installation 6.1 Check after opening the carton 6.1.1 Please check the device according to “Packing list”, and confirm if every part have been accorded with the packing list, including in the user’s manual, maintenance document and accessories.
  • Page 23 PBS V5.1 User’s Manual Figure 6-2 Back view of the device with battery 6.2.2 Taking the battery out a. Open the battery cover according to the battery loading process. b. Take out the battery in turn. c. Install the battery cover according to the way after the battery loading process Warning: The battery must be installed in strict accordance with the electrode polarity within the battery tank.
  • Page 24 PBS V5.1 User’s Manual 6.2.3 Probe Installation Indication Indication Indication point 2 point 3 point 4 Indication point 1 Figure 6-3 Probe socket Figure 6-4 Probe plug a. Find the indication point 1 and 2 on the probe socket indication point 3 and 4 as shown in Figure 6-3 &...
  • Page 25: Chapter 7 Operation Manual

    PBS V5.1 User’s Manual Chapter 7 Operation Manual 7.1 Preparation before switching on the device 7.1.1 Check if the battery is properly installed. 7.1.2 Checking if the host and probe are connected properly. 7.1.3 Check if the device is fixed on the trolley steadily according to “Mobile Workstation Installation Instructions”...
  • Page 26: Main System Interface

    PBS V5.1 User’s Manual Figure 7-2 Login Interface 7.3.2 Touch the digital icons on the display screen to input the password, and will automatically enter into the system if the password are correct. Note: 1) The password consists of four digits; 2) Please contact the dealer or manufacturer for the password before the first use or after losing the password;...
  • Page 27 PBS V5.1 User’s Manual Table 7-1 Illustration for status bar Icon name Icon Description Operator ID The operation ID can be input in reset interface. Expert Operator Type Easy Normal Connection Printer status Abnormal connection or shortage of printing paper Normal Connection of probe Probe status Abnormal Connection of probe...
  • Page 28 PBS V5.1 User’s Manual After entering main system operation interface, both touch function keys and panel buttons can be used to operate. Figure 7-4 Touch function keys Figure 7-5 Panel Buttons a. Press button or touch icon to start barcode scanning; scan the patients’ examination application bar code to get the patient ID.
  • Page 29: Pre-Scan Interface And Scan Interface

    PBS V5.1 User’s Manual Note: Each status icon in the status bar changes according to the system settings or system detection. The function of the touch function icon will vary with the interface. If the device is in the trial period, the main interface will display "Please calibrate again before: ****-**-**"...
  • Page 30: Scan Result Interface

    PBS V5.1 User’s Manual 7.5.2 Pre-scan status of expert mode will show the real-time display of ultrasonic scanning images; In simple mode, only the identified liquid dark area is shown, see figure 7-7 yellow area. 7.5.3 Expert mode will display the bladder contour in real time, and the position state of the bladder will be indicated by the contour color.
  • Page 31 PBS V5.1 User’s Manual 7.6.1 The interface effect of the scan results is shown in figure 7-9, indicating the current measurement result value and position prompt. Figure 7-9 The interface effect of the scan results a. Press right button or touch , back to the “main interface”.
  • Page 32: Scanning Results Two-Color Interface And Grayscale Interface

    PBS V5.1 User’s Manual the grayscale image are overlapped. If the deviation is too large, please do manual correction or re-scanning. 7.7 Scanning results two-color interface and grayscale interface. 7.7.1 The interface effect of the scan result of two-color interface is as shown in figure 7-10.
  • Page 33: Result Correction Interface

    PBS V5.1 User’s Manual b. Press or touch , Enter the “save edit interface”. c. Press or touch , switch between grayscale and 2- color graph. d. Press or touch , switch the current display images as the last four images of the scan result.
  • Page 34: Save & Edit Interface

    PBS V5.1 User’s Manual b. Touch to select the image which needs to re-locate bladder edge, and re-identify bladder edge at the touched position as the center of bladder dark area. The new envelope of bladder edge can be displayed. c.
  • Page 35: Browse Historical Record Interface

    PBS V5.1 User’s Manual a. Press button or scan the touch icon to scan the barcode of the patient examination application form to get the patients’ ID. b. Input the patients’ name from the soft keyboard in the touch screen. c.
  • Page 36 It is possible that one of the patient ID and name recorded in the test record is null based on the input condition while saving. PBSV5.1 with English version is not configured for WLAN. Press PRINT button or upload icon has no effect for the device.
  • Page 37: Check Record Detail Interface

