Advertisement

DIGITAL THERMOMETER
INSTRUCTION MANUAL

1. GENERAL DESCRIPTION

This 3 1/2 digital thermometer with highly stable performance and high accuracy is driven by 9V
battery. Any K-type thermocouple can be used as the temperature sensor. It adopts new style holster
with preventing dropping design, can be comfortably handled.
This instruction manual covers information on operation and safety. For better use, please read the
relevant information carefully and observe all the notes strictly. Please read this operation manual
carefully before your operation!

2. GENERAL SPECIFICATIONS

2.1 Display: 3 1/2 digit large LCD display, Max. display 1999.
2.2 Polarity display: automatically shows polarity symbol "-".
2.3 Over range display: "1" or "-1"
2.4 Low battery indication: LCD displays "
2.5 Sampling rate: approx. 2.5 times/sec.
2.6 Storage environment: relative humidity 0%-80% at -10~60℃ or 10~140℉.
2.7 Working temperature: 0~50℃ or 32~122℉
2.8 Max input voltage: DC 60V or AC 24V.
2.9 Battery: 9V battery 006P or IEC6F22 or NEDA1604.
2.10 Battery life: 200 hours for alkaline battery, and 100 hours for carbon battery.
2.11 Dimension: 130mm (L) × 95mm (W) × 28mm (H).
2.12 Weight: approx.240g (including battery).
2.13 Accessories: manual, gift box, TP01 thermocouple, 9V battery.

3. TECHNICAL DATA

3.1 Error coefficient: 0.1 × accuracy/℃ when the temperature is lower than 18℃ or higher than 28℃.
3.2 Accuracy: ± (a%× reading + digit).
3.3 Accuracy adjustment environment: 23℃± 5℃.
3.4 Measurement range: (-50~1300)℃ (-50~1999)℉
Resolution
Range
0.1℃
-50℃~199.9℃
0.1℉
-50℉~199.9℉
".
Accuracy
±(0.2%+1℃)
±(0.2%+1℉)
-50℃~1000℃
1℃
1000℃~1300℃
-50℉~1000℉
1℉
1000℉~1999℉
Note:
1. For obtaining an accurate measurement result, after you turn on the meter, please wait for 3 minutes
to warm-up before taking measurements.
2. The accuracy above doesn't include the temperature error of probe. Please refer to the accuracy of
the probe for reference.
3.5 Specifications of temperature probe
Model
TP01
Specification
(-50~250)℃
Range
(-58~1832)℉
(-58~482)℉
Contact point type
Contact type
Dimension
Bead
Heat resistance
Handle
260℃
Extension line
length: 1m
Thermocouple
Accuracy
± 1.5℃
For measuring
For measuring temperature of liquid and
Application
temperature in
narrow space
Picture ①
Picture
Picture ①
TP01
Picture ③
Surface probe
± (0.3%+2℃)
±(0.5%+2℃)
±(0.3%+2℉)
±(0.5%+2℉)
TP02
TP03
(-50~750)℃
(-50~1300)℃
(-58~2372)℉
Insert type
Insert type
Ø3.2×150mm
Ø6.2×300mm
Heat resistance
Heat resistance
150℃
150℃
Ø12×100mm
Ø12×100mm
(-20~90)℃ length: 1m
K (CA)NiCr/NiAl alloy thermocouple
± 0.4%
± 0.3%
semisolid
Picture ②
Picture ②
TP02, TP03
Surface probe
(-50~750)℃
(-58~1382)℉
Contact type
Ø15×100mm
Heat resistance
150℃
Ø12×90mm
For measuring
surface
temperature
Picture ③

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the TP-01 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for HTC TP-01

  • Page 1 ± (0.3%+2℃) DIGITAL THERMOMETER -50℃~1000℃ 1℃ ±(0.5%+2℃) 1000℃~1300℃ INSTRUCTION MANUAL ±(0.3%+2℉) -50℉~1000℉ 1℉ ±(0.5%+2℉) 1000℉~1999℉ 1. GENERAL DESCRIPTION Note: 1. For obtaining an accurate measurement result, after you turn on the meter, please wait for 3 minutes This 3 1/2 digital thermometer with highly stable performance and high accuracy is driven by 9V to warm-up before taking measurements.
  • Page 2 4. OPERATION INSTRUCTION measurement result, the insert depth should be 15-20 times of the diameter of the probe. 4.1 Preparation before measurement 4.1.1 Check the installation of the battery. If “ ” shows up, please replace it with a new battery. 4.1.2 Make sure the setting of function keys is correct.

This manual is also suitable for:

Tp-02Tp-03

Table of Contents