    PBS V5.1 User’s Manual 7.11 Check Record Detail Interface 7.11.1 Figure 7-15 is the measurement record detail interface. Figure 7-15 Measurement Record Detail Interface 7.11.2 Measurement record detail are including in patient ID, name, measurement result, measurement date, the grayscale of the scan. The current case can be printed or deleted. 7.11.3 Instruction a.
  • Page 38: System Reset Interface

    PBS V5.1 User’s Manual g. Press button or touch to delete this record. Please touch the delete icon for 2 times to delete the measurement record. 7.12 System reset Interface 7.12.1 Figure 7-16 is System reset Interface. Figure 7-16 System reset Interface 7.12.2 Operator: Set the operator ID and display 3 characters only.
  • Page 39 PBS V5.1 User’s Manual ress & button or touch & to move and to select the option of “the operator”. b. Press button or touch to delete the input operator information. c. Press button or touch to start barcode scanning, scan operator code to get the operator ID.
  • Page 40 PBS V5.1 User’s Manual c. Press button or touch to save the settings and return back to the main system interface. Or save all settings after other settings are set. 7.12.4 Scan planes: Set scan planes quantity. Operation Function Keys as following Figure 7-19 Operation Function Keys of Scan Planes Setting Instruction a.
  • Page 41 PBS V5.1 User’s Manual Figure 7-20 Operation Function Keys of WLAN Setting Instruction a. Press & button or touch to move to select the option of WLAN. b. Press button or touch to set on and off of the WLAN. c.
  • Page 42 Or save all settings after other settings are set. Note: PBSV5.1 with English version is not configured for WLAN setting.WLAN setting has no effect for the device. 7.12.6 Sound: set the key sound...
  • Page 43 PBS V5.1 User’s Manual a. Press & button or touch & to move to select the option of sound. b. Press button or touch to set the key sound on/off. c. Press button or touch to save the settings and return back to the main system interface.
  • Page 44 PBS V5.1 User’s Manual Figure 7-24 Memory Format the sub-menu. c. Press button or touch to select memory format. The formatting progress and status will be displayed in the position of formatting Settings value, and return to the formatting menu after completing. d.
  • Page 45 PBS V5.1 User’s Manual Figure 7-25 Operation Function Keys of Mode Reset Instruction a. Press & button or touch & to move to select the “mode”. b. Press button or touch to switch the scan mode human/phantom. c. Press button or touch to save the settings and return back to the main system interface.
  • Page 46 PBS V5.1 User’s Manual Fig 7-26 Probe Calibration Reset Operation keys Instruction a. Press button & button or touch & to move and select “Probe Calibration” . b. Press button or touch to set the value adding 1. c. Press button or touch to set the value minus 1.
  • Page 47 PBS V5.1 User’s Manual Figure 7-27 Volume Calibration Operation Keys Instruction a. Press button & button or touch & to move to select “ Volume Calibration” b. Press button or touch to set the value adding 1. c. Press button or touch to set the value minus 1.
  • Page 48 PBS V5.1 User’s Manual Figure 7-28 Time Setting Function Keys Instructions: a. press button and or touch , move selection items and select the "Time setting" option. b. press button or touch , enter the time setting sub-menu, the interface is shown in Figure 7-29: Figure 7-29 Time Setting Sub-menu c.
  • Page 49 PBS V5.1 User’s Manual d. Press button or touch , adjust the underlined character value; e. Press button or touch , finish time setting, return to time setting menu. f. press scan button or touch , save the setting and return to the main menu.
  • Page 50 Or wait for other settings and then press this to save all settings. Note: PBSV5.1 with English version is not configured for setting of server configuration. This setting of server configuration has no effect for the device. 7.12.13 Touch screen calibration: Calibrate the touch screen.
  • Page 51 PBS V5.1 User’s Manual Figure 7-32 Touch Screen Calibration Operation Function Keys Instructions: a. Press button or touch , move selection items and select the "Touch screen calibration" option. b. Press button or touch , enter the touch screen calibration interface as shown in Figure 7-33.
  • Page 52 PBS V5.1 User’s Manual c. Press scan button or touch , save the setting and return to the main menu. Or wait for other settings and then press this to save all settings. 7.12.14 About: Display device information, including manufacturer information and software version information.
  • Page 53: Device Verification Interface

    PBS V5.1 User’s Manual Figure 7-35 Device Information Interface i. Press scan button or touch , exit device information interface. ,save the setting and return to the main c. Press scan button or touch menu. Or wait for other settings and then press this to save all settings. 7.13 Device verification interface When the trial time expires, the device verification interface will automatically pop up after the device is powered on.
  • Page 54: Location Prompt

    PBS V5.1 User’s Manual 7.13.1 Send the serial number of the device to the distributor to obtain the check number corresponding to the serial number. 7.13.2 Enter the activation code by touching the character keys. 7.13.3 Press button or touch , verify the product.
  • Page 55: Test Record Printing

    PBS V5.1 User’s Manual Head Head Right Right Left Left Foot Foot Figure 7-41 Figure 7-42 Move the probe towards patient’s feet. Move the probe towards patient’s head Head Head Right Right Left Left Foot Foot Figure 7-43 Figure 7-44 Move the probe to right Move the probe to left 7.15 Test record printing...
  • Page 56: Thermal Printer

    PBS V5.1 User’s Manual 7.16 Thermal printer 7.16.1 The device has a built-in thermal printer. 7.16.2 Operation for changing the thermal printer paper ① Figure 7-46 Figure 7-47 Figure 7-48 Figure 7-49 a. As shown in Figure 7-46, open the cover ①. b.
  • Page 57: Energy Saving Mode

    Note: When in USB mode, do not copy unrelated files to the bladder scanner device on the PC; PBSV5.1 with English version is not configured for WIFI wireless network export. WIFI wireless network export has no effect for the device.
  • Page 58: Over Temperature Tips

    PBS V5.1 User’s Manual 7.18.1 In order to prolong the working time of the device. In the state of being turned on, the device will automatically turn off if there is no operation within 5 minutes and enter the power-saving mode. In power-saving mode, the power indicator is still green. 7.18.2 After the device enters the energy-saving mode, clicking anywhere on the touch screen or pressing any key (except the power switch key) can wake up the device and enter the normal working state.
  • Page 59 Warning: 1) Meike will supply the charger for every device. If the user purchase the out sourcing charger, it must meet the requirements of IEC/EN 60950-1, and Meike will be not responsible for any safety accident.
  • Page 60 PBS V5.1 User’s Manual 3) The power adapter provided with this device meet the requirements of the IEC 60601-1:2005+A1:2012 standard. 4) In order to ensure the safety of the operator and the patient, the device cannot be started for detection during the charging.
  • Page 61: Chapter 8 Clinical Application

    PBS V5.1 User’s Manual Chapter 8 Clinical Application 8.1 Operation Steps: 8.1.1 Patient should be in supine posture when it is scanned with the device. 8.1.2 The coupling agent will be evenly put on the acoustic window of the probe. 8.1.3 Put the probe placed in the human abdomen at about 2 cm above the pubic bone and the direction of probe must be slightly sloped towards to the head of patient, aiming at the...
  • Page 62 PBS V5.1 User’s Manual Figure 8-1. Too large red edge line Figure 8-2 Too large red edge line Figure 8-3 Too large red edge line Figure 8-4 Too small red line Figure 8-5 Too small red line Figure 8-6 red line is not in dark fluid area Figure 8-7 Non-fluid area Figure 8-8 Non-fluid area Figure 8-9...
  • Page 63 PBS V5.1 User’s Manual Note: If Bladder profile envelope is too big, then its volume will be too big If Bladder profile envelope is too small, then its volume will be small If there is no Bladder profile envelope, then the result will be small If Bladder profile envelope is not in dark fluid area, its scanned result will be ineffective.
  • Page 64: Chapter 9 Transportation And Storage

    PBS V5.1 User’s Manual Chapter 9 Transportation and Storage 9.1 Transportation The marks on the device packaging case conform to the requirement of ISO 780:2015 “Packaging-Distribution packaging-Graphical symbols for handling and storage of packages”. With shockproof facility inside of case, be fit for air, railway, highway and steamboat transportation.
  • Page 65: Chapter 10 Maintenance And Service

    PBS V5.1 User’s Manual Chapter 10 Maintenance and Service 10.1 Maintenance for probe 10.1.1 As the probe is valuable and damageable part, any collision or falling is forbidden. Especially give attention to the acoustic window of probe, must guard against any damage, even any minim scar also will reduce the effect and its life.
  • Page 66: Register Method

    PBS V5.1 User’s Manual 10.3.1 Once the device has problem, firstly check and make sure all of their connection are OK, and enough electricity of battery. If the device still doesn’t work after the above work, please contact your local distributor or manufacturer. 10.3.2 the following necessary information must be needed when contacting: a.Model of the device and its configuration;...
  • Page 67: Chapter 11 Simple Problem Elimination

    PBS V5.1 User’s Manual Chapter 11 Simple Problem Elimination To guarantee the normal use of this device, the solutions for trouble shooting are as below (For reference only): 1) Check that the battery is properly installed and that the battery has a certain charge 2) Check the connection between probe and device and lock up.
  • Page 68: Appendix A Probe Calibration

    PBS V5.1 User’s Manual Appendix A Probe Calibration A. 1 Fix the probe on a holder and make sure that the probe is placed vertically on the center of acoustic window of phantom (standard phantom 160 ml ) See Figure A-1. Figure A-1 the place of probe A.
  • Page 69 PBS V5.1 User’s Manual Figure A-5 ∠a When 0°< ∠a < 15°, 0 < value of scanning adjust < 10 Formula: When 15°< ∠a < 30°, 10 < value of scanning adjust< 20; When 30°< ∠a < 45°, 20 < value of scanning adjust < 30; When 45°<...
  • Page 70 PBS V5.1 User’s Manual Formula: When 0°< ∠a < 15°, 0 < value of scanning adjust < 10 When 15°< ∠a < 30°, 10 < value of scanning adjust < 20; When 30°< ∠a < 45°, 20 < value of scanning adjust < 30; When 45°<...
  • Page 71: Appendix B Volume Calibration

    PBS V5.1 User’s Manual Appendix B Volume Calibration B. 1 Fix the probe on a holder and make sure that the probe is placed vertically on the center of acoustic window of phantom ( standard phantom 160 ml ) See Figure B-1 Figure B-1 The place of probe B.2 Please press “OK”...
  • Page 72 PBS V5.1 User’s Manual Figure B-3 Pre-scan image B.4 Find the biggest round dark area(Target) by moving probe, adjust the distance between probe and phantom, let the round dark area (Target) be in the middle of fan-image in pre-scan. The red line shown in Figure B-3 is central line, the object measuring must be across this center line, the edge of round dark area is green.
  • Page 73 PBS V5.1 User’s Manual 160 (phantom volume ) – mean value = calibrating value B.8 Select Calibration in the system reset interface for the volume calibration. B.9 Phantom Parameter: Phantom Parameter: Sound speed of background material: 1540 m/s±10 m/s,23℃±3℃ Attenuation of background material:0.7 dB/(cm•MHz)±0.05 dB/(cm•MHz),23℃±3℃ Protective plate:≤0.4 mm he internal target parameter of the Phantom Density: 1021±1 kg/m...
  • Page 74: Appendix C Precautions For 18650 Battery Use

    PBS V5.1 User’s Manual Appendix C Precautions for 18650 Battery use C.1 Prohibition for 18650 Battery C.1.1 Prohibit to collide the battery with sharp parts. C.1.2 Keep a clean environment, avoid sharp components exist around battery. C.1.3 Do not perforate the battery with nail or other edge tools. C.1.4 Do not transport or keep together with metallic things, such as necklace and barrette.
  • Page 75 PBS V5.1 User’s Manual 4) When using external charger to charge 18650 battery, the voltage of single battery must be less than 4.2 V during charging. Warning: Power supply must be cut off once if the battery is found out of the way during recharging, for instance, expand, smoke, break, then before touch or closing to battery in 4 hour later and this battery is forbidden to be used again.
  • Page 76 PBS V5.1 User’s Manual 8)Prohibit use or place of the battery in high temp. ( e.g: strong sunshine, fire or place with a higher than 60℃ inside of car) environment. C.5.2 you must wash with running water and see the doctor in case the electrolyte leaks out and touch your skin, eyes or other part of body, C.5.3 Do not use any damaged battery ( such as broken battery, breakage of shell, the smell of electrolyte, leakiness, etc.)
  • Page 77: Appendix D Guidance And Manufacturer's Declaration

    Guidance and manufacturer’s declaration – electromagnetic emissions The PBSV5.1 Palm bladder scanner is intended for use in the electromagnetic environment specified below. The customer or the user of the SECP-II should assure that it is used in such an environment.
  • Page 78 Guidance and manufacturer’s declaration – electromagnetic immunity The PBSV5.1 Palm bladder scanner is intended for use in the electromagnetic environment specified below. The customer or the user of the PBSV5.1 Palm bladder scanner should assure that it is used in such an environment. IEC 60601 test...
  • Page 79 RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the PBSV5.1 Palm bladder scanner is used exceeds the applicable RF compliance level above, the PBSV5.1 Palm bladder scanner should be observed to verify normal operation.
  • Page 80 PBS V5.1 User’s Manual For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